healthmark ProChek-II, Model Number PT-313, PT-201-12
healthmark ProChek-II, Model Number PT-313, PT-201-12
Recalling Firm: Healthmark Industries Co., Inc.
Grund für den Rückruf / Gefahr
TranslatedTest manufactured with an expired ingredient which may present as either a false negative or positive.
Produktbeschreibung & Identifikation
Translatedhealthmark ProChek-II, Model Number PT-313, PT-201-12
Additional Source Details
| Field | Value |
|---|---|
| City | Fraser |
| State | MI |
| Event id | 99537 |
| Address 1 | 18600 Malyn Blvd |
| Address 2 | N/A |
| Code info | PT-313 - Serial/Lot Numbers: 3000535312, 3000535333, 3000536416, 3000536423, 3000537810, 3000537822, 3000542453, 3000545623, 3000550099, 3000550512, 313069, 313072, 313073, 313074, 313075, 313076, 313077, 313078, 313079, 313080, 313081, 313082, 313083, 313084, 313085, 313086, 313087, 313088, 313089, 313090, 313091, 313092, 313093, 313094, 313095, 313096, 313097, 313098, 313099, 313100, 313101, 313102, 313106, 313107; H-PT-201-12 - Serial/Lot Numbers: 7P0110, 7P0111, 7P0113, 7P0122. |
| Postal code | 48026-3496 |
| Report date | 20260902 |
| Product type | Devices |
| Product quantity | 5090 units |
| Reason for recall | Test manufactured with an expired ingredient which may present as either a false negative or positive. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260729 |
| Initial firm notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
| Center classification date | 20260826 |
Consumer Action Protocol: What You Should Do Immediately
Safety StepsInspect the NDC code, batch number, or control number on your medicine bottle or blister packaging to confirm whether your unit is included.
Do not abruptly stop critical prescribed medication without physician advice. Contact your dispensing pharmacist immediately for a safe alternative.
Under statutory consumer protection rules, consumers are entitled to a full refund, certified replacement, or free repair kit without needing the original sales receipt.
If you suffered an injury, illness, or property damage from this product, file an official incident report with regulatory authorities (FDA Devices).
Your Legal Rights as a Consumer
Recalls are legally binding corrective safety actions overseen by statutory government bodies. Recalling entities are legally prohibited from charging consumers for official safety remedies. If a retailer or dealership refuses your recall remedy, notify the issuing regulatory agency immediately.
Frequently Asked Questions About This Recall
FAQ GuideQ1:What should I do immediately if I own or bought healthmark ProChek-II, Model Number PT-313, PT-201-12?
Check the lot number and NDC on your prescription label. For essential medications, consult your doctor or dispensing pharmacist immediately before abruptly stopping your dose.
Q2:Can I get a full refund or free repair for this recall?
Yes. Recalling manufacturers and retail merchants are required under consumer protection regulations to offer a statutory remedy—typically a full refund, certified repair kit, or replacement unit.
Q3:Do I need my original sales receipt to claim a remedy or refund?
In most recall cases, no receipt is required. Because safety recalls are matters of public health and statutory compliance, retailers and manufacturers routinely issue refunds, store credit, or replacement kits upon presentation of the affected product or lot code.
Q4:Where can I verify official information for this safety alert?
This safety alert is aggregated directly from official regulatory agency records (FDA Devices). You can click the 'Official Agency Alert' link on this page to view the primary government notice and contact the manufacturer directly.
Prüfen Sie weitere Produkte in Ihrem Haushalt
Kostenlose ToolsSafety recalls are issued daily across foods, medications, vehicles, and electronics. Use our instant tools to check your home for active safety warnings:
Übersicht
- Rückrufendes UnternehmenHealthmark Industries Co., Inc.
- StatusOngoing
- RisikostufeClass II
- DistributionWorldwide distribution - US Nationwide and the countries of Canada, France, Hong Kong, South Korea.
📢 Warnung teilen
Alert family and friends to keep them safe. Copy this quick alert message:
🚨 SAFETY ALERT: Healthmark Industries Co., Inc. - healthmark ProChek-II, Model Number PT-313, PT-201-12 Check if your household is affected: https://therecallfile.com/de/recalls/healthmark-prochek-ii-model-number-pt-313-pt-201-12/FDA-Z-3031-2026
Alarm-Liste abonnieren
Erhalten Sie sofortige E-Mail-Benachrichtigungen bei kritischen Produktrückrufen.
Support The Recall File
Help keep official recall alerts free, accessible, and fast for everyone.