Allura Xper FD10/10; Model Number: 722011;
Allura Xper FD10/10; Model Number: 722011;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Allura Xper FD10F; Model Number: 722002;
Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU DAVINCI, REF CDS982779N; 9) GYN DAVINCI, REF CDS982780V; 10) UNIVERSAL ROBOTIC ORLANDO, REF CDS982857K; 11) ROBOTIC GU/GYN, REF CDS983393L; 12) GYN LAPAROSCOPY CDS, REF CDS984656Q; 13) GYN LAPAROSCOPY CDS, REF CDS984656R; 14) DAVINCI CDS UCC-LF, REF CDS984663J; 15) ROBOTIC UROLOGY PROSTATE, REF CDS985467J; 16) DAVINCI PROSTATECTOMY, REF DYNJG901002B; 17) ROBOTIC PACK IMF 100913- LF, REF DYNJVB91075C; 18) ROBOTIC PACK, REF DYNJ36840L; 19) ROBOTIC PACK MNMC, REF DYNJ38152D; 20) WMC GYN LAPAROSCOPY PACK-LF, REF DYNJ50649J; 21) ROBOTIC GASTRIC BYPASS PACK, REF DYNJ52096P; 22) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564D; 23) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564F; 24) GYN ROBOTIC PACK, REF DYNJ52581D; 25) GYN LAPAROSCOPY PACK, REF DYNJ53782J; 26) UNIVERSAL DAVINCI ROBOTIC P, REF DYNJ53968P; 27) WALKER GYN-UROLOGY ROBOTIC PK, REF DYNJ57458B; 28) GYN LAPAROSCOPY, REF DYNJ62136B; 29) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392F; 30) TOTAL LAPAROSCOPIC HYST PACK, REF DYNJ63456B; 31) ROBOTIC PACK, REF DYNJ67177D; 32) MH GYN LAPAROSCOPY PACK-LF, REF DYNJ84433B; 33) DAVINCI, REF DYNJ900963M; 34) LEX DAVINCI GYN MAJOR, REF DYNJ900965N; 35) LEX DAVINCI UROLOGY, REF DYNJ900967N; 36) COR DAVINCI, REF DYNJ901018O; 37) GYN ROBOTIC, REF DYNJ901046M; 38) DAVINCI, REF DYNJ901076L; 39) GYN ROBOTIC, REF DYNJ901796K; 40) LOU GENERAL ROBOT, REF DYNJ901844N; 41) LOU GYN LAPAROSCOPY, REF DYNJ901845L; 42) LOU GYN LAPAROSCOPY, REF DYNJ901845M; 43) LOU LITHOTOMY ROBOTIC, REF DYNJ901850N; 44) LEX GYN MINOR LAPAROSCOPY, REF DYNJ902029G; 45) GYN ROBOT, REF DYNJ902513J; 46) CHN OR DAVINCI PROSTATE, REF DYNJ902540O; 47) WMC ROBOTIC PACK - LF, REF DYNJ902583N; 48) ROBOTIC GYN & GU, REF DYNJ903459J; 49) GYN LAPAROSCOPY-LF, REF DYNJ903647I; 50) ROBOTIC UROLOGY, REF DYNJ903827F; 51) LITHOTOMY ROBOTIC CDS, REF DYNJ903948R; 52) GYN ROBOTIC, REF DYNJ903983P; 53) GYN LAP, REF DYNJ903990Q; 54) ROBOTIC-LF, REF DYNJ904254L; 55) ROBOTIC, REF DYNJ904343G; 56) ROBOT GYN, REF DYNJ904572L; 57) ROBOTIC PROSTECTOMY, REF DYNJ904597G; 58) GYN LAPAROSCOPY-LF, REF DYNJ905289G; 59) GYN ROBOTIC, REF DYNJ905730J; 60) URO ROBOTIC, REF DYNJ905746J; 61) COLON KIT, REF DYNJ905964G; 62) CHN OR ROBOTIC GYN, REF DYNJ905985O; 63) CW ROBOT, REF DYNJ906051C; 64) CW ROBOT, REF DYNJ906051D; 65) ROBOT, REF DYNJ906112B; 66) CDS ROBOTIC GYN CH, REF DYNJ906265D; 67) ROBOTIC GYN PACK, REF DYNJ906265F; 68) KIT GYN LAPAROSCOPY RFD, REF DYNJ906322F; 69) GYN LAPAROSCOPY, REF DYNJ906357C; 70) ROBOTIC COLON, REF DYNJ906379D; 71) GYN DAVINCI FOAKS, REF DYNJ906444C; 72) DAVINCI, REF DYNJ906543B; 73) CHN GYN ONCOLOGY, REF DYNJ906731K; 74) CHN ROBOTIC CYSTECTOMY, REF DYNJ907072F; 75) WMC THORACIC ROBOTIC, REF DYNJ907153C; 76) WMC THORACIC ROBOTIC, REF DYNJ907153D; 77) GYN ROBOTIC, REF DYNJ907492F; 78) KIT LAP GYN NTX, REF DYNJ907599C; 79) KIT ROBOTIC GYN, REF DYNJ907635D; 80) KIT