États-Unis Recalls

Rappels de sécurité de la FDA, CPSC, NHTSA et de l’USDA FSIS.

Showing 24 of 17,562 results
Medical DeviceUSFDA Devices

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

GE Medical Systems Information Technologies Inc
FDA DevicesClass II

GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.

Aug 12, 2026GE Medical Systems Information Technologies Inc
Medical DeviceUSFDA Devices

Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Pride Mobility Products Corp.
FDA DevicesClass II

Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.

Aug 12, 2026Pride Mobility Products Corp.
Vehicle RecallUSNHTSA

Odyne Systems, LLC - High Voltage Battery May Fail and Cause Fire

Odyne Systems, LLC
NHTSAELECTRICAL SYSTEM

Odyne Systems, LLC - High Voltage Battery May Fail and Cause Fire

Battery failure can also increase the risk of a fire.

Aug 12, 2026Odyne Systems, LLC
Medical DeviceUSFDA Devices

Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF

Freudenberg Medical, LLC
FDA DevicesClass II

Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF

Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

Aug 12, 2026Freudenberg Medical, LLC
Medical DeviceUSFDA Devices

Radial Jaw 4 Jumbo Biopsy Forceps with the following descriptions and UPN: Radial Jaw 4 Jumbo 240cm Box 5; Outer Box UPN: M00513363; Inner Box UPN: M00513360.

Boston Scientific Corporation
FDA DevicesClass II

Radial Jaw 4 Jumbo Biopsy Forceps with the following descriptions and UPN: Radial Jaw 4 Jumbo 240cm Box 5; Outer Box UPN: M00513363; Inner Box UPN: M00513360.

Potential for breach in sterile barrier of device packaging.

Aug 12, 2026Boston Scientific Corporation
Medical DeviceUSFDA Devices

Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

Siemens Healthcare Diagnostics, Inc.
FDA DevicesClass II

Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.

Aug 12, 2026Siemens Healthcare Diagnostics, Inc.
Medical DeviceUSFDA Devices

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

TISSIUM SA
FDA DevicesClass II

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Aug 12, 2026TISSIUM SA
Medical DeviceUSFDA Devices

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.

Oxos Medical Inc
FDA DevicesClass II

Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.

Due to a defective circuit board that overheats

Aug 12, 2026Oxos Medical Inc
Medical DeviceUSFDA Devices

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

TISSIUM SA
FDA DevicesClass II

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Aug 12, 2026TISSIUM SA
Medical DeviceUSFDA Devices

FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.

INTEGRATED MEDCRAFT LLC
FDA DevicesClass II

FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.

Devices could potentially contain an occluded or partially occluded Guedel Airway tip.

Aug 12, 2026INTEGRATED MEDCRAFT LLC
Food & BeverageUSFDA Food

Blank Slate Creamery VEGAN NON-DAIRY FROZEN DESSERT COCONUT FUDGE SANDWICHES, 4.5 oz clear plastic package

Blank Slate Creamery Production
FDA FoodClass I

Blank Slate Creamery VEGAN NON-DAIRY FROZEN DESSERT COCONUT FUDGE SANDWICHES, 4.5 oz clear plastic package

undeclared egg

Aug 12, 2026Blank Slate Creamery Production
Food & BeverageUSFDA Food

Clover Hill Dairy LLP, Flavored Cheese (horseradish). Sold in approximate 1/2-lb, 1-lb, 2.5-lb, 5-lb, and 40-lb units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.

Clover Hill Dairy, LLC
FDA FoodClass I

Clover Hill Dairy LLP, Flavored Cheese (horseradish). Sold in approximate 1/2-lb, 1-lb, 2.5-lb, 5-lb, and 40-lb units. Unknown information: packaging, labeling, or total quantity of product distributed. Keep refrigerated.

Product contaminated with Listeria monocytogenes.

Aug 12, 2026Clover Hill Dairy, LLC
Medical DeviceUSFDA Devices

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

MUNEVO GMBH
FDA DevicesClass II

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

Aug 12, 2026MUNEVO GMBH
PharmaceuticalUSFDA Drugs

Cyclophosphamide for Injection, USP, 1 gram/vial, One Single-Dose Vial, Rx only, Manufactured for: Long Grove Pharmaceuticals, LLC, Rosemont, IL 60018. Manufactured in Taiwan, NDC 81298-8112-1

Sunny Pharmtech Inc.
FDA DrugsClass I

Cyclophosphamide for Injection, USP, 1 gram/vial, One Single-Dose Vial, Rx only, Manufactured for: Long Grove Pharmaceuticals, LLC, Rosemont, IL 60018. Manufactured in Taiwan, NDC 81298-8112-1

Presence of Particulate Matter; identified as stainless steel

Aug 12, 2026Sunny Pharmtech Inc.
PharmaceuticalUSFDA Drugs

Mirabegron Extended-Release Tablets, 25 mg, 30-count bottle, Rx only, Manufactured by: Zydus Lifesciences Ltd., Matoda, Ahmedabad, India, Distributed by: Zydus Pharmaceuticals (USA) Inc., Pennington, NJ 08534, NDC 70710-1159-3.

