Safety Recall Directory
Browse and filter the latest official safety alerts globally.
Cuban Crackers Galletas Cubanas, Net Wt 8Oz (226g). Distributed by Best Floridian, LLC. UPC 5700300060
Cuban Crackers Galletas Cubanas, Net Wt 8Oz (226g). Distributed by Best Floridian, LLC. UPC 5700300060
Undeclared colors FD&C yellow #5, FD&C yellow #6, and FD&C red 40
Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/10Shaft/Left/Std/139mm Model/Catalog Number: 02.426.681S Product Description: 2 Column Distal Radius Plate - 6 Head, 10 Shaft, Standard Left - 139mm
Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/10Shaft/Left/Std/139mm Model/Catalog Number: 02.426.681S Product Description: 2 Column Distal Radius Plate - 6 Head, 10 Shaft, Standard Left - 139mm
The supplier manufactured anatomical left plates with an incorrect thread orientation.
Brand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product Description: iT20 collects physiological parameters by ECG cables and SpO2 sensors, then achieves data analyzing and processing. After that, data will be sent to MFM-CMS via Wi-Fi. The parameters supported are ECG, SpO2 and PR. Component: No
Brand Name: EDAN Product Name: Telemetry Transmitter Model/Catalog Number: iT20 Product Description: iT20 collects physiological parameters by ECG cables and SpO2 sensors, then achieves data analyzing and processing. After that, data will be sent to MFM-CMS via Wi-Fi. The parameters supported are ECG, SpO2 and PR. Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Brand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS, MFM-CNS Lite Product Description: The MFM-CNS and MFM-CNS Lite are clinical data managing software applications. Both applications manage clinical data of fetal monitoring and uterine activity, and the MFM-CNS additional monitors maternal vital signs. Data are automatically acquired from bedside monitors, for the purpose of collecting. processing and saving the patient and/or clinical data that is normally provided on record papers and/or separate bedside monitors. They provide electronic medical records and operate with off-the-shelf software and hardware. Component: No
Brand Name: EDAN Product Name: Central Monitoring System Model/Catalog Number: MFM-CNS, MFM-CNS Lite Product Description: The MFM-CNS and MFM-CNS Lite are clinical data managing software applications. Both applications manage clinical data of fetal monitoring and uterine activity, and the MFM-CNS additional monitors maternal vital signs. Data are automatically acquired from bedside monitors, for the purpose of collecting. processing and saving the patient and/or clinical data that is normally provided on record papers and/or separate bedside monitors. They provide electronic medical records and operate with off-the-shelf software and hardware. Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Anthony's Barbecue & Dip-It Sauce, 15 oz. (425g) glass bottle, UPC 6 89076 62272 1. Anthony's BBQ Sauce PO Box 3239 Harbor, OR 97415. The ingredients statement on the recalled bottle is declared as "***Ingredients: Tomatoes, Brown Sugar, Distilled Water, Distilled Vinegar, Garlic, Onion, Liquid Smoke, Salt, Mustard Seed, Natural Flavor, Corn syrup, Turmeric, Paprika, Pepper, Hydrolyzed soy protein, Caramel color, 1/10 of 1% Benzoate of Soda***.
Anthony's Barbecue & Dip-It Sauce, 15 oz. (425g) glass bottle, UPC 6 89076 62272 1. Anthony's BBQ Sauce PO Box 3239 Harbor, OR 97415. The ingredients statement on the recalled bottle is declared as "***Ingredients: Tomatoes, Brown Sugar, Distilled Water, Distilled Vinegar, Garlic, Onion, Liquid Smoke, Salt, Mustard Seed, Natural Flavor, Corn syrup, Turmeric, Paprika, Pepper, Hydrolyzed soy protein, Caramel color, 1/10 of 1% Benzoate of Soda***.
Undeclared allergen (fish). Anthony's Barbecue & Dip-It Sauce label does not declare Worcestershire (Anchovy).
Rifton Equipment, E-Pacer. Model/Catalog Number: K660. The E-Pacer is a gait training device with power height adjustment that facilitates the sit-to-stand motion.
Rifton Equipment, E-Pacer. Model/Catalog Number: K660. The E-Pacer is a gait training device with power height adjustment that facilitates the sit-to-stand motion.
Potential for fraying on the body support strap.
Brand Name: LEASEIR Product Name: LEASEIR MHR XCELL Model/Catalog Number: MHR 110-b Software Version: 1.5.0.0 Product Description: The Leaseir MHR Xcell is a surgical laser instrument for use in general and plastic surgery and dermatology, intended for hair removal and treatment of pseudo folliculitis barbae on all skin types (Fitzpatrick I-VI). The process implies the generation of intense light pulses at specific wavelengths. The specific nature of the energies and pulse durations cause the desired effect (in the hair follicles and oxyhemoglobin in the blood) heating them sufficiently and destroying them without unnecessary damage to surrounding tissue. This is achieved by controlled emission of laser radiation to the target tissue. The system consists of a main console and two interchangeable applicators that delivers pulsed light in the range of 800-820nm with a peak in the 810nm in different spot sizes. Two of the applicators emitting radiation at 810 nm wavelength for two areas 13.5¿15 mm¿ and 14.5¿25.5 mm¿. Two different operation modes are available: static mode and dynamic mode, which are differentiated basically by the frequency range defined for each mode (1-4 Hz for static and 10 Hz for dynamic). The principle of operation consists in the photons travel along the axis and reflected again back into the crystal, continuing the chain reaction, while photons travelling in different directions leave the crystal. In one of the two mirrors, a tiny hole allows a small amount of light to leak out and the resulting beam is focused with lenses and is emitted from the laser. The total energy emitted by the Leaseir MHR Xcell is produced by an array of diodes. The laser emission is activated by the hand piece trigger, deliver a continuous pulse pattern while the button is pressed. Component: NO
Brand Name: LEASEIR Product Name: LEASEIR MHR XCELL Model/Catalog Number: MHR 110-b Software Version: 1.5.0.0 Product Description: The Leaseir MHR Xcell is a surgical laser instrument for use in general and plastic surgery and dermatology, intended for hair removal and treatment of pseudo folliculitis barbae on all skin types (Fitzpatrick I-VI). The process implies the generation of intense light pulses at specific wavelengths. The specific nature of the energies and pulse durations cause the desired effect (in the hair follicles and oxyhemoglobin in the blood) heating them sufficiently and destroying them without unnecessary damage to surrounding tissue. This is achieved by controlled emission of laser radiation to the target tissue. The system consists of a main console and two interchangeable applicators that delivers pulsed light in the range of 800-820nm with a peak in the 810nm in different spot sizes. Two of the applicators emitting radiation at 810 nm wavelength for two areas 13.5¿15 mm¿ and 14.5¿25.5 mm¿. Two different operation modes are available: static mode and dynamic mode, which are differentiated basically by the frequency range defined for each mode (1-4 Hz for static and 10 Hz for dynamic). The principle of operation consists in the photons travel along the axis and reflected again back into the crystal, continuing the chain reaction, while photons travelling in different directions leave the crystal. In one of the two mirrors, a tiny hole allows a small amount of light to leak out and the resulting beam is focused with lenses and is emitted from the laser. The total energy emitted by the Leaseir MHR Xcell is produced by an array of diodes. The laser emission is activated by the hand piece trigger, deliver a continuous pulse pattern while the button is pressed. Component: NO
The console label for affected devices is missing the "DANGER" symbol.
Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261
Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261
Inflation device handle may detach from the syringe during procedure.
Rifton Equipment, TRAM. Model/Catalog Number: K310. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training.
Rifton Equipment, TRAM. Model/Catalog Number: K310. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training.
Potential for fraying on the body support strap.
Safety concerns over imported counterfeit melatonin products
Safety concerns over imported counterfeit melatonin products
We are issuing an updated safety advisory regarding imported counterfeit unregistered melatonin products....
BeneHeart C Series AED: Non-rechargeable public automated external defibrillator
BeneHeart C Series AED: Non-rechargeable public automated external defibrillator
Mindray Medical Australia Pty Ltd is conducting a product correction for the BeneHeart C Series Automated External Defibrillator (AED)....
Baby harness strap lock buckle clip adjuster

Baby harness strap lock buckle clip adjuster
This product does not comply with the mandatory standard for child restraints for use in motor vehicles .
Australian Greenlip Abalone, Frozen, 9 pieces, Net Wt. 23.5 OZ UPC: 9335533013342
Australian Greenlip Abalone, Frozen, 9 pieces, Net Wt. 23.5 OZ UPC: 9335533013342
Potential Listeria Monocytogenes contamination
Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Software Version: N/A Product Description: The Ion Endoluminal System is a robotically assisted platform for minimally invasive peripheral lung biopsy. Ion consists of a System Cart and Controller that are used to navigate a flexible Catheter within the pulmonary tract using endoscopic visualization for diagnostic and therapeutic procedures. The Ion Fully Articulating Catheter (referred to as the Catheter) consists of the housing, sensor, input discs, shaft, tool channel, and tip. The Catheter is installed on the carriage of the arm. The Catheter is provided non-sterile and is a multi-use device. The Ion Peripheral Vision Probe (referred to as the Vision Probe or VP) is an endoscope that is compatible with, and inserted into, the Catheter tool channel. The VP provides direct visualization of the patient s airways during navigation. The VP attaches to the arm and has a collar that is rotated and then pulled to release it from the arm. The VP is provided non-sterile and is a multi-use device that can be used up to 5 times. The Vision Probe Bag (referred to as VPB) installs over the Instrument Arm for temporary storage of the VP when it is not inserted within the Catheter tool channel. The VPB is a single-use device provided sterile (folded within a pouch) by Ethylene Oxide (EO) sterilization method. Refer to HHE for additional details. Component: N/A
Brand Name: Ion Endoluminal System Product Name: Ion Vision Probe Bag Model/Catalog Number: 490127 Software Version: N/A Product Description: The Ion Endoluminal System is a robotically assisted platform for minimally invasive peripheral lung biopsy. Ion consists of a System Cart and Controller that are used to navigate a flexible Catheter within the pulmonary tract using endoscopic visualization for diagnostic and therapeutic procedures. The Ion Fully Articulating Catheter (referred to as the Catheter) consists of the housing, sensor, input discs, shaft, tool channel, and tip. The Catheter is installed on the carriage of the arm. The Catheter is provided non-sterile and is a multi-use device. The Ion Peripheral Vision Probe (referred to as the Vision Probe or VP) is an endoscope that is compatible with, and inserted into, the Catheter tool channel. The VP provides direct visualization of the patient s airways during navigation. The VP attaches to the arm and has a collar that is rotated and then pulled to release it from the arm. The VP is provided non-sterile and is a multi-use device that can be used up to 5 times. The Vision Probe Bag (referred to as VPB) installs over the Instrument Arm for temporary storage of the VP when it is not inserted within the Catheter tool channel. The VPB is a single-use device provided sterile (folded within a pouch) by Ethylene Oxide (EO) sterilization method. Refer to HHE for additional details. Component: N/A
Due to probe bags that may have a compromised or incomplete sterile pouch seal.
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
TS-10/TS-10H Tube Sorter, automatic sorting system for sorting barcoded sample tubes, Catalog Number BN168792, running software versions 00-17
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12 Product Description: The X8 X10 X12 Patient Monitor (hereinafter called X series) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. Component: No
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: X8, X10, X12 Product Description: The X8 X10 X12 Patient Monitor (hereinafter called X series) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time. Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70, iM80 Product Description: The iM series Patient Monitor including iM50, iM60, iM70 and iM80 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate. Component: No
Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70, iM80 Product Description: The iM series Patient Monitor including iM50, iM60, iM70 and iM80 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate. Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2
SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.
