ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories
ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Coach and Equipment Mfg. Corp. - Wheelchair Lift May Inadvertently Move
Coach and Equipment Mfg. Corp. - Wheelchair Lift May Inadvertently Move
A wheelchair lift that continues to move while occupied increases the risk of injury.
Aluminum Trailer Company (ATC) - Wireless Remote May Fail to Stop Winch
Aluminum Trailer Company (ATC) - Wireless Remote May Fail to Stop Winch
A winch that operates unintentionally may increase the risk of injury.
Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.
Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Orthofix MINIRAIL, SHORT, LENGTHENER REF: M103, Unilateral orthopedic/trauma external fixation device
Orthofix MINIRAIL, SHORT, LENGTHENER REF: M103, Unilateral orthopedic/trauma external fixation device
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
ABIOMED Impella 5.5 Smart Assist Set, JP, Pump Set Codes 1000211, 1000212; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
ABIOMED Impella 5.5 Smart Assist Set, JP, Pump Set Codes 1000211, 1000212; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Papablic Replacement Brush Heads. Model Number: Papablic18B
Papablic Replacement Brush Heads. Model Number: Papablic18B
Potential for brush heads to break.
Papablic Replacement Brush Heads. Model Number: PAPA317AA.
Papablic Replacement Brush Heads. Model Number: PAPA317AA.
Potential for brush heads to break.
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Clinere, Carbamide Peroxide, a) 2 bottles of Clinere Carbamide Peroxide (2 x 0.50 mL) 6.5%, 0.5 fl. oz. (15 mL); b) 1 bottle of Clinere Carbamide Peroxide (0,50 fl. oz. (15 mL), Dist. by: Quest Products LLC, Pleasant Prairie, WI 53158, NDC 68229-102-01.
Clinere, Carbamide Peroxide, a) 2 bottles of Clinere Carbamide Peroxide (2 x 0.50 mL) 6.5%, 0.5 fl. oz. (15 mL); b) 1 bottle of Clinere Carbamide Peroxide (0,50 fl. oz. (15 mL), Dist. by: Quest Products LLC, Pleasant Prairie, WI 53158, NDC 68229-102-01.
SubPotent Drug: low pH and significantly reduced assay (~0.6% vs expected ~6 7%).
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK-LF DYNJ61651J
Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK-LF DYNJ61651J
Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.
ABIOMED Impella 5.5 SmartAssist S2 Set, JP, Pump Set Codes 1000459, 1000460; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
ABIOMED Impella 5.5 SmartAssist S2 Set, JP, Pump Set Codes 1000459, 1000460; containing Introducer Product Code 0052-3006-JP. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
T-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL
T-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
Daimler Trucks North America, LLC - Insufficient Illumination from Low Beam Headlights
Daimler Trucks North America, LLC - Insufficient Illumination from Low Beam Headlights
Insufficient low beam headlight illumination can reduce driver visibility, increasing the risk of a crash.
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.
Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
ABIOMED Intro Kit, 14Fr x 13 and 25 cm, PKGD, JP, Pump Set Code 0052-0023-JP; containing Introducer Product Code 0052-3025-JP. heart pump and accessories
ABIOMED Intro Kit, 14Fr x 13 and 25 cm, PKGD, JP, Pump Set Code 0052-0023-JP; containing Introducer Product Code 0052-3025-JP. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation device
Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation device
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
Optima XR200amx, Mobile Digital-Ready Radiographic System
Optima XR200amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Optima XR240amx, Mobile Digital Radiographic System
Optima XR240amx, Mobile Digital Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Rapid Refill Continuous Injection System, UPN M00566001
Rapid Refill Continuous Injection System, UPN M00566001
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
Orthofix MINIRAIL, ARTICULATED, FIXATOR, VERTICAL AXIS REF: M122, Unilateral orthopedic/trauma external fixation device
Orthofix MINIRAIL, ARTICULATED, FIXATOR, VERTICAL AXIS REF: M122, Unilateral orthopedic/trauma external fixation device
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
ABIOMED Impella CP Introducer Kit 14 Fr, 25 cm, Pump Set Code 0052-0039, containing Kit, 14Fr Introducer, 25cm, Long Taper, Introducer Product Code 0052-3053. heart pump and accessories
ABIOMED Impella CP Introducer Kit 14 Fr, 25 cm, Pump Set Code 0052-0039, containing Kit, 14Fr Introducer, 25cm, Long Taper, Introducer Product Code 0052-3053. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
One-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
One-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.