ABIOMED Insertion Kit, Pkgd, JP, Pump Set Code 0052-0011-JP; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
ABIOMED Insertion Kit, Pkgd, JP, Pump Set Code 0052-0011-JP; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
ABIOMED Impella RP Flex with SmartAssist Set, Pump Set Code 1000323, containing Kit, 23Fr Introducer, 11cm, Sterile, 23 Fr, 11 cm, Product Code 2000342. heart pump and accessories
ABIOMED Impella RP Flex with SmartAssist Set, Pump Set Code 1000323, containing Kit, 23Fr Introducer, 11cm, Sterile, 23 Fr, 11 cm, Product Code 2000342. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula
Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula
Medtronic received two complaints of loose foreign material confirmed as containing trace levels of blood.
Arterial Line Kit 20ga x 6, REF: 400115A
Arterial Line Kit 20ga x 6, REF: 400115A
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
Robotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the product, which may expose a section of instrument arm, which may result in procedure delay or conversion to an alternate minimally invasive approach. Retrospectively reported recall.
Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK-LF DYNJ61651J
Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK-LF DYNJ61651J
Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.
Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.
Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Coach and Equipment Mfg. Corp. - Wheelchair Lift May Inadvertently Move
Coach and Equipment Mfg. Corp. - Wheelchair Lift May Inadvertently Move
A wheelchair lift that continues to move while occupied increases the risk of injury.
Hiab USA Inc. - Crane Column May Fail
Hiab USA Inc. - Crane Column May Fail
A cracked weld may result in the collapse of the crane column when in use, increasing the risk of injury to the operator or others nearby.
Daimler Trucks North America, LLC - Insufficient Illumination from Low Beam Headlights
Daimler Trucks North America, LLC - Insufficient Illumination from Low Beam Headlights
Insufficient low beam headlight illumination can reduce driver visibility, increasing the risk of a crash.
Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.
Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.
Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204
ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
General Biospy Tray, REF: GUTS1000
General Biospy Tray, REF: GUTS1000
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Optima XR220amx, Mobile Digital-Ready Radiographic System
Optima XR220amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
meijer, Carbamide Peroxide 6.5%, 0.5 FL OZ (15 mL), Washer Bulb included, NDC 41250-835-33, DIST. BY MEIJER DISTRIBUTION, INC., GRAND RAPIDS, MI 49544.
meijer, Carbamide Peroxide 6.5%, 0.5 FL OZ (15 mL), Washer Bulb included, NDC 41250-835-33, DIST. BY MEIJER DISTRIBUTION, INC., GRAND RAPIDS, MI 49544.
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).
TopCare Health, CARBAMIDE PEROXIDE 6.5%, 0.5 FL OZ (15 mL) a) Kit - WASHER BULB & DROPS INCLUDED (NDC 36800-835-33); b) 1 bottle (NDC 36800-835-34), DISTRIBUTED BY TOPCO ASSOCIATES LLC, ELK GROVE VILLAGE, IL 60007.
TopCare Health, CARBAMIDE PEROXIDE 6.5%, 0.5 FL OZ (15 mL) a) Kit - WASHER BULB & DROPS INCLUDED (NDC 36800-835-33); b) 1 bottle (NDC 36800-835-34), DISTRIBUTED BY TOPCO ASSOCIATES LLC, ELK GROVE VILLAGE, IL 60007.
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).
Nystatin Cream USP, (100,000 USP Nystatin Units), NET WT 15 g, Rx only, 15 g tube, Manufactured by Padagis, Yeruham, Israel, NDC 45802-059-35
Nystatin Cream USP, (100,000 USP Nystatin Units), NET WT 15 g, Rx only, 15 g tube, Manufactured by Padagis, Yeruham, Israel, NDC 45802-059-35
Subpotent Drug: Low out of specification assay results performed during long-term stability testing.
CVS Health, Carbamide Peroxide, 6.5%, 0.5 FL Oz (15 mL), a) Bottle (NDC 51316-822-00.); b) Kit (1 BULB, Syringe & Drops- NDC 51316-823-00.), Distributed by: CVS Pharmacy., Inc. One CVS Drive, Woonsocket, RI 02895,
CVS Health, Carbamide Peroxide, 6.5%, 0.5 FL Oz (15 mL), a) Bottle (NDC 51316-822-00.); b) Kit (1 BULB, Syringe & Drops- NDC 51316-823-00.), Distributed by: CVS Pharmacy., Inc. One CVS Drive, Woonsocket, RI 02895,
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).
AUDIOLOGIST'S CHOICE, Carbamide Peroxide 6.5%, 0.5 Fl Oz (15 mL) a) Kit - Washer Bulb Included (NDC 59256-001-02); b) 1 bottle (NDC 59256-836-34), Distributed By: Oaktree Products Inc., St. Louis. MO 63005.
AUDIOLOGIST'S CHOICE, Carbamide Peroxide 6.5%, 0.5 Fl Oz (15 mL) a) Kit - Washer Bulb Included (NDC 59256-001-02); b) 1 bottle (NDC 59256-836-34), Distributed By: Oaktree Products Inc., St. Louis. MO 63005.
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).
GOOD NEIGHBOR PHARMACY, Carbamide Peroxide 6.5%, 0.5 Fl Oz (15 mL), a) Kit - Washer Bulb Included, (NDC 46122-556-05); b) 1 bottle (NDC 46122-557-05), Distributed by: AmerisourceBergen, 1 West First Avenue, Conshohocken, PA 19428.
GOOD NEIGHBOR PHARMACY, Carbamide Peroxide 6.5%, 0.5 Fl Oz (15 mL), a) Kit - Washer Bulb Included, (NDC 46122-556-05); b) 1 bottle (NDC 46122-557-05), Distributed by: AmerisourceBergen, 1 West First Avenue, Conshohocken, PA 19428.
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).