États-Unis Recalls
Rappels de sécurité de la FDA, CPSC, NHTSA et de l’USDA FSIS.
Allura Xper FD10/10; Model Number: 722027;
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Benadryl, 25'S 2PK Box, UPC 815556020316
Benadryl, 25'S 2PK Box, UPC 815556020316
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
Volkswagen Group of America, Inc. - Park Outside - High Voltage Battery May Catch Fire
Volkswagen Group of America, Inc. - Park Outside - High Voltage Battery May Catch Fire
A vehicle fire while parked or driving can increase the risk of injury.
Ford Motor Company - Inadequate Windshield Bonding/FMVSS 212
Ford Motor Company - Inadequate Windshield Bonding/FMVSS 212
In the event of a crash, windshield detachment can increase the risk of injury.
Keystone RV Company - Ramp Door May Break During Use
Keystone RV Company - Ramp Door May Break During Use
A ramp door that breaks during use increases the risk of injury.
Allura Xper FD20 OR Table; Model Number: 722023;
Allura Xper FD20 OR Table; Model Number: 722023;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Oshkosh Corporation - Wheel May Detach from Vehicle
Oshkosh Corporation - Wheel May Detach from Vehicle
A detached wheel can cause a loss of vehicle control or become a road hazard for other vehicles, increasing the risk of a crash.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-PLASTIC CRANIOFACIAL PK-LF, Medline kit SKU DYNJ0395085V; 2) NH PERI DISC PACK, Medline kit SKU DYNJ66311C; 3) DISCECTOMY PACK, Medline kit SKU DYNJ86770; 4) TOTAL KNEE, Medline kit SKU DYNJ909643F.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-PLASTIC CRANIOFACIAL PK-LF, Medline kit SKU DYNJ0395085V; 2) NH PERI DISC PACK, Medline kit SKU DYNJ66311C; 3) DISCECTOMY PACK, Medline kit SKU DYNJ86770; 4) TOTAL KNEE, Medline kit SKU DYNJ909643F.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Plum Duo Infusion System, List Number 40002-04-01
Plum Duo Infusion System, List Number 40002-04-01
ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.
Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU DAVINCI, REF CDS982779N; 9) GYN DAVINCI, REF CDS982780V; 10) UNIVERSAL ROBOTIC ORLANDO, REF CDS982857K; 11) ROBOTIC GU/GYN, REF CDS983393L; 12) GYN LAPAROSCOPY CDS, REF CDS984656Q; 13) GYN LAPAROSCOPY CDS, REF CDS984656R; 14) DAVINCI CDS UCC-LF, REF CDS984663J; 15) ROBOTIC UROLOGY PROSTATE, REF CDS985467J; 16) DAVINCI PROSTATECTOMY, REF DYNJG901002B; 17) ROBOTIC PACK IMF 100913- LF, REF DYNJVB91075C; 18) ROBOTIC PACK, REF DYNJ36840L; 19) ROBOTIC PACK MNMC, REF DYNJ38152D; 20) WMC GYN LAPAROSCOPY PACK-LF, REF DYNJ50649J; 21) ROBOTIC GASTRIC BYPASS PACK, REF DYNJ52096P; 22) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564D; 23) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564F; 24) GYN ROBOTIC