ROBOTIC GYN, REF DYNJ907647B; 81) ROBOTIC-LF, REF DYNJ907910F; 82) ROBOTIC GENERAL, REF DYNJ907919C; 83) ROBOTIC BASE, REF DYNJ908447D; 84) ROBOTIC BASE, REF DYNJ908447F; 85) GYN, REF DYNJ908504A; 86) GU DAVINCI KIT, REF DYNJ908783F; 87) ROBOTIC PROSTATE, REF DYNJ909120D; 88) OSC ROBOT BARIATRIC, REF DYNJ909547; 89) LEX DAVINCI PROSTATE, REF DYNJ909548A; 90) GYN-URO ROBOT, REF DYNJ910137A; 91) PROSTATECTOMY ROBOTIC, REF DYNJ910991; 92) GYN ROBOTIC, REF DYNJ910995; 93) ROBOTIC, REF DYNJ9225761X; 94) DAVINCI ROBOTIC PACK-LF, REF DYNJ9288198X; 95) MB GYN ROBOTIC ADD, REF DYNJ9701096M; 96) PHS722500FCK, REF PHS722500G.
Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU DAVINCI, REF CDS982779N; 9) GYN DAVINCI, REF CDS982780V; 10) UNIVERSAL ROBOTIC ORLANDO, REF CDS982857K; 11) ROBOTIC GU/GYN, REF CDS983393L; 12) GYN LAPAROSCOPY CDS, REF CDS984656Q; 13) GYN LAPAROSCOPY CDS, REF CDS984656R; 14) DAVINCI CDS UCC-LF, REF CDS984663J; 15) ROBOTIC UROLOGY PROSTATE, REF CDS985467J; 16) DAVINCI PROSTATECTOMY, REF DYNJG901002B; 17) ROBOTIC PACK IMF 100913- LF, REF DYNJVB91075C; 18) ROBOTIC PACK, REF DYNJ36840L; 19) ROBOTIC PACK MNMC, REF DYNJ38152D; 20) WMC GYN LAPAROSCOPY PACK-LF, REF DYNJ50649J; 21) ROBOTIC GASTRIC BYPASS PACK, REF DYNJ52096P; 22) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564D; 23) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564F; 24) GYN ROBOTIC PACK, REF DYNJ52581D; 25) GYN LAPAROSCOPY PACK, REF DYNJ53782J; 26) UNIVERSAL DAVINCI ROBOTIC P, REF DYNJ53968P; 27) WALKER GYN-UROLOGY ROBOTIC PK, REF DYNJ57458B; 28) GYN LAPAROSCOPY, REF DYNJ62136B; 29) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392F; 30) TOTAL LAPAROSCOPIC HYST PACK, REF DYNJ63456B; 31) ROBOTIC PACK, REF DYNJ67177D; 32) MH GYN LAPAROSCOPY PACK-LF, REF DYNJ84433B; 33) DAVINCI, REF DYNJ900963M; 34) LEX DAVINCI GYN MAJOR, REF DYNJ900965N; 35) LEX DAVINCI UROLOGY, REF DYNJ900967N; 36) COR DAVINCI, REF DYNJ901018O; 37) GYN ROBOTIC, REF DYNJ901046M; 38) DAVINCI, REF DYNJ901076L; 39) GYN ROBOTIC, REF DYNJ901796K; 40) LOU GENERAL ROBOT, REF DYNJ901844N; 41) LOU GYN LAPAROSCOPY, REF DYNJ901845L; 42) LOU GYN LAPAROSCOPY, REF DYNJ901845M; 43) LOU LITHOTOMY ROBOTIC, REF DYNJ901850N; 44) LEX GYN MINOR LAPAROSCOPY, REF DYNJ902029G; 45) GYN ROBOT, REF DYNJ902513J; 46) CHN OR DAVINCI PROSTATE, REF DYNJ902540O; 47) WMC ROBOTIC PACK - LF, REF DYNJ902583N; 48) ROBOTIC GYN & GU, REF DYNJ903459J; 49) GYN LAPAROSCOPY-LF, REF DYNJ903647I; 50) ROBOTIC UROLOGY, REF DYNJ903827F; 51) LITHOTOMY ROBOTIC CDS, REF DYNJ903948R; 52) GYN ROBOTIC, REF DYNJ903983P; 53) GYN LAP, REF DYNJ903990Q; 54) ROBOTIC-LF, REF DYNJ904254L; 55) ROBOTIC, REF DYNJ904343G; 56) ROBOT GYN, REF DYNJ904572L; 57) ROBOTIC PROSTECTOMY, REF DYNJ904597G; 58) GYN LAPAROSCOPY-LF, REF DYNJ905289G; 59) GYN ROBOTIC, REF DYNJ905730J; 60) URO ROBOTIC, REF DYNJ905746J; 61) COLON KIT, REF DYNJ905964G; 62) CHN OR ROBOTIC GYN, REF DYNJ905985O; 63) CW ROBOT, REF DYNJ906051C; 64) CW ROBOT, REF DYNJ906051D; 65) ROBOT, REF DYNJ906112B; 66) CDS ROBOTIC GYN CH, REF DYNJ906265D; 67) ROBOTIC GYN PACK, REF DYNJ906265F; 68) KIT GYN LAPAROSCOPY RFD, REF DYNJ906322F; 69) GYN LAPAROSCOPY, REF DYNJ906357C; 70) ROBOTIC COLON, REF DYNJ906379D; 71) GYN DAVINCI FOAKS, REF DYNJ906444C; 72) DAVINCI, REF DYNJ906543B; 73) CHN GYN ONCOLOGY, REF DYNJ906731K; 74) CHN ROBOTIC CYSTECTOMY, REF DYNJ907072F; 75) WMC THORACIC ROBOTIC, REF DYNJ907153C; 76) WMC THORACIC ROBOTIC, REF DYNJ907153D; 77) GYN ROBOTIC, REF DYNJ907492F; 78) KIT LAP GYN NTX, REF DYNJ907599C; 79) KIT ROBOTIC GYN, REF DYNJ907635D; 80) KIT ROBOTIC GYN, REF DYNJ907647B; 81) ROBOTIC-LF, REF DYNJ907910F; 82) ROBOTIC GENERAL, REF DYNJ907919C; 83) ROBOTIC BASE, REF DYNJ908447D; 84) ROBOTIC BASE, REF DYNJ908447F; 85) GYN, REF DYNJ908504A; 86) GU DAVINCI KIT, REF DYNJ908783F; 87) ROBOTIC PROSTATE, REF DYNJ909120D; 88) OSC ROBOT BARIATRIC, REF DYNJ909547; 89) LEX DAVINCI PROSTATE, REF DYNJ909548A; 90) GYN-URO ROBOT, REF DYNJ910137A; 91) PROSTATECTOMY ROBOTIC, REF DYNJ910991; 92) GYN ROBOTIC, REF DYNJ910995; 93) ROBOTIC, REF DYNJ9225761X; 94) DAVINCI ROBOTIC PACK-LF, REF DYNJ9288198X; 95) MB GYN ROBOTIC ADD, REF DYNJ9701096M; 96) PHS722500FCK, REF PHS722500G.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
Allura Xper FD10/10; Model Number: 722005;
Allura Xper FD10/10; Model Number: 722005;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Allura Xper FD10/10; Model Number: 722027;
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Swan, White Clear Alcohol, 50%, 16oz, 12 count
Swan, White Clear Alcohol, 50%, 16oz, 12 count
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
SMV DSX New Line, System, Tomography, Computed, Emission
SMV DSX New Line, System, Tomography, Computed, Emission
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Allura Xper FD20 Biplane; Model Number: 722008;
Allura Xper FD20 Biplane; Model Number: 722008;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
DAYQUIL COLD & FLU, 32CT/2PK
DAYQUIL COLD & FLU, 32CT/2PK
CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.
Allura Xper FD20 OR Table; Model Number: 722023;
Allura Xper FD20 OR Table; Model Number: 722023;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
Volkswagen Group of America, Inc. - High-Voltage Battery May Overheat and Cause Fire
Volkswagen Group of America, Inc. - High-Voltage Battery May Overheat and Cause Fire
A high-voltage battery that overheats increases the risk of a fire.
Alexis Fire Equipment Company - Loss of Vehicle Control from Loose Axle Clamp Fasteners
Alexis Fire Equipment Company - Loss of Vehicle Control from Loose Axle Clamp Fasteners
A loss of vehicle control increases the risk of a crash.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-LE FORT MANDIBLE PACK-LF, Medline Kit SKU DYNJ0785864R; 2) RR-ASC EAR PACK-LF, Medline Kit SKU DYNJ0792607R.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-LE FORT MANDIBLE PACK-LF, Medline Kit SKU DYNJ0785864R; 2) RR-ASC EAR PACK-LF, Medline Kit SKU DYNJ0792607R.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Brand Name: PALACOS Product Name: PALACOS MV pro 40; PALACOS MV pro 80 Model/Catalog Number: 5150633; 5150635 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
Brand Name: PALACOS Product Name: PALACