Dispositifs Médicaux Recalls

Defective surgical instruments, diagnostic equipment, and health device recalls.

Showing 24 of 2,119 results
Medical DeviceUSFDA Devices

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

ETAC A/S
FDA DevicesClass II

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Feb 11, 2026ETAC A/S
Medical DeviceUSFDA Devices

NAEOTOM Alpha.Prime Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS

Siemens Medical Solutions USA, Inc
FDA DevicesClass II

NAEOTOM Alpha.Prime Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS

To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Feb 11, 2026Siemens Medical Solutions USA, Inc
Medical DeviceUSFDA Devices

Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327

Vortex Surgical Inc.
FDA DevicesClass II

Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Feb 11, 2026Vortex Surgical Inc.
Medical DeviceUSFDA Devices

PRISMAFLEX ST100 SET US, Product Code 107636US; Dialyzer, High Permeability With Or Without Sealed Dialysate System

VANTIVE US HEALTHCARE LLC
FDA DevicesClass II

PRISMAFLEX ST100 SET US, Product Code 107636US; Dialyzer, High Permeability With Or Without Sealed Dialysate System

There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

Feb 11, 2026VANTIVE US HEALTHCARE LLC
Medical DeviceUSFDA Devices

Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Olympus Corporation of the Americas
FDA DevicesClass II

Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Devices which did not undergo thermoforming could deform and lose performance.

Feb 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Olympus Corporation of the Americas
FDA DevicesClass II

Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Devices which did not undergo thermoforming could deform and lose performance.

Feb 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705

Encore Medical, LP
FDA DevicesClass II

EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

Feb 11, 2026Encore Medical, LP
Medical DeviceUSFDA Devices

Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Olympus Corporation of the Americas
FDA DevicesClass II

Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Devices which did not undergo thermoforming could deform and lose performance.

Feb 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing pTau 217 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse pTau 217 Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Fujirebio Diagnostics, Inc.
FDA DevicesClass II

Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description: Lumipulse pTau 217 Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing pTau 217 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse pTau 217 Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Feb 11, 2026Fujirebio Diagnostics, Inc.
Medical DeviceUSFDA Devices

Vortex Surgical I2 Injection Kit, VS0500

Vortex Surgical Inc.
FDA DevicesClass II

Vortex Surgical I2 Injection Kit, VS0500

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Feb 11, 2026Vortex Surgical Inc.
Medical DeviceUSFDA Devices

NAEOTOM Alpha Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS

Siemens Medical Solutions USA, Inc
FDA DevicesClass II

NAEOTOM Alpha Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS

To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Feb 11, 2026Siemens Medical Solutions USA, Inc
Medical DeviceUSFDA Devices

Vortex Surgical TID Pharos Illuminated Depressor, VS0801B

Vortex Surgical Inc.
FDA DevicesClass II

Vortex Surgical TID Pharos Illuminated Depressor, VS0801B

XXX

Feb 11, 2026Vortex Surgical Inc.
Medical DeviceUSFDA Devices

Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse ¿-Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators set 235454 Component: No

Fujirebio Diagnostics, Inc.
FDA DevicesClass II

Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse ¿-Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators set 235454 Component: No

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Feb 11, 2026Fujirebio Diagnostics, Inc.
Medical DeviceUSFDA Devices

Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.

A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.

Feb 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD

Vortex Surgical Inc.
FDA DevicesClass II

Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Feb 11, 2026Vortex Surgical Inc.
Medical DeviceUSFDA Devices

Vivoo pH Test

Changchun Wancheng Bio-Electron Co., Ltd.
FDA DevicesClass II

Vivoo pH Test

Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.

Feb 11, 2026Changchun Wancheng Bio-Electron Co., Ltd.
Medical DeviceUSFDA Devices

Vivoo Sodium Test; Vivoo Vitamin C Test; Vivoo Hydration Test

Changchun Wancheng Bio-Electron Co., Ltd.
FDA DevicesClass II

Vivoo Sodium Test; Vivoo Vitamin C Test; Vivoo Hydration Test

Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.

Feb 11, 2026Changchun Wancheng Bio-Electron Co., Ltd.
Medical DeviceUSFDA Devices

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491

Biomerieux Inc
FDA DevicesClass II

VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 ***Update 4/2/26 -- Additional product references not manufactured at time of initial notification. These products will not be distributed within the United States.*** AST-N484 Ref. 425181 AST-N486 Ref. 425183 AST-N491 Ref. 425206 AST-N494 Ref. 425224 AST-N485 Ref. 425182 AST-XN41 Ref. 425204 AST-XN42 Ref. 425309 AST-N504 Ref. 425491

Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.

Feb 11, 2026Biomerieux Inc
Medical DeviceUSFDA Devices

Rumex Disposable Diamond Dusted ILM Elevator, 12-7523

Vortex Surgical Inc.
FDA DevicesClass II

Rumex Disposable Diamond Dusted ILM Elevator, 12-7523

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Feb 11, 2026Vortex Surgical Inc.
Medical DeviceUSFDA Devices

The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips; URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein; Vivoo Ketone Test

Changchun Wancheng Bio-Electron Co., Ltd.
FDA DevicesClass II

The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips; URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein; Vivoo Ketone Test

Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.

Feb 11, 2026Changchun Wancheng Bio-Electron Co., Ltd.
Medical DeviceUSFDA Devices

HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;

Boston Scientific Corporation
FDA DevicesClass I

HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Feb 11, 2026Boston Scientific Corporation
Medical DeviceUSFDA Devices

Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Olympus Corporation of the Americas
FDA DevicesClass II

Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Devices which did not undergo thermoforming could deform and lose performance.

Feb 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

PRISMAFLEX M150 SET, Product Code 109990; Dialyzer, High Permeability With Or Without Sealed Dialysate System

VANTIVE US HEALTHCARE LLC
FDA DevicesClass II

PRISMAFLEX M150 SET, Product Code 109990; Dialyzer, High Permeability With Or Without Sealed Dialysate System

There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

Feb 11, 2026VANTIVE US HEALTHCARE LLC
Medical DeviceUSFDA Devices

Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Olympus Corporation of the Americas
FDA DevicesClass II

Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Devices which did not undergo thermoforming could deform and lose performance.

Feb 11, 2026Olympus Corporation of the Americas