Medical DeviceUSFDA Devices

Endo-Model Replacement Plateau; Item Number: 15-8521/15;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model Replacement Plateau; Item Number: 15-8521/15;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Medical DeviceUSFDA Devices

GEM Premier 5000 PAK; Part No. 00055407510.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000 PAK; Part No. 00055407510.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

GEM Premier 5000; Part No. 00055445008.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000; Part No. 00055445008.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

GEM Premier 5000 PAK; Part No. 00055407508.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000 PAK; Part No. 00055407508.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).

Physio-Control, Inc.
FDA DevicesClass II

PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).

Due to pediatric defibrillator electrode delamination

Mar 18, 2026Physio-Control, Inc.
Medical DeviceUSFDA Devices

GEM Premier 5000 PAK; Part No. 00055360011.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000 PAK; Part No. 00055360011.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Food & BeverageUSFDA Food

Made Fresh Salads brand Caramel Apple Cream Cheese; 5 lb white plastic tub

Made Fresh Salads Inc
FDA FoodClass I

Made Fresh Salads brand Caramel Apple Cream Cheese; 5 lb white plastic tub

Products may be contaminated with Listeria monocytogenes.

Mar 18, 2026Made Fresh Salads Inc
Food & BeverageUSFDA Food

VEGETARIAN CHICKEN FLAVOR RAMEN Express Net wt 3oz (85g) Ingredients: Enriched unbleached flour (wheat flour***) Palm oil***" Contains: Wheat, Soy. UPC 8 10037 81004 2 Produced by PALMETTO GOURMET FOODS, INC SALUDA, SC 29138

Palmetto Gourmet Foods Inc.
FDA FoodClass II

VEGETARIAN CHICKEN FLAVOR RAMEN Express Net wt 3oz (85g) Ingredients: Enriched unbleached flour (wheat flour***) Palm oil***" Contains: Wheat, Soy. UPC 8 10037 81004 2 Produced by PALMETTO GOURMET FOODS, INC SALUDA, SC 29138

Product contains undeclared Yellow #5

Mar 18, 2026Palmetto Gourmet Foods Inc.
Medical DeviceUSFDA Devices

GEM Premier 5000 PAK; Part No. 00055407511.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000 PAK; Part No. 00055407511.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

Penner Pacific Bathing Spa, Model Numbers 350010-1, 350000-2, 350010-X

Penner Patient Care, Inc.
FDA DevicesClass III

Penner Pacific Bathing Spa, Model Numbers 350010-1, 350000-2, 350010-X

The device does not bear a unique device identifier.

Mar 18, 2026Penner Patient Care, Inc.
Medical DeviceUSFDA Devices

Penner Pacific Bathing Spa, Model Numbers 860010-1L

Penner Patient Care, Inc.
FDA DevicesClass III

Penner Pacific Bathing Spa, Model Numbers 860010-1L

The device does not bear a unique device identifier.

Mar 18, 2026Penner Patient Care, Inc.
Medical DeviceUSFDA Devices

GEM Premier 5000; Part No: 00055415011.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000; Part No: 00055415011.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Food & BeverageUSFDA Food

VEGETARIAN CHICKEN FLAVOR RAMEN Express Net wt 3oz (85g) Ingredients: Enriched unbleached flour (wheat flour***) Vegetable oil***" Contains: Wheat, Soy. UPC 8 10037 81061 5 Produced by PALMETTO GOURMET FOODS, INC SALUDA, SC 29138

Palmetto Gourmet Foods Inc.
FDA FoodClass II

VEGETARIAN CHICKEN FLAVOR RAMEN Express Net wt 3oz (85g) Ingredients: Enriched unbleached flour (wheat flour***) Vegetable oil***" Contains: Wheat, Soy. UPC 8 10037 81061 5 Produced by PALMETTO GOURMET FOODS, INC SALUDA, SC 29138

Product contains undeclared Yellow #5

Mar 18, 2026Palmetto Gourmet Foods Inc.
Medical DeviceUSFDA Devices

GEM Premier 5000; Part No. 00055430011.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000; Part No. 00055430011.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

Endo-Model Replacement Plateau; Item Number: 15-2836/11;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model Replacement Plateau; Item Number: 15-2836/11;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Medical DeviceUSFDA Devices

CLARITY II Laser System; Model No. 1110200210.

Lutronic Corporation
FDA DevicesClass II

CLARITY II Laser System; Model No. 1110200210.

Reports of devices sparking/popping and potentially burning patients.

Mar 18, 2026Lutronic Corporation
Medical DeviceUSFDA Devices

GEM Premier 5000 PAK; Part No. 00055415004.

Instrumentation Laboratory
FDA DevicesClass II

GEM Premier 5000 PAK; Part No. 00055415004.

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Mar 18, 2026Instrumentation Laboratory
Medical DeviceUSFDA Devices

Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Medical DeviceUSFDA Devices

Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.

Medartis AG
FDA DevicesClass II

Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis of the hand, forearm, shoulder and foot. Component: Not applicable.

Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.

Mar 18, 2026Medartis AG
Medical DeviceUSFDA Devices

Penner Pacific Bathing Spa, Model Numbers 390010-1, 390010-X, 390010-2

Penner Patient Care, Inc.
FDA DevicesClass III

Penner Pacific Bathing Spa, Model Numbers 390010-1, 390010-X, 390010-2

The device does not bear a unique device identifier.

Mar 18, 2026Penner Patient Care, Inc.
Medical DeviceUSFDA Devices

Endo-Model Replacement Plateau; Item Number: 15-2835/12;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model Replacement Plateau; Item Number: 15-2835/12;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Medical DeviceUSFDA Devices

Endo-Model Replacement Plateau; Item Number: 15-8521/11;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model Replacement Plateau; Item Number: 15-8521/11;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Medical DeviceUSFDA Devices

Endo-Model Replacement Plateau; Item Number: 15-0027/16;

Waldemar Link GmbH & Co. KG (Mfg Site)
FDA DevicesClass II

Endo-Model Replacement Plateau; Item Number: 15-0027/16;

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Mar 18, 2026Waldemar Link GmbH & Co. KG (Mfg Site)
Vehicle RecallUSNHTSA

Keystone RV Company - Improperly Secured Wiring May Be Damaged

Keystone RV Company
NHTSAGeneral Alert

Keystone RV Company - Improperly Secured Wiring May Be Damaged

Damaged wiring may cause an electrical fault, increasing the risk of a fire.

Mar 18, 2026Keystone RV Company