MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
Recalling Firm: Medtronic MiniMed, Inc.
Grund für den Rückruf / Gefahr
TranslatedAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Produktbeschreibung & Identifikation
TranslatedMiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
Additional Source Details
| Field | Value |
|---|---|
| City | Northridge |
| State | CA |
| Event id | 98186 |
| Address 1 | 18000 Devonshire St |
| Code info | Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 740G Insulin Pump: 763000504908, 763000504915, 763000505073, 763000484637, 763000505042, 763000504861, 763000629984, 763000504946, 763000504878, 763000559557, 763000504960, 763000505080, 763000505066, 763000629991, 763000505004, 763000504922, 763000504892, 763000505059, 763000504977, 763000505028, 763000505172, 763000505127, 763000505134, 763000505158, 763000505141, 763000505103, 763000505110, 763000504991, 763000505165, 763000504939, 763000504953, 763000504984, 763000505189, 763000851224, 763000851217, 763000769451, 763000769468: MINIMED 740G SYSTEM USER GUIDE mg/dL(M000149C002 1, M000149C004 1, M000149C005 1, M000150C005 1, M000150C002 1, M000150C021 2, M000149C016 1, M000149C021 1, M000150C004 1, M000150C025 1, M000150C030 1, M000149C014 1, M000149C006 1, M000150C003 1, M000150C015 1, M000150C028 1, M000150C018 1, M000150C022 1, M000150C007 1, M000149C017 1, M000149C008 1, M000150C008 1, M000150C032 1, M000150C029 1, M0001... [TRUNCATED] |
| Postal code | 91325-1219 |
| Report date | 20260408 |
| Product type | Devices |
| Product quantity | 24,946 |
| Reason for recall | All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260213 |
| Initial firm notification | Letter |
| Center classification date | 20260402 |
Consumer Action Protocol: What You Should Do Immediately
Safety StepsInspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.
Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.
Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.
If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).
Your Legal Rights as a Consumer
Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.
Frequently Asked Questions About This Recall
FAQ GuideQ1:What should I do immediately if I own or bought MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)?
Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.
Q2:Can I get a full refund or free repair for this recall?
Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.
Q3:Do I need my original sales receipt to claim a remedy or refund?
In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.
Q4:Where can I verify official information for this safety alert?
This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.
Prüfen Sie weitere Produkte in Ihrem Haushalt
Kostenlose ToolsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Übersicht
- Rückrufendes UnternehmenMedtronic MiniMed, Inc.
- StatusOngoing
- RisikostufeClass II
- DistributionWorldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Macedonia, Malawi, Malta, Martinique, Mauritius, Mayotte, Moldavia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Qatar, Reunion, Romania, Russia, San Marino, Saudi Arabia, Serbia, Sierra Leone, Slovakia, Slovenia, South Africa, Spain, St.Pier,Miquel., Sudan, Sweden, Switzerland, Syrian Arab Republic, Tanzania, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Yemen, Zambia, Zimbabwe.
📢 Warnung teilen
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: Medtronic MiniMed, Inc. - MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862) Check if your household is affected: https://therecallfile.com/de/recalls/minimed-740g-insulin-pump-mmt-1811-mmt-1812-mmt-1861-mmt-1862/FDA-Z-1746-2026
Alarm-Liste abonnieren
Erhalten Sie sofortige E-Mail-Benachrichtigungen bei kritischen Produktrückrufen.
Support The Recall File
Help keep official recall alerts free, accessible, and fast for everyone.