Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483C; 4) CV DRAPING/BASIN 2PART PACK, Medline Kit Number/SKU DYNJ89414; 5) CVG BASIN PACK SC, Medline Kit Number/SKU DYNJ38052G; 6) T/A BASIN SET, Medline Kit Number/SKU DYNJ47941F; 7) VALVE BASIN PACK SC, Medline Kit Number/SKU DYNJ38054G.

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as: 1) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483A; 2) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483B; 3) CV DRAPING/BASIN, Medline Kit Number/SKU DYNJ908483C; 4) CV DRAPING/BASIN 2PART PACK, Medline Kit Number/SKU DYNJ89414; 5) CVG BASIN PACK SC, Medline Kit Number/SKU DYNJ38052G; 6) T/A BASIN SET, Medline Kit Number/SKU DYNJ47941F; 7) VALVE BASIN PACK SC, Medline Kit Number/SKU DYNJ38054G.

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline convenience kits: ABD PACK DYNJ81673B BARIATRIC DYNJ902515K BARIATRIC + DYNJ904119O BARIATRIC CDS CDS983883J BWNBORO LAP CHOLE PACK-LF DYNJ32836B CCSB KARIN JWIRE LUMPECTOMY DYNJ910459B CHRISTUS CHILDRENS LAP CHOLE DYNJ61165G CUH MINOR PACK-LF DYNJ60095B FREE FLAP CUSTOM PACK B DYNJ82736B FREE NIPPLE GRAFT DYNJ911160B GEN MINOR DYNJ900906L GEN SURG LAPAROSCOPY PACK DYNJ55248F GEN SURG MINOR PACK DYNJ69142F GENERAL LAP DYNJ61475A DYNJT3759 GENERAL LAP PACK DYNJ61515I DYNJT4452 GENERAL LAP PACK-LF DYNJ37208D GENERAL LAPAROSCOPY DYNJ905870G GENERAL LAPAROSCOPY PACK DYNJ17443T DYNJ45263B DYNJT4504 LAP CHOLE DYNJ45664D DYNJ912005 LAP CHOLE PACK DYNJ81695B LAP CHOLE PACK-LF DYNJ0614113C DYNJ0618369V LAP CHOLE PROCEDURE-LF CDS981184D LAP CHOLE RFD DYNJ906302F LAP CHOLE SURGICOUNT DYNJ45531G LAP CHOLE TOTE DYNJ908339F LAP GASTRIC BYPASS CDS CDS860146M LAP GASTRIC PACK DYNJ62065A LAP HERNIA RFD DYNJ906301F LAP PACK DYNJ38871J DYNJ51389A LAPAROSCOPIC CDS920082AC LAPAROSCOPIC PACK DYNJ52605K LAPAROSCOPIC PACK-LF PHS41736G LAPAROSCOPY PACK DYNJT8313 LAPAROSCOPY PACK-LF DYNJ02309D DYNJ0551175T LAPAROSCOPY UM RF DYNJ902324D LAPAROSCOPY-LF DYNJ903241B LAPAROTOMY PACK-LF DYNJ21091O LARGE ABDOMINAL PACK-LF PHS41737D LEX GENERAL LAPAROSCOPY DYNJ900969K MINOR GENERAL PACK DYNJT3884 MINOR PACK-LF DYNJ0105608I MINOR PK DYNJ61335D MINOR PROCEDURE CDS CDS981793G PEDIATRIC MOSES DYNJ58864B PK, GEN-LAPAROSCOPY DYNJ53837B RMC GENERAL LAPAROSCOPIC DYNJ907087C ROBOTIC GENERAL DYNJ906861D SM-GENERAL LAP PACK-LF DYNJ0780078W SMJ GENERAL ABDOMINAL PACK-LF DYNJ25765Q UOFL HOSP GENERAL LAPAROSCOPY DYNJ909321A