PACK, REF DYNJ52581D; 25) GYN LAPAROSCOPY PACK, REF DYNJ53782J; 26) UNIVERSAL DAVINCI ROBOTIC P, REF DYNJ53968P; 27) WALKER GYN-UROLOGY ROBOTIC PK, REF DYNJ57458B; 28) GYN LAPAROSCOPY, REF DYNJ62136B; 29) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392F; 30) TOTAL LAPAROSCOPIC HYST PACK, REF DYNJ63456B; 31) ROBOTIC PACK, REF DYNJ67177D; 32) MH GYN LAPAROSCOPY PACK-LF, REF DYNJ84433B; 33) DAVINCI, REF DYNJ900963M; 34) LEX DAVINCI GYN MAJOR, REF DYNJ900965N; 35) LEX DAVINCI UROLOGY, REF DYNJ900967N; 36) COR DAVINCI, REF DYNJ901018O; 37) GYN ROBOTIC, REF DYNJ901046M; 38) DAVINCI, REF DYNJ901076L; 39) GYN ROBOTIC, REF DYNJ901796K; 40) LOU GENERAL ROBOT, REF DYNJ901844N; 41) LOU GYN LAPAROSCOPY, REF DYNJ901845L; 42) LOU GYN LAPAROSCOPY, REF DYNJ901845M; 43) LOU LITHOTOMY ROBOTIC, REF DYNJ901850N; 44) LEX GYN MINOR LAPAROSCOPY, REF DYNJ902029G; 45) GYN ROBOT, REF DYNJ902513J; 46) CHN OR DAVINCI PROSTATE, REF DYNJ902540O; 47) WMC ROBOTIC PACK - LF, REF DYNJ902583N; 48) ROBOTIC GYN & GU, REF DYNJ903459J; 49) GYN LAPAROSCOPY-LF, REF DYNJ903647I; 50) ROBOTIC UROLOGY, REF DYNJ903827F; 51) LITHOTOMY ROBOTIC CDS, REF DYNJ903948R; 52) GYN ROBOTIC, REF DYNJ903983P; 53) GYN LAP, REF DYNJ903990Q; 54) ROBOTIC-LF, REF DYNJ904254L; 55) ROBOTIC, REF DYNJ904343G; 56) ROBOT GYN, REF DYNJ904572L; 57) ROBOTIC PROSTECTOMY, REF DYNJ904597G; 58) GYN LAPAROSCOPY-LF, REF DYNJ905289G; 59) GYN ROBOTIC, REF DYNJ905730J; 60) URO ROBOTIC, REF DYNJ905746J; 61) COLON KIT, REF DYNJ905964G; 62) CHN OR ROBOTIC GYN, REF DYNJ905985O; 63) CW ROBOT, REF DYNJ906051C; 64) CW ROBOT, REF DYNJ906051D; 65) ROBOT, REF DYNJ906112B; 66) CDS ROBOTIC GYN CH, REF DYNJ906265D; 67) ROBOTIC GYN PACK, REF DYNJ906265F; 68) KIT GYN LAPAROSCOPY RFD, REF DYNJ906322F; 69) GYN LAPAROSCOPY, REF DYNJ906357C; 70) ROBOTIC COLON, REF DYNJ906379D; 71) GYN DAVINCI FOAKS, REF DYNJ906444C; 72) DAVINCI, REF DYNJ906543B; 73) CHN GYN ONCOLOGY, REF DYNJ906731K; 74) CHN ROBOTIC CYSTECTOMY, REF DYNJ907072F; 75) WMC THORACIC ROBOTIC, REF DYNJ907153C; 76) WMC THORACIC ROBOTIC, REF DYNJ907153D; 77) GYN ROBOTIC, REF DYNJ907492F; 78) KIT LAP GYN NTX, REF DYNJ907599C; 79) KIT ROBOTIC GYN, REF DYNJ907635D; 80) KIT ROBOTIC GYN, REF DYNJ907647B; 81) ROBOTIC-LF, REF DYNJ907910F; 82) ROBOTIC GENERAL, REF DYNJ907919C; 83) ROBOTIC BASE, REF DYNJ908447D; 84) ROBOTIC BASE, REF DYNJ908447F; 85) GYN, REF DYNJ908504A; 86) GU DAVINCI KIT, REF DYNJ908783F; 87) ROBOTIC PROSTATE, REF DYNJ909120D; 88) OSC ROBOT BARIATRIC, REF DYNJ909547; 89) LEX DAVINCI PROSTATE, REF DYNJ909548A; 90) GYN-URO ROBOT, REF DYNJ910137A; 91) PROSTATECTOMY ROBOTIC, REF DYNJ910991; 92) GYN ROBOTIC, REF DYNJ910995; 93) ROBOTIC, REF DYNJ9225761X; 94) DAVINCI ROBOTIC PACK-LF, REF DYNJ9288198X; 95) MB GYN ROBOTIC ADD, REF DYNJ9701096M; 96) PHS722500FCK, REF PHS722500G.
Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU DAVINCI, REF CDS982779N; 9) GYN DAVINCI, REF CDS982780V; 10) UNIVERSAL ROBOTIC ORLANDO, REF CDS982857K; 11) ROBOTIC GU/GYN, REF CDS983393L; 12) GYN LAPAROSCOPY CDS, REF CDS984656Q; 13) GYN LAPAROSCOPY CDS, REF CDS984656R; 14) DAVINCI CDS UCC-LF, REF CDS984663J; 15) ROBOTIC UROLOGY PROSTATE, REF CDS985467J; 16) DAVINCI PROSTATECTOMY, REF DYNJG901002B; 17) ROBOTIC PACK IMF 100913- LF, REF DYNJVB91075C; 18) ROBOTIC PACK, REF DYNJ36840L; 19) ROBOTIC PACK MNMC, REF DYNJ38152D; 20) WMC GYN LAPAROSCOPY PACK-LF, REF DYNJ50649J; 21) ROBOTIC GASTRIC BYPASS PACK, REF DYNJ52096P; 22) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564D; 23) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564F; 24) GYN ROBOTIC PACK, REF DYNJ52581D; 25) GYN LAPAROSCOPY PACK, REF DYNJ53782J; 26) UNIVERSAL DAVINCI ROBOTIC P, REF DYNJ53968P; 27) WALKER GYN-UROLOGY ROBOTIC PK, REF DYNJ57458B; 28) GYN LAPAROSCOPY, REF DYNJ62136B; 29) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392F; 30) TOTAL LAPAROSCOPIC HYST PACK, REF DYNJ63456B; 31) ROBOTIC PACK, REF DYNJ67177D; 32) MH GYN LAPAROSCOPY PACK-LF, REF DYNJ84433B; 33) DAVINCI, REF DYNJ900963M; 34) LEX DAVINCI GYN MAJOR, REF DYNJ900965N; 35) LEX DAVINCI UROLOGY, REF DYNJ900967N; 36) COR DAVINCI, REF DYNJ901018O; 37) GYN ROBOTIC, REF DYNJ901046M; 38) DAVINCI, REF DYNJ901076L; 39) GYN ROBOTIC, REF DYNJ901796K; 40) LOU GENERAL ROBOT, REF DYNJ901844N; 41) LOU GYN LAPAROSCOPY, REF DYNJ901845L; 42) LOU GYN LAPAROSCOPY, REF DYNJ901845M; 43) LOU LITHOTOMY ROBOTIC, REF DYNJ901850N; 44) LEX GYN MINOR LAPAROSCOPY, REF DYNJ902029G; 45) GYN ROBOT, REF DYNJ902513J; 46) CHN OR DAVINCI PROSTATE, REF DYNJ902540O; 47) WMC ROBOTIC PACK - LF, REF DYNJ902583N; 48) ROBOTIC GYN & GU, REF DYNJ903459J; 49) GYN LAPAROSCOPY-LF, REF DYNJ903647I; 50) ROBOTIC UROLOGY, REF DYNJ903827F; 51) LITHOTOMY ROBOTIC CDS, REF DYNJ903948R; 52) GYN ROBOTIC, REF DYNJ903983P; 53) GYN LAP, REF DYNJ903990Q; 54) ROBOTIC-LF, REF DYNJ904254L; 55) ROBOTIC, REF DYNJ904343G; 56) ROBOT GYN, REF DYNJ904572L; 57) ROBOTIC PROSTECTOMY, REF DYNJ904597G; 58) GYN LAPAROSCOPY-LF, REF DYNJ905289G; 59) GYN ROBOTIC, REF DYNJ905730J; 60) URO ROBOTIC, REF DYNJ905746J; 61) COLON KIT, REF DYNJ905964G; 62) CHN OR ROBOTIC GYN, REF DYNJ905985O; 63) CW ROBOT, REF DYNJ906051C; 64) CW ROBOT, REF DYNJ906051D; 65) ROBOT, REF DYNJ906112B; 66) CDS ROBOTIC GYN CH, REF DYNJ906265D; 67) ROBOTIC GYN PACK, REF DYNJ906265F; 68) KIT GYN LAPAROSCOPY RFD, REF DYNJ906322F; 69) GYN LAPAROSCOPY, REF DYNJ906357C; 70) ROBOTIC COLON, REF DYNJ906379D; 71) GYN DAVINCI FOAKS, REF DYNJ906444C; 72) DAVINCI, REF DYNJ906543B; 73) CHN GYN ONCOLOGY, REF DYNJ906731K; 74) CHN ROBOTIC CYSTECTOMY, REF DYNJ907072F; 75) WMC THORACIC ROBOTIC, REF DYNJ907153C; 76) WMC THORACIC ROBOTIC, REF DYNJ907153D; 77) GYN ROBOTIC, REF DYNJ907492F; 78) KIT LAP GYN NTX, REF DYNJ907599C; 79) KIT ROBOTIC GYN, REF DYNJ907635D; 80) KIT ROBOTIC GYN, REF DYNJ907647B; 81) ROBOTIC-LF, REF DYNJ907910F; 82) ROBOTIC GENERAL, REF DYNJ907919C; 83) ROBOTIC BASE, REF DYNJ908447D; 84) ROBOTIC BASE, REF DYNJ908447F; 85) GYN, REF DYNJ908504A; 86) GU DAVINCI KIT, REF DYNJ908783F; 87) ROBOTIC PROSTATE, REF DYNJ909120D; 88) OSC ROBOT BARIATRIC, REF DYNJ909547; 89) LEX DAVINCI PROSTATE, REF DYNJ909548A; 90) GYN-URO ROBOT, REF DYNJ910137A; 91) PROSTATECTOMY ROBOTIC, REF DYNJ910991; 92) GYN ROBOTIC, REF DYNJ910995; 93) ROBOTIC, REF DYNJ9225761X; 94) DAVINCI ROBOTIC PACK-LF, REF DYNJ9288198X; 95) MB GYN ROBOTIC ADD, REF DYNJ9701096M; 96) PHS722500FCK, REF PHS722500G.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Atwater's, Holiday Cookie Tin, 19.65 ounces, Round metal tin with a window lid
Atwater's, Holiday Cookie Tin, 19.65 ounces, Round metal tin with a window lid
Undeclared tree nut allergens: almonds, pecans, and walnuts.
