Vereinigte Staaten Recalls
Verbraucherrückrufe von FDA, CPSC, NHTSA und USDA FSIS.
Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20
Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20
Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
Dabigatran etexilate Capsules, 150 mg, 20-count carton (2x10), Rx only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7255-10.
Dabigatran etexilate Capsules, 150 mg, 20-count carton (2x10), Rx only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7255-10.
Failed Impurities/Degradation Specifications: OOS result was observed during the related substance test analysis with respect to Impurity II, Impurity III and Total impurity testing for 12 Month Long-Term Stability.
Ventriculostomy Kit, REF: 46154
Ventriculostomy Kit, REF: 46154
Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Medline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU DYNJ65065A: 2) EXTREMITY/FOOT PACK Medline Kit SKU DYNJ65066B: 3) ANTERIOR CERVICAL-SMH Medline Kit SKU DYNJ907718B: 4) LAMINECTOMY-SMH Medline Kit SKU DYNJ907732C: 5) KYPHOPLASTY Medline Kit SKU DYNJ907984C: 6) LUMBAR LAMINECTOMY Medline Kit SKU DYNJ907987D: 7) KYPHOPLASTY TRAY Medline Kit SKU DYNJ908413C: 8) CRANIOTOMY Medline Kit SKU DYNJ910626: 9) NEURO SPINE Medline Kit SKU DYNJ910639.
Medline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU DYNJ65065A: 2) EXTREMITY/FOOT PACK Medline Kit SKU DYNJ65066B: 3) ANTERIOR CERVICAL-SMH Medline Kit SKU DYNJ907718B: 4) LAMINECTOMY-SMH Medline Kit SKU DYNJ907732C: 5) KYPHOPLASTY Medline Kit SKU DYNJ907984C: 6) LUMBAR LAMINECTOMY Medline Kit SKU DYNJ907987D: 7) KYPHOPLASTY TRAY Medline Kit SKU DYNJ908413C: 8) CRANIOTOMY Medline Kit SKU DYNJ910626: 9) NEURO SPINE Medline Kit SKU DYNJ910639.
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116I
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116I
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Medline medical convenience kits labeled as: 1) SHOULDER PACK, Medline Kit SKU DYNJ39590O; 2) SHOULDER PACK, Medline Kit SKU DYNJ88566; 3) MAJOR ORTHO KIT, Medline Kit SKU DYNJ911535.
Medline medical convenience kits labeled as: 1) SHOULDER PACK, Medline Kit SKU DYNJ39590O; 2) SHOULDER PACK, Medline Kit SKU DYNJ88566; 3) MAJOR ORTHO KIT, Medline Kit SKU DYNJ911535.
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. UROLOGY ROBOTIC PROCEDURE PACK DYNJ0945779U
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. UROLOGY ROBOTIC PROCEDURE PACK DYNJ0945779U
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Island Pacific IMITATION CRAB STICKS NET WT: 1.0 LB (454g) UPC: 0000000175029 PRECOOKED SERVE HOT OR COLD INGREDIENTS: POLLOCK AND/OR THREADFIN BREAM, WATER, POTATO STARCH, TAPIOCA STARCH, CORN STARCH, SALT, SUGAR, MIRIN (RICE WINE), SOYBEAN OIL, SOYBEAN PROTEIN, CRAB EXTRACT, ARTIFICIAL CRAB FLAVOR, COLOR ADDED (CARMINE). CONTAINS: FISH (POLLACK AND/OR THREADFIN BREAM), CRUSTACEAN SHELLFISH (CRAB), AND SOYBEAN. PRODUCT OF CHINA ISLAND PACIFIC SUPERMARKET 738 Epperson Drive City of Industry CA 91748
Island Pacific IMITATION CRAB STICKS NET WT: 1.0 LB (454g) UPC: 0000000175029 PRECOOKED SERVE HOT OR COLD INGREDIENTS: POLLOCK AND/OR THREADFIN BREAM, WATER, POTATO STARCH, TAPIOCA STARCH, CORN STARCH, SALT, SUGAR, MIRIN (RICE WINE), SOYBEAN OIL, SOYBEAN PROTEIN, CRAB EXTRACT, ARTIFICIAL CRAB FLAVOR, COLOR ADDED (CARMINE). CONTAINS: FISH (POLLACK AND/OR THREADFIN BREAM), CRUSTACEAN SHELLFISH (CRAB), AND SOYBEAN. PRODUCT OF CHINA ISLAND PACIFIC SUPERMARKET 738 Epperson Drive City of Industry CA 91748
Undeclared sodium tripolyphosphate (STTP)
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter
FARASTAR Pulsed Field Ablation Generator, Product ID M004PFCE61M401; Percutaneous Cardiac Ablation Catheter
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. DAVINCI PROSTATE PACK DYNJ68477L DYNJ68477M
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. DAVINCI PROSTATE PACK DYNJ68477L DYNJ68477M
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
Due to device not having FDA Clearance
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. ROBOTIC CDS980523M
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. ROBOTIC CDS980523M
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.
Flexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal: GYN, REF: 26153EAU/1.
Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.
MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.
The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12
Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12
Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.