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits: ABD PACK DYNJ81673B BARIATRIC DYNJ902515K BARIATRIC + DYNJ904119O BARIATRIC CDS CDS983883J BWNBORO LAP CHOLE PACK-LF DYNJ32836B CCSB KARIN JWIRE LUMPECTOMY DYNJ910459B CHRISTUS CHILDRENS LAP CHOLE DYNJ61165G CUH MINOR PACK-LF DYNJ60095B FREE FLAP CUSTOM PACK B DYNJ82736B FREE NIPPLE GRAFT DYNJ911160B GEN MINOR DYNJ900906L GEN SURG LAPAROSCOPY PACK DYNJ55248F GEN SURG MINOR PACK DYNJ69142F GENERAL LAP DYNJ61475A DYNJT3759 GENERAL LAP PACK DYNJ61515I DYNJT4452 GENERAL LAP PACK-LF DYNJ37208D GENERAL LAPAROSCOPY DYNJ905870G GENERAL LAPAROSCOPY PACK DYNJ17443T DYNJ45263B DYNJT4504 LAP CHOLE DYNJ45664D DYNJ912005 LAP CHOLE PACK DYNJ81695B LAP CHOLE PACK-LF DYNJ0614113C DYNJ0618369V LAP CHOLE PROCEDURE-LF CDS981184D LAP CHOLE RFD DYNJ906302F LAP CHOLE SURGICOUNT DYNJ45531G LAP CHOLE TOTE DYNJ908339F LAP GASTRIC BYPASS CDS CDS860146M LAP GASTRIC PACK DYNJ62065A LAP HERNIA RFD DYNJ906301F LAP PACK DYNJ38871J DYNJ51389A LAPAROSCOPIC CDS920082AC LAPAROSCOPIC PACK DYNJ52605K LAPAROSCOPIC PACK-LF PHS41736G LAPAROSCOPY PACK DYNJT8313 LAPAROSCOPY PACK-LF DYNJ02309D DYNJ0551175T LAPAROSCOPY UM RF DYNJ902324D LAPAROSCOPY-LF DYNJ903241B LAPAROTOMY PACK-LF DYNJ21091O LARGE ABDOMINAL PACK-LF PHS41737D LEX GENERAL LAPAROSCOPY DYNJ900969K MINOR GENERAL PACK DYNJT3884 MINOR PACK-LF DYNJ0105608I MINOR PK DYNJ61335D MINOR PROCEDURE CDS CDS981793G PEDIATRIC MOSES DYNJ58864B PK, GEN-LAPAROSCOPY DYNJ53837B RMC GENERAL LAPAROSCOPIC DYNJ907087C ROBOTIC GENERAL DYNJ906861D SM-GENERAL LAP PACK-LF DYNJ0780078W SMJ GENERAL ABDOMINAL PACK-LF DYNJ25765Q UOFL HOSP GENERAL LAPAROSCOPY DYNJ909321A

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

ConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS5-120LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

ConMed Corporation
FDA DevicesClass II

ConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS5-120LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Potential for overpressure alerts.

Jun 24, 2026ConMed Corporation
Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as BIOBURDEN TEST KIT, Kit Uro Robot Cystectomy, etc. (see the recall documents for a full list of products)

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as BIOBURDEN TEST KIT, Kit Uro Robot Cystectomy, etc. (see the recall documents for a full list of products)

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.

BD SWITZERLAND SARL
FDA DevicesClass II

BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.

Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.

Jun 24, 2026BD SWITZERLAND SARL
Medical DeviceUSFDA Devices

Medline convenience kits: ADD-A-CATH ARTERIAL INSERTION ARTNC185 PED SCHOOL AGE PICC DRSG TRAY DYNDC3291

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits: ADD-A-CATH ARTERIAL INSERTION ARTNC185 PED SCHOOL AGE PICC DRSG TRAY DYNDC3291

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

MEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ908097B; 4) CRANI, Medline Kit Number/SKU DYNJ908097C; 5) CRANI, Medline Kit Number/SKU DYNJ908097D; 6) DR TUCKER ANT TOTAL HIP, Medline Kit Number/SKU DYNJ912029; 7) JEWISH CRANI PACK, Medline Kit Number/SKU DYNJ55637B; 8) L SCOPE PACK, Medline Kit Number/SKU DYNJ43930A; 9) LAMI PACK, Medline Kit Number/SKU DYNJ67767; 10) LAVH NEW BRAUNFELS, Medline Kit Number/SKU DYNJ61136F; 11) SOI KNEE ARTHROSCOPY, Medline Kit Number/SKU DYNJV0172Q; 12) T AND A PACK, Medline Kit Number/SKU DYNJ86229.