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
So Delicious Dairy Free Salted Caramel Cluster Cashewmilk, Non-Dairy Frozen Dessert, One Pint (473 mL). UPC 744473 476138. Manufactured By/For So Delicious Dairy Free Eugene, OR 97402 So Delicious Dairy Free Salted Caramel Cluster Cashewmilk, Non-Dairy Frozen Dessert. Net wt. 500mL. UPC 744473 276134. Product of the U.S.A. Distributed by Danone Inc., Boucherville, Quebec J4B 1E6
So Delicious Dairy Free Salted Caramel Cluster Cashewmilk, Non-Dairy Frozen Dessert, One Pint (473 mL). UPC 744473 476138. Manufactured By/For So Delicious Dairy Free Eugene, OR 97402 So Delicious Dairy Free Salted Caramel Cluster Cashewmilk, Non-Dairy Frozen Dessert. Net wt. 500mL. UPC 744473 276134. Product of the U.S.A. Distributed by Danone Inc., Boucherville, Quebec J4B 1E6
Foreign material
Bayer 2Pk- 25 CT
Bayer 2Pk- 25 CT
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
Allura Xper FD20/20; Model Number: 722038;
Allura Xper FD20/20; Model Number: 722038;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
DPA161 Shrimp Paste (Klong Kone) 'L' , 32 oz, 12 count , PLastic Jar, 12 packages / cases
DPA161 Shrimp Paste (Klong Kone) 'L' , 32 oz, 12 count , PLastic Jar, 12 packages / cases
Product tested high for lead.
Personal Care, Vaporizing Chest Rub, 4 oz, 12 count, UPC 048155903319
Personal Care, Vaporizing Chest Rub, 4 oz, 12 count, UPC 048155903319
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.
Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System.
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
Ukrop's 6" Pound Cake with Buttercream Icing/Pink Roses NET WT 44 OZ (2.75lb/1.25kg) CONTAINS: Egg, Milk, Wheat Made in a production area that also uses soy and tree nuts. UPC 7 22515 91452 7 Ukrop's Homestyle Foods, LLC Richmond, VA 23230
Ukrop's 6" Pound Cake with Buttercream Icing/Pink Roses NET WT 44 OZ (2.75lb/1.25kg) CONTAINS: Egg, Milk, Wheat Made in a production area that also uses soy and tree nuts. UPC 7 22515 91452 7 Ukrop's Homestyle Foods, LLC Richmond, VA 23230
Undeclared Soy. Label did not include soy, which was used in product.
Halls Cough Drops a) BAGS, MENTHOL, 12CT; b) BAGS, HONEY LEM, 12CT; c) BAGS, DEFENSE CITRUS, 12CT; d) BAGS, CHERRY, 12CT; e) M. CHERRY, 20 STK; f) BRAZIL, Cherry, 21CT/27.5G; f) LEMON. HONEY, 20CT #1180E; g) LIME FRESH, 20CT, #1182; h) MENTHOL BLUE, 20CT, #1180A; i) BRAZIL, Menthol, 21CT/27.5G, j) M . STRAWBERRY, 20 STK; k) DEFENSE, VITAMIN C, 20CT, #1181
Halls Cough Drops a) BAGS, MENTHOL, 12CT; b) BAGS, HONEY LEM, 12CT; c) BAGS, DEFENSE CITRUS, 12CT; d) BAGS, CHERRY, 12CT; e) M. CHERRY, 20 STK; f) BRAZIL, Cherry, 21CT/27.5G; f) LEMON. HONEY, 20CT #1180E; g) LIME FRESH, 20CT, #1182; h) MENTHOL BLUE, 20CT, #1180A; i) BRAZIL, Menthol, 21CT/27.5G, j) M . STRAWBERRY, 20 STK; k) DEFENSE, VITAMIN C, 20CT, #1181
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
PEPCID COMPLETE - 25CT
PEPCID COMPLETE - 25CT
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
HerbsForever- Gastro Care; Ayruvedic Herbal Supplement; 90 VEGE Capsules; Extract 800 MG each capsule. UPC: 8 07814 00124 3
HerbsForever- Gastro Care; Ayruvedic Herbal Supplement; 90 VEGE Capsules; Extract 800 MG each capsule. UPC: 8 07814 00124 3
Potential undeclared allergen ingredient (Wheat)