Halyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.
Halyard NEURO IR PERIPHERAL (PS 64405) kit. Model Number: UINR44-01.
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Field Safety Notices: 4 to 8 May 2026
Field Safety Notices: 4 to 8 May 2026
List of Field Safety Notices from 4 to 8 May 2026....
Field Safety Notices: 27 April to 1 May 2026
Field Safety Notices: 27 April to 1 May 2026
List of Field Safety Notices from 27 April to 1 May 2026....
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.
Philips Ingenia Ambition X with MR Elastography (MRE). 1. Model Number (REF): 781356. 2. Model Number (REF): 782109. 3. Model Number (REF): 782138. 4. Model Number (REF): 782160.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
TGA warning on the risks of importing unapproved peptide products
TGA warning on the risks of importing unapproved peptide products
We are warning consumers about the risks of importing and using unapproved peptide products that are promoted online....
Field Safety Notices: 1 May 2026
Field Safety Notices: 1 May 2026
List of Field Safety Notices for 1 May 2026....
Medline medical procedure kits, containing Medline Neuro Sponges (961 in total), NEURO,
Medline medical procedure kits, containing Medline Neuro Sponges (961 in total), NEURO,
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Heraeus, PALAMIX duo. Material Number: 66057897.
Heraeus, PALAMIX duo. Material Number: 66057897.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINE PACK-LF, Medline Kit Number/SKU PHS719414B
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINE PACK-LF, Medline Kit Number/SKU PHS719414B
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Heraeus, PALAMIX uno. Material Number: 66057893.
Heraeus, PALAMIX uno. Material Number: 66057893.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670
Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: LATEX FREE GENERAL PURPOSE TRAY, Medline Kit SKU 75670
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medline Kit Number/SKU DYNJ47568B; 3. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555B; 4. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555C; 5. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555D.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medline Kit Number/SKU DYNJ47568B; 3. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555B; 4. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555C; 5. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555D.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
i-STAT G3+ cartridge; List Number: 03P78-26;
i-STAT G3+ cartridge; List Number: 03P78-26;
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
Brand Name: Leksell Vantage Arc System REF: 1053958
Brand Name: Leksell Vantage Arc System REF: 1053958
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLY
Philips Bridge Prep Kit REF: K12-09098C Sterile EO, Rx ONLY
catheter may experience resistance when being advanced over the guidewire
i-STAT EG7+ cartridge; List Number: 03P76-25;
i-STAT EG7+ cartridge; List Number: 03P76-25;
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use
BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use
BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.
Philips Bridge Prep Kit REF: K12-09098B Sterile EO, Rx ONLY
Philips Bridge Prep Kit REF: K12-09098B Sterile EO, Rx ONLY
catheter may experience resistance when being advanced over the guidewire