Medline Industries, LP
FDA DevicesClass II

MEDLINE Medical Procedure Kits labeled as: 1) CRANI, Medline Kit Number/SKU DYNJ908097; 2) CRANI, Medline Kit Number/SKU DYNJ908097A; 3) CRANI, Medline Kit Number/SKU DYNJ908097B; 4) CRANI, Medline Kit Number/SKU DYNJ908097C; 5) CRANI, Medline Kit Number/SKU DYNJ908097D; 6) DR TUCKER ANT TOTAL HIP, Medline Kit Number/SKU DYNJ912029; 7) JEWISH CRANI PACK, Medline Kit Number/SKU DYNJ55637B; 8) L SCOPE PACK, Medline Kit Number/SKU DYNJ43930A; 9) LAMI PACK, Medline Kit Number/SKU DYNJ67767; 10) LAVH NEW BRAUNFELS, Medline Kit Number/SKU DYNJ61136F; 11) SOI KNEE ARTHROSCOPY, Medline Kit Number/SKU DYNJV0172Q; 12) T AND A PACK, Medline Kit Number/SKU DYNJ86229.

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-UPM; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-HC; ARTERIAL LINE KIT: 20 GA X 5IN (12 CM)/ASK-04020-HMC; RADIAL ARTERY CATHETER KIT: 20 GA X 30MM/ASK-04020-MMC; RA CATH KIT: 20 GA X 1-1/2IN (3.81 CM)/ASK-04020-SV; ARTERIAL CATH KIT: 20 GA X 5 IN (12 CM)/ASK-04100-AMC-S1; ARTERIAL CATHETER KIT: 20 GA X 12 CM/ASK-04100-NS; RADIAL ARTERY KIT/ASK-04500-AH; RAD ART:20GAX3.81CM / ART LINE:20GAX12CM/ASK-04500-HF-S; ARTERIAL LINE KIT: 1L 20 GA X 12 CM/ASK-04510-HA; ARTERIAL LINE KIT: 20 GA. X 5 IN (12 CM)/ASK-04510-HF; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-IHC1; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-MSK2; ARTERIAL CATH KIT: 20GA X 5IN/ASK-04510-NS; ARTERIAL CATHETER KIT: 20 GA X 12 CM/FS-04510-S; ARTERIAL LINE KIT/HF-04510-1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-BSM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45CM)/ASK-04020-HH1; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-JMH1; RA CATH KIT: 20 GA X 1-3/4 IN/ASK-04020-MAS; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-MCC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MHT; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MI; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MIHS; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MUSC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-NCB; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-SPH; RA CATH SET: 20 GA X 1-3/4IN (4.45CM)/ASK-04020-UPM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-UR1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-X1A; RA CATH KIT: 20 GA X 1-3/4IN/WBH-04020-1; CVC KIT: 20 GA X 12CM/ASK-04510-UCLA; ARTERIAL ACCESS TRAY WITH .025IN SWG/ASK-04001-BW1

ARROW INTERNATIONAL, LLC
FDA DevicesClass I

Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-UPM; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-HC; ARTERIAL LINE KIT: 20 GA X 5IN (12 CM)/ASK-04020-HMC; RADIAL ARTERY CATHETER KIT: 20 GA X 30MM/ASK-04020-MMC; RA CATH KIT: 20 GA X 1-1/2IN (3.81 CM)/ASK-04020-SV; ARTERIAL CATH KIT: 20 GA X 5 IN (12 CM)/ASK-04100-AMC-S1; ARTERIAL CATHETER KIT: 20 GA X 12 CM/ASK-04100-NS; RADIAL ARTERY KIT/ASK-04500-AH; RAD ART:20GAX3.81CM / ART LINE:20GAX12CM/ASK-04500-HF-S; ARTERIAL LINE KIT: 1L 20 GA X 12 CM/ASK-04510-HA; ARTERIAL LINE KIT: 20 GA. X 5 IN (12 CM)/ASK-04510-HF; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-IHC1; ARTERIAL LINE KIT: 20 GA X 5 IN (12 CM)/ASK-04510-MSK2; ARTERIAL CATH KIT: 20GA X 5IN/ASK-04510-NS; ARTERIAL CATHETER KIT: 20 GA X 12 CM/FS-04510-S; ARTERIAL LINE KIT/HF-04510-1; RA CATH KIT: 20 GA X 1-3/4IN/ASK-04020-BSM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45CM)/ASK-04020-HH1; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-JMH1; RA CATH KIT: 20 GA X 1-3/4 IN/ASK-04020-MAS; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-MCC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MHT; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MI; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-MIHS; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-MUSC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-NCB; RA CATH KIT: 20 GA X 1 3/4 IN/ASK-04020-PMC; RA CATH KIT: 20 GA X 1-3/4IN (4.45 CM)/ASK-04020-SPH; RA CATH SET: 20 GA X 1-3/4IN (4.45CM)/ASK-04020-UPM; RA CATH KIT: 20 GA X 1 3/4 IN (4.45 CM)/ASK-04020-UR1; RA CATH KIT: 20 GA X 4.45 CM/NA-04020-X1A; RA CATH KIT: 20 GA X 1-3/4IN/WBH-04020-1; CVC KIT: 20 GA X 12CM/ASK-04510-UCLA; ARTERIAL ACCESS TRAY WITH .025IN SWG/ASK-04001-BW1

