Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
Recalling Firm: Trividia Health, Inc.
Grund für den Rückruf / Gefahr
TranslatedThe system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Produktbeschreibung & Identifikation
TranslatedOwner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
Betroffene Produkte
TranslatedOwner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
Additional Source Details
| Field | Value |
|---|---|
| City | Fort Lauderdale |
| State | FL |
| Event id | 98317 |
| Address 1 | 2400 Nw 55th Ct |
| Address 2 | N/A |
| Code info | Device Item/UDI-DI: REA4001-01/311917173894, REA4002-01/96295129267, REA4007-01/50428354766, REA4011-01/87701427398, REA4019-01/708820685659, REA4020-01/41415072787, REA4025-01/93351022784, REA4031-01/41220009169, REA4054-01/10939742445, REA4066-01/11822590051, REA4081-01/21292015015, REA4081-40/21292007164, REA4081-43/21292007584, REA4089-01/52569139042, REA4094-01/681131240154, REA4094-01RX/681131403214, REA4203-01/41260007354, REA4211-01/10939958617, REA4H01-01/21292007010, REA4H01-40/21292007447, REA4H01-43/21292007911, International: REA4i03-01/N/A, REA4i03-02/N/A, REA4i04-00/21292012731, REA4i81-11/N/A, REA4i82-11/21292012212, REA4i82-12/21292009991. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX AIR Owner's Booklet/ REA4TVH03 Rev 57 and prior, TRUE METRIX AIR Owner's Booklet (English)/ REA4TVH35 Rev 51 and prior, TRUE METRIX AIR Owner's Booklet (Spanish)/ REA4TVH35S Rev 51 and prior, Relion (Walmart) TRUE METRIX AIR Owners Booklet/ REA4RLN03 Rev 52 and ... [TRUNCATED] |
| Postal code | 33309-2672 |
| Report date | 20260318 |
| Product type | Devices |
| Product quantity | 3,678,026 |
| Reason for recall | The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260206 |
| Initial firm notification | Letter |
| Center classification date | 20260312 |
Consumer Action Protocol: What You Should Do Immediately
Safety StepsInspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.
Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.
Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.
If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).
Your Legal Rights as a Consumer
Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.
Frequently Asked Questions About This Recall
FAQ GuideQ1:What should I do immediately if I own or bought Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Descripti...?
Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.
Q2:Can I get a full refund or free repair for this recall?
Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.
Q3:Do I need my original sales receipt to claim a remedy or refund?
In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.
Q4:Where can I verify official information for this safety alert?
This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.
Prüfen Sie weitere Produkte in Ihrem Haushalt
Kostenlose ToolsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Übersicht
- Rückrufendes UnternehmenTrividia Health, Inc.
- StatusOngoing
- RisikostufeClass I
- DistributionWorldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
📢 Warnung teilen
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: Trividia Health, Inc. - Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12 Check if your household is affected: https://therecallfile.com/de/recalls/owner-s-booklets-and-instructions-for-use-that-are-used-with-the-follo/FDA-Z-1547-2026
Alarm-Liste abonnieren
Erhalten Sie sofortige E-Mail-Benachrichtigungen bei kritischen Produktrückrufen.
Support The Recall File
Help keep official recall alerts free, accessible, and fast for everyone.