Zydus Pharmaceuticals (USA) Inc
FDA DrugsClass II

Mirabegron Extended-Release Tablets, 25 mg, 30-count bottle, Rx only, Manufactured by: Zydus Lifesciences Ltd., Matoda, Ahmedabad, India, Distributed by: Zydus Pharmaceuticals (USA) Inc., Pennington, NJ 08534, NDC 70710-1159-3.

CGMP: Due to Out of Specification (OOS) result for N-Nitroso Mirabegron impurity.

Aug 12, 2026Zydus Pharmaceuticals (USA) Inc
Medical DeviceUSFDA Devices

Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary Bone Saw Blade Assembly 16mm

Biomet, Inc.
FDA DevicesClass II

Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary Bone Saw Blade Assembly 16mm

Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

Aug 12, 2026Biomet, Inc.
Medical DeviceUSFDA Devices

Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5336, 9323-5338, 9323-5340 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

United Orthopedic Corp.
FDA DevicesClass II

Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5336, 9323-5338, 9323-5340 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

Aug 12, 2026United Orthopedic Corp.
Medical DeviceUSFDA Devices

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

TISSIUM SA
FDA DevicesClass II

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Aug 12, 2026TISSIUM SA
Food & BeverageUSFDA Food

La Colonia Foods Salvadoreno Requeson Soft Ricotta, Net Wt: 14 oz (367g), packaged into retail-sized clear plastic containers. Keep refrigerated.

La Colonia Foods Llc
FDA FoodClass I

La Colonia Foods Salvadoreno Requeson Soft Ricotta, Net Wt: 14 oz (367g), packaged into retail-sized clear plastic containers. Keep refrigerated.

Products may be contaminated with Listeria monocytogenes.

Aug 12, 2026La Colonia Foods Llc
Medical DeviceUSFDA Devices

Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

United Orthopedic Corp.
FDA DevicesClass II

Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

Aug 12, 2026United Orthopedic Corp.
Medical DeviceUSFDA Devices

Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm

Biomet, Inc.
FDA DevicesClass II

Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm

Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

Aug 12, 2026Biomet, Inc.
Medical DeviceUSFDA Devices

Merge PACS, medical imaging diagnostic workstation

Merge Healthcare, Inc.
FDA DevicesClass II

Merge PACS, medical imaging diagnostic workstation

Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.

Aug 12, 2026Merge Healthcare, Inc.
Medical DeviceUSFDA Devices

Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Pride Mobility Products Corp.
FDA DevicesClass II

Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.

Aug 12, 2026Pride Mobility Products Corp.
Medical DeviceUSFDA Devices

Radial Jaw 4 Large Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 LC 240cm Box 20; Outer Box UPN: M00513321; Inner Box UPN: M00513320. 2. Radial Jaw 4 LC 240cm Box 40; Outer Box UPN: M00513322; Inner Box UPN: M00513320. 3. Radial Jaw 4 LC w/Needle 240cm Box 20; Outer Box UPN: M00513331; Inner Box UPN: M00513330. 4. Radial Jaw 4 LC w/Needle 240cm Box 40; Outer Box UPN: M00513332; Inner Box UPN: M00513330. 5. Radial Jaw 4 LC w/Needle 240cm Box 5; Outer Box UPN: M00513333; Inner Box UPN: M00513330.

Boston Scientific Corporation
FDA DevicesClass II

Radial Jaw 4 Large Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 LC 240cm Box 20; Outer Box UPN: M00513321; Inner Box UPN: M00513320. 2. Radial Jaw 4 LC 240cm Box 40; Outer Box UPN: M00513322; Inner Box UPN: M00513320. 3. Radial Jaw 4 LC w/Needle 240cm Box 20; Outer Box UPN: M00513331; Inner Box UPN: M00513330. 4. Radial Jaw 4 LC w/Needle 240cm Box 40; Outer Box UPN: M00513332; Inner Box UPN: M00513330. 5. Radial Jaw 4 LC w/Needle 240cm Box 5; Outer Box UPN: M00513333; Inner Box UPN: M00513330.

Potential for breach in sterile barrier of device packaging.

Aug 12, 2026Boston Scientific Corporation