Dispositifs Médicaux Recalls
Defective surgical instruments, diagnostic equipment, and health device recalls.
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Probes may rupture/burst during activation
CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP15K/B, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT19K/D, CENFT23K, CENFT23K/A, CENFT23K/B, CENFT23K/D, CENFT27K, CENFT27K/A, CENFT27K/B, CENFT27K/D, CENFT31K, CENFT31K/A, CENFT31K/B, CENFT31K/D
CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP15K/B, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT19K/D, CENFT23K, CENFT23K/A, CENFT23K/B, CENFT23K/D, CENFT27K, CENFT27K/A, CENFT27K/B, CENFT27K/D, CENFT31K, CENFT31K/A, CENFT31K/B, CENFT31K/D
16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.
Philips L12-5 Ultrasound Transducer
Philips L12-5 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Probes may rupture/burst during activation
Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd
If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
Philips S4-1 Ultrasound Transducer
Philips S4-1 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Philips C9-4 Ultrasound Transducer
Philips C9-4 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.
I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.
I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Philips L17-5 Ultrasound Transducer
Philips L17-5 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.
The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N, 37351-34-N, 37351-36-N, 37351-38-N, 37351-40-N, 37351-42-N, 37351-44-N, 37351-46-N, 37351-48-N, 37351-50-N, 37351-55-N, 37351-60-N, 37351-65-N, 37351-70-N, 37351-75-N, 37351-80-N, 37351-85-N. 2. Humeral Plate, Distal; Article Numbers: 21241-5, 21241-7, 21242-7, 21243-10, 21244-9.
Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N, 37351-34-N, 37351-36-N, 37351-38-N, 37351-40-N, 37351-42-N, 37351-44-N, 37351-46-N, 37351-48-N, 37351-50-N, 37351-55-N, 37351-60-N, 37351-65-N, 37351-70-N, 37351-75-N, 37351-80-N, 37351-85-N. 2. Humeral Plate, Distal; Article Numbers: 21241-5, 21241-7, 21242-7, 21243-10, 21244-9.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Probes may rupture/burst during activation
Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
Philips 3D6-2 Ultrasound Transducer
Philips 3D6-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.