LH One Step Ovulation Test Device (Urine)
LH One Step Ovulation Test Device (Urine)
Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Product Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog Number: 81550 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges (IRC): Lumipulse G pTau 217 Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G pTau 217 Plasma IRC), for the quantitative measurement of pTau 217 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology by a specific two-step immunoassay method on the LUMIPULSE G System. 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-phosphorylated Tau (217) monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-Tau monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains chemical stabilizers in 50 mM Tris buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327
Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Triumph Motorcycles America, Ltd. - Alternator Wiring May Short and Overheat
Triumph Motorcycles America, Ltd. - Alternator Wiring May Short and Overheat
Overheated wiring increases the risk of a crash or fire.
JUICE from the Raw, cashew milk
JUICE from the Raw, cashew milk
Under-processed product may contain pathogenic bacteria that could harm to humans.
Sofit, Seedless Red Dates, 12 oz package, Plastic bags, 50 packages per case
Sofit, Seedless Red Dates, 12 oz package, Plastic bags, 50 packages per case
Undeclared sulfites
Bread and Roll Mix; 50 lb. multi-wall, kraft, poly-lined moisture proof bags
Bread and Roll Mix; 50 lb. multi-wall, kraft, poly-lined moisture proof bags
Undeclared Milk Allergen
spicely ORGANIC garlic salt NO ARTIFICIAL COLORING NO GLUTEN - NO MSG - NO SOY NO PRESERVATIVES - NO WHEAT NO SUGAR - NO SALT - NO IRRADIATION Net wt: 3.4oz (96g) Ingredients: Salt, *Garlic *Organic www.spicely.com Organic Spices Inc, Fremont CA 94538 UPC: 826998007643
spicely ORGANIC garlic salt NO ARTIFICIAL COLORING NO GLUTEN - NO MSG - NO SOY NO PRESERVATIVES - NO WHEAT NO SUGAR - NO SALT - NO IRRADIATION Net wt: 3.4oz (96g) Ingredients: Salt, *Garlic *Organic www.spicely.com Organic Spices Inc, Fremont CA 94538 UPC: 826998007643
Product contains salt and is mislabeled
Spice Cake Mix; 50 lb. multi-wall, kraft, poly-lined moisture proof bags
Spice Cake Mix; 50 lb. multi-wall, kraft, poly-lined moisture proof bags
Undeclared Milk Allergen
la ferme aux biches - conserves de cerf
la ferme aux biches - conserves de cerf
perte de maitrise du process de sterilisation
H3 Hintermann Ankle Replacement System
H3 Hintermann Ankle Replacement System
The TGA has concerns about an increased risk of revision as shown in the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) data when compared to other fixed and mobile b...
Green Lumber Natural Fuel For Men capsule, packaged in 2, 4, 10, and 30-count blister packs. GreenLumber.com, 2618 San Miguel Drive, Suite #296, Newport Beach, CA 92660
Green Lumber Natural Fuel For Men capsule, packaged in 2, 4, 10, and 30-count blister packs. GreenLumber.com, 2618 San Miguel Drive, Suite #296, Newport Beach, CA 92660
Marketed without an Approved NDA/ANDA: FDA analysis found this product to contain tadalafil
General Rubber (Thailand) Co., Ltd - Tires May Crack and Separate/FMVSS 139
General Rubber (Thailand) Co., Ltd - Tires May Crack and Separate/FMVSS 139
Tire failure can cause a sudden loss of tire pressure, increasing the risk of a crash.
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.