Rifton Equipment, Low Base TRAM. Model/Catalog Number: K320. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training. The low base variant includes smaller casters and lower base frame so that the legs fit underneath beds.
Rifton Equipment, Low Base TRAM. Model/Catalog Number: K320. The TRAM is a powered lift device for transferring clients in a seated position or assisting in a sit-to-stand transfer and gait training. The low base variant includes smaller casters and lower base frame so that the legs fit underneath beds.
Potential for fraying on the body support strap.
Thyroid, USP, Rx only, Net Wt: 0.50kg, For Manufacturing, Processing or Repackaging Use Only, Specialty Process Labs, Phoenix, AZ 85034, NDC 81305-100-02
Thyroid, USP, Rx only, Net Wt: 0.50kg, For Manufacturing, Processing or Repackaging Use Only, Specialty Process Labs, Phoenix, AZ 85034, NDC 81305-100-02
Subpotent Drug
GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legend PET Gantry 16cm - MEA, Omni Legend PET Gantry 16cm Haifa LM, Omni Legend PET Gantry 16cm Mobile, Omni Legend PET Gantry 21cm - BJ for Global, Omni Legend PET Gantry 21cm - Global Haifa LM, Omni Legend PET Gantry 21cm - Mobile, Omni Legend PET Gantry 21cm WSO LM, Omni Legend PET Gantry 32cm - MEA, Omni Legend PET Gantry 32cm Haifa LM, Omni Legend PET Gantry 32cm Mobile, OMNI PET Gantry 15cm, Omni PET Gantry 15cm - WSO, OMNI PET Gantry 15cm Cardiac- WSO, OMNI PET Gantry 15cm ES - WSO, Omni PET Gantry 15cm Mobile WSO, Omni PET Gantry 21cm - WSO, Omni PET Gantry 21cm Mobile - WSO, OMNI PET Gantry 30cm, Omni PET Gantry 30cm - WSO; System, Tomography, Computed, Emission
GE Healthcare Omni Legend, sold under the following names: OMNI Legend 16, Omni Legend 21, OMNI Legend 32, Omni Legend PET Gantry 16cm - MEA, Omni Legend PET Gantry 16cm Haifa LM, Omni Legend PET Gantry 16cm Mobile, Omni Legend PET Gantry 21cm - BJ for Global, Omni Legend PET Gantry 21cm - Global Haifa LM, Omni Legend PET Gantry 21cm - Mobile, Omni Legend PET Gantry 21cm WSO LM, Omni Legend PET Gantry 32cm - MEA, Omni Legend PET Gantry 32cm Haifa LM, Omni Legend PET Gantry 32cm Mobile, OMNI PET Gantry 15cm, Omni PET Gantry 15cm - WSO, OMNI PET Gantry 15cm Cardiac- WSO, OMNI PET Gantry 15cm ES - WSO, Omni PET Gantry 15cm Mobile WSO, Omni PET Gantry 21cm - WSO, Omni PET Gantry 21cm Mobile - WSO, OMNI PET Gantry 30cm, Omni PET Gantry 30cm - WSO; System, Tomography, Computed, Emission
There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most easily identified in transaxial slices on the acquisition console (both corrected and non-corrected for attenuation).
Brand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6 Express, F9 Express Product Description: The devices are bedside fetal and maternal monitors, which are used to monitor the physiological parameters of pregnant women including the fetus from 28 weeks gestation. The devices may be used antepartum as well as during labor and delivery. Component: No
Brand Name: EDAN Product Name: Fetal & Maternal Monitor Model/Catalog Number: F6, F9, F6 Express, F9 Express Product Description: The devices are bedside fetal and maternal monitors, which are used to monitor the physiological parameters of pregnant women including the fetus from 28 weeks gestation. The devices may be used antepartum as well as during labor and delivery. Component: No
As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.