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Jun 24, 2026ARROW INTERNATIONAL, LLC
Medical DeviceUSFDA Devices

ConMed AirSeal 12mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS12-120LPI. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

ConMed Corporation
FDA DevicesClass II

ConMed AirSeal 12mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS12-120LPI. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Potential for overpressure alerts.

Jun 24, 2026ConMed Corporation
Medical DeviceUSFDA Devices

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Smith & Nephew, Inc.
FDA DevicesClass II

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

Jun 24, 2026Smith & Nephew, Inc.
Medical DeviceUSFDA Devices

ALYON Surgical Lighting System

Steris Corporation
FDA DevicesClass II

ALYON Surgical Lighting System

The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

Jun 24, 2026Steris Corporation
Medical DeviceUSFDA Devices

Description/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP, DRAINAGE KIT/ASK-01000-NW, THORACENTESIS KIT/ASK-01000-UPMS, PNEUMOTHORAX KIT/AK-01500, PNEUMOTHORAX KIT (NO HEIMLICH VALVE)/ASK-01500, PNEUMOTHORAX KIT/ASK-01500-JCM

ARROW INTERNATIONAL, LLC
FDA DevicesClass I

Description/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP, DRAINAGE KIT/ASK-01000-NW, THORACENTESIS KIT/ASK-01000-UPMS, PNEUMOTHORAX KIT/AK-01500, PNEUMOTHORAX KIT (NO HEIMLICH VALVE)/ASK-01500, PNEUMOTHORAX KIT/ASK-01500-JCM

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Jun 24, 2026ARROW INTERNATIONAL, LLC
Medical DeviceUSFDA Devices

ConMed AirSeal 8mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. Catalog Number: iAS8-100LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

ConMed Corporation
FDA DevicesClass II

ConMed AirSeal 8mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 100mm Length. Catalog Number: iAS8-100LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Potential for overpressure alerts.

Jun 24, 2026ConMed Corporation
Medical DeviceUSFDA Devices

Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C

AMD Medicom Inc.
FDA DevicesClass II

Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C; 0.75x0.75in, Non-Sterile/55995C; 1.0x1.0in, Non-Sterile/55996C; 1.0x3.0in, Non-Sterile/55997C; 0.5x1.5in, Non-Sterile/55998C

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

Jun 24, 2026AMD Medicom Inc.
Medical DeviceUSFDA Devices

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Datex-Ohmeda Inc
FDA DevicesClass I

Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Jun 24, 2026Datex-Ohmeda Inc
Medical DeviceUSFDA Devices

Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500

GE Medical Systems, LLC
FDA DevicesClass II

Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Jun 24, 2026GE Medical Systems, LLC
Medical DeviceUSFDA Devices

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C

Medline Industries, LP
FDA DevicesClass II

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F

Medline Industries, LP
FDA DevicesClass II

Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Jun 24, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Description/REF: ACCESS TRAY/ASK-04001-MC3

ARROW INTERNATIONAL, LLC
FDA DevicesClass I

Description/REF: ACCESS TRAY/ASK-04001-MC3

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Jun 24, 2026ARROW INTERNATIONAL, LLC
Medical DeviceUSFDA Devices

Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A

Inspire Medical Systems Inc.
FDA DevicesClass II

Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A

Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.

Jun 24, 2026Inspire Medical Systems Inc.