Medical DeviceUSFDA Devices

FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

Boston Scientific Corporation
FDA DevicesClass II

FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Jul 29, 2026Boston Scientific Corporation
PharmaceuticalUSFDA Drugs

Rohto Cooling Eye Drops Cool Relief (naphazoline hydrochloride 0.012%, polysorbate 80 0.2%), Sterile, 0.4 FL OZ (13 mL) each, Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 310742010749, NDC 10742-8141-1

Rohto-Mentholatum (Vietnam) Co., Ltd.
FDA DrugsClass II

Rohto Cooling Eye Drops Cool Relief (naphazoline hydrochloride 0.012%, polysorbate 80 0.2%), Sterile, 0.4 FL OZ (13 mL) each, Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 310742010749, NDC 10742-8141-1

Lack of Assurance of Sterility

Jul 29, 2026Rohto-Mentholatum (Vietnam) Co., Ltd.
Medical DeviceUSFDA Devices

FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

Boston Scientific Corporation
FDA DevicesClass II

FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Jul 29, 2026Boston Scientific Corporation
PharmaceuticalUSFDA Drugs

Dabigatran etexilate Capsules, 110 mg, 60-count carton (6x10), Rx Only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7254-68.

Ascend Laboratories, LLC
FDA DrugsClass III

Dabigatran etexilate Capsules, 110 mg, 60-count carton (6x10), Rx Only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7254-68.

Failed Impurities/Degradation Specifications: OOS result was observed during the related substance test analysis with respect to Impurity II, Impurity III and Total impurity testing for 12 Month Long-Term Stability.

Jul 29, 2026Ascend Laboratories, LLC
PharmaceuticalUSFDA Drugs

Cetirizine Hydrochloride Tablets USP 5 mg, 100-count bottle, Manufactured by: Unique Pharmaceuticals Labs, (A Div. of J.B. Chemicals & Pharmaceuticals, Ltd.), Mumbai 400 030, India. Distributed by: Rising Pharma Holdings, Inc., East Brunswick, NJ 08816, NDC 16571-401-10.

JB Chemicals and Pharmaceuticals Ltd
FDA DrugsClass I

Cetirizine Hydrochloride Tablets USP 5 mg, 100-count bottle, Manufactured by: Unique Pharmaceuticals Labs, (A Div. of J.B. Chemicals & Pharmaceuticals, Ltd.), Mumbai 400 030, India. Distributed by: Rising Pharma Holdings, Inc., East Brunswick, NJ 08816, NDC 16571-401-10.

Cross Contamination with Other Products: Customer complaints for appearance of discolored tablets and red dots observed on Cetirizine Hydrochloride Tablets USP 5 mg. Determined to be Ranitidine.

Jul 29, 2026JB Chemicals and Pharmaceuticals Ltd
PharmaceuticalUSFDA Drugs

Dabigatran etexilate Capsules, 150 mg, 20-count carton (2x10), Rx only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7255-10.

Ascend Laboratories, LLC
FDA DrugsClass III

Dabigatran etexilate Capsules, 150 mg, 20-count carton (2x10), Rx only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7255-10.

Failed Impurities/Degradation Specifications: OOS result was observed during the related substance test analysis with respect to Impurity II, Impurity III and Total impurity testing for 12 Month Long-Term Stability.

Jul 29, 2026Ascend Laboratories, LLC
Medical DeviceUSFDA Devices

Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P

Medline Industries, LP
FDA DevicesClass II

Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P

Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.

Jul 29, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).

Prismatik Dentalcraft, Inc.
FDA DevicesClass II

Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).

Due to device not having FDA Clearance

Jul 29, 2026Prismatik Dentalcraft, Inc.
Medical DeviceUSFDA Devices

3mensio Workstation (Vascular Fenestrated) software

PIE Medical Imaging B.V.
FDA DevicesClass II

3mensio Workstation (Vascular Fenestrated) software

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Jul 29, 2026PIE Medical Imaging B.V.
PharmaceuticalUSFDA Drugs

Levothyroxine Sodium Tablets, USP, 150 mcg (0.0150 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal Health, Dublin, OH 43017. NDC Bag: 55154-3563-0; NDC Blister: 0904-6956-61

Major Pharmaceuticals
FDA DrugsClass II

Levothyroxine Sodium Tablets, USP, 150 mcg (0.0150 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal Health, Dublin, OH 43017. NDC Bag: 55154-3563-0; NDC Blister: 0904-6956-61

Subpotent Drug

Jul 29, 2026Major Pharmaceuticals
Medical DeviceUSFDA Devices

SeaSpine Polyaxial Head REF Number: 41-3010

SEASPINE ORTHOPEDICS CORPORATION
FDA DevicesClass II

SeaSpine Polyaxial Head REF Number: 41-3010

Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.

