États-Unis Recalls
Rappels de sécurité de la FDA, CPSC, NHTSA et de l’USDA FSIS.
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791
Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791
Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.
Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper FD10/10; Model Numbers: 722005, 722011, 722027. 4. Allura Xper FD10C; Model Number: 722001. 5. Allura Xper FD20; Model Numbers: 722006, 722012, 722028. 6. Allura Xper FD20 Biplane; Model Numbers: 722008, 722013. 7. Allura Xper FD20 Biplane OR Table; Model Number: 722025. 8. Allura Xper FD20 OR Table; Model Numbers: 722015, 722023, 722035. 9. Allura Xper FD20/10; Model Number: 722029. 10. Allura Xper FD20/15; Model Number: 722058. 11. Allura Xper FD20/15 OR Table; Model Number: 722059. 12. Allura Xper FD20/20; Model Number: 722038. 13. Allura Xper FD20/20 OR Table; Model Number: 722039. 14. Allura Xper R9 7 M12; Model Number: 722065. 15. Allura Xper R9 7 M20; Model Number: 722066. Interventional Fluoroscopic X-Ray System.
Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper FD10/10; Model Numbers: 722005, 722011, 722027. 4. Allura Xper FD10C; Model Number: 722001. 5. Allura Xper FD20; Model Numbers: 722006, 722012, 722028. 6. Allura Xper FD20 Biplane; Model Numbers: 722008, 722013. 7. Allura Xper FD20 Biplane OR Table; Model Number: 722025. 8. Allura Xper FD20 OR Table; Model Numbers: 722015, 722023, 722035. 9. Allura Xper FD20/10; Model Number: 722029. 10. Allura Xper FD20/15; Model Number: 722058. 11. Allura Xper FD20/15 OR Table; Model Number: 722059. 12. Allura Xper FD20/20; Model Number: 722038. 13. Allura Xper FD20/20 OR Table; Model Number: 722039. 14. Allura Xper R9 7 M12; Model Number: 722065. 15. Allura Xper R9 7 M20; Model Number: 722066. Interventional Fluoroscopic X-Ray System.
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
STRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620
STRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620
There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).
FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).
Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
Pumpkin Tree Peter Rabbit Organics Banana & Strawberry fruit puree, 4oz/113g pouch, UPC 815367010780, PTO Item 10720. Distributed by: PT Organics Ltd, 333 South State St, Suite V-142, Lake Oswego, OR 97034. Product is pasteurized blend of banana and strawberry purees sold in single serve spouted flexible film pouch (4oz/113g) with red cap. It is intended for consumption by children aged +6 months. Each master case contains 10/4oz. pouches of PT Organics Banana Strawberry 1006034. The bar code on master case label is 00815367010797.
Pumpkin Tree Peter Rabbit Organics Banana & Strawberry fruit puree, 4oz/113g pouch, UPC 815367010780, PTO Item 10720. Distributed by: PT Organics Ltd, 333 South State St, Suite V-142, Lake Oswego, OR 97034. Product is pasteurized blend of banana and strawberry purees sold in single serve spouted flexible film pouch (4oz/113g) with red cap. It is intended for consumption by children aged +6 months. Each master case contains 10/4oz. pouches of PT Organics Banana Strawberry 1006034. The bar code on master case label is 00815367010797.
Possible foreign object (thin, soft plastic strands).
Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A
Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A
Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.
Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.
Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).
Clover Hill Dairy LLP, Soft Cuajada in Brine, sold in five (5) gallon and two (2) gallon buckets. Unknown information: labeling. Keep refrigerated.
Clover Hill Dairy LLP, Soft Cuajada in Brine, sold in five (5) gallon and two (2) gallon buckets. Unknown information: labeling. Keep refrigerated.
Product contaminated with Listeria monocytogenes.
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Clover Hill Dairy LLP, Pepper Jack Cheese Varieties (Snack Pack Assorted Pepperjack Stix (Jalapeno Cheddar, Pepper Jack)). Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Clover Hill Dairy LLP, Pepper Jack Cheese Varieties (Snack Pack Assorted Pepperjack Stix (Jalapeno Cheddar, Pepper Jack)). Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Product contaminated with Listeria monocytogenes.
Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Clover Hill Dairy LLP, mild cheese variety (Snack Pack Assorted Regular Mild Stix White Colby, Marble Jack, Yummy Cheddar). Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Clover Hill Dairy LLP, mild cheese variety (Snack Pack Assorted Regular Mild Stix White Colby, Marble Jack, Yummy Cheddar). Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Product contaminated with Listeria monocytogenes.
Clover Hill Dairy LLP, Sharp Cheese Snack Pack Sharp Stix White or Yellow. Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Clover Hill Dairy LLP, Sharp Cheese Snack Pack Sharp Stix White or Yellow. Sold in approximate 0.40-lb (6.4oz) units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Product contaminated with Listeria monocytogenes.
Clover Hill Dairy LLP, Mild Cheese Variety (Yummy Cheddar, White Cheddar, White Colby, Monterey Jack, Marble Jack). Sold in 1/2-lb, 1-lb, 2.5-lb, 5-lb, and 40-lb units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Clover Hill Dairy LLP, Mild Cheese Variety (Yummy Cheddar, White Cheddar, White Colby, Monterey Jack, Marble Jack). Sold in 1/2-lb, 1-lb, 2.5-lb, 5-lb, and 40-lb units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.
Product contaminated with Listeria monocytogenes.
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475635 Product Description: Intramedullary Bone Saw Blade Assembly 15mm
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475635 Product Description: Intramedullary Bone Saw Blade Assembly 15mm
Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.