Dispositifs Médicaux Recalls
Defective surgical instruments, diagnostic equipment, and health device recalls.
Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software Version: ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200) ARVIS Shoulder Software v2.0.3 (IN-23630) Product Description: The ARVIS system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. The ARVIS Surgical Navigation System is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements such as changes in leg length in Hip Arthroplasty. The system is compatible with straight acetabular impactors and with specific offset impactors, identified in the instructions for use, for which an adapter has been validated. Example orthopedic surgical procedures include but are not limited to: " Total Knee Arthroplasty " Unicompartmental Knee Arthroplasty: Tibial Transverse Resection " Hip Arthroplasty The ARVIS" head mounted display is for displaying augmented reality visualization and information to the user intraoperatively. The augmented/virtual displayed information should not be relied upon solely for absolute positional/alignment information and should always be used in conjunction with the displayed stereotaxic information. Component: NA
Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software Version: ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200) ARVIS Shoulder Software v2.0.3 (IN-23630) Product Description: The ARVIS system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. The ARVIS Surgical Navigation System is indicated for assisting the surgeon in the positioning and alignment of implants relative to reference alignment axes and landmarks in stereotactic orthopedic surgery. The system aids the surgeon in making intraoperative measurements such as changes in leg length in Hip Arthroplasty. The system is compatible with straight acetabular impactors and with specific offset impactors, identified in the instructions for use, for which an adapter has been validated. Example orthopedic surgical procedures include but are not limited to: " Total Knee Arthroplasty " Unicompartmental Knee Arthroplasty: Tibial Transverse Resection " Hip Arthroplasty The ARVIS" head mounted display is for displaying augmented reality visualization and information to the user intraoperatively. The augmented/virtual displayed information should not be relied upon solely for absolute positional/alignment information and should always be used in conjunction with the displayed stereotaxic information. Component: NA
Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder
Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder
Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
Blanketrol II Model 222S; Thermal regulating system.
Blanketrol II Model 222S; Thermal regulating system.
Overheating of the 115V CPU board may cause the device to shut down.
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
One-Piece Guedel Airway, size 00, ISO 5.0, Blue. Oropharyngeal Airway Device.
One-Piece Guedel Airway, size 00, ISO 5.0, Blue. Oropharyngeal Airway Device.
Devices could potentially be occluded or partially occluded.
Halyard LAVH PACK, Model/Catalog Number: USAR053-11 containing Clearify Visualization System (5303907).
Halyard LAVH PACK, Model/Catalog Number: USAR053-11 containing Clearify Visualization System (5303907).
The kit component was subjected to EO sterilization, contrary to the manufacturer recommendation, which may have led to the component leaking in a kit.
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
One-Piece Guedel Airway, size 000, ISO 3.5, Light Green. Oropharyngeal Airway Device.
One-Piece Guedel Airway, size 000, ISO 3.5, Light Green. Oropharyngeal Airway Device.
Devices could potentially be occluded or partially occluded.
Innercool Coolblue Console, Model STx; Thermal regulating system.
Innercool Coolblue Console, Model STx; Thermal regulating system.
Overheating of the 115V CPU board may cause the device to shut down.
OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.
OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.
Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Blanketrol III, Model 233; Thermal regulating system.
Blanketrol III, Model 233; Thermal regulating system.
Overheating of the 115V CPU board may cause the device to shut down.
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
An MHRA investigation has found increased risks of wear and corrosion associated with multiple cobalt-chrome (Co-Cr) modular neck hip stem components, including a potential increased occurrence of rev...
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy
Boston Scientific ENROUTE Transcarotid Neuroprotection System, Material Number SR-200-NPS; Percutaneous catheter
Boston Scientific ENROUTE Transcarotid Neuroprotection System, Material Number SR-200-NPS; Percutaneous catheter
Potential for arterial sheath tip separation or partial tip separation during use.
Ami HTX
Ami HTX
Momentary absence of On-Screen "X-Ray On" banner during run readiness check.
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Boston Scientific ENROUTE Transcarotid Neuroprotection System Plus, Material Number SR-250-NPS; Percutaneous catheter
Boston Scientific ENROUTE Transcarotid Neuroprotection System Plus, Material Number SR-250-NPS; Percutaneous catheter
Potential for arterial sheath tip separation or partial tip separation during use.
Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev A, Model/Catalog Number 6123N R; foot implant
Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev A, Model/Catalog Number 6123N R; foot implant
Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
Lago X
Lago X
Momentary absence of On-Screen "X-Ray On" banner during run readiness check.