Jul 29, 2026SEASPINE ORTHOPEDICS CORPORATION
Vehicle RecallUSNHTSA

Toyota Motor Engineering & Manufacturing - Side Curtain Air Bag May Deploy Improperly

Toyota Motor Engineering & Manufacturing
NHTSAAIR BAGS

Toyota Motor Engineering & Manufacturing - Side Curtain Air Bag May Deploy Improperly

Improper air bag deployment can increase the risk of injury during a crash.

Jul 29, 2026Toyota Motor Engineering & Manufacturing
Medical DeviceUSFDA Devices

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

Boston Scientific Corporation
FDA DevicesClass II

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Jul 29, 2026Boston Scientific Corporation
PharmaceuticalUSFDA Drugs

Levothyroxine Sodium Tablets, USP, 112 mcg (0.0112 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6954-61

Major Pharmaceuticals
FDA DrugsClass II

Levothyroxine Sodium Tablets, USP, 112 mcg (0.0112 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6954-61

Subpotent Drug

Jul 29, 2026Major Pharmaceuticals
Medical DeviceUSFDA Devices

Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.

Emano Metrics, Inc.
FDA DevicesClass II

Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.

Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.

Jul 29, 2026Emano Metrics, Inc.
Vehicle RecallUSNHTSA

Toyota Motor Engineering & Manufacturing - Side Curtain Air Bag May Deploy Improperly

Toyota Motor Engineering & Manufacturing
NHTSAAIR BAGS

Toyota Motor Engineering & Manufacturing - Side Curtain Air Bag May Deploy Improperly

Improper air bag deployment can increase the risk of injury during a crash.

Jul 29, 2026Toyota Motor Engineering & Manufacturing
Medical DeviceUSFDA Devices

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

Boston Scientific Corporation
FDA DevicesClass II

FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter

Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.

Jul 29, 2026Boston Scientific Corporation
PharmaceuticalUSFDA Drugs

Glycopyrrolate Oral Solution, 1 mg/5 mL, delivers 5 mL, Rx only, Pkg: Precision Dose, Inc, S. Beloil, IL 61080, unit-dose oral syringes, NDCs 68094-073-01 (oral syringe) and 68094-073-58 (case).

Precision Dose Inc.
FDA DrugsClass II

Glycopyrrolate Oral Solution, 1 mg/5 mL, delivers 5 mL, Rx only, Pkg: Precision Dose, Inc, S. Beloil, IL 61080, unit-dose oral syringes, NDCs 68094-073-01 (oral syringe) and 68094-073-58 (case).

Failed Impurities/Degradation Specifications

Jul 29, 2026Precision Dose Inc.
Medical DeviceUSFDA Devices

Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B

Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.

Jul 29, 2026Medline Industries, LP
PharmaceuticalUSFDA Drugs

Levothyroxine Sodium Tablets, USP, 88 mcg (0.088 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6952-61

Major Pharmaceuticals
FDA DrugsClass II

Levothyroxine Sodium Tablets, USP, 88 mcg (0.088 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6952-61

Subpotent Drug

Jul 29, 2026Major Pharmaceuticals
Medical DeviceUSFDA Devices

MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.

Medline Industries, LP
FDA DevicesClass I

MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.

The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.

Jul 29, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

ICU Medical Inc.
FDA DevicesClass II

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

Jul 29, 2026ICU Medical Inc.
Medical DeviceUSFDA Devices

FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter

Boston Scientific Corporation
FDA DevicesClass II

FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter

Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.

Jul 29, 2026Boston Scientific Corporation
Medical DeviceUSFDA Devices

Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

Ferno-Washington Inc
FDA DevicesClass II

Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.

Jul 29, 2026Ferno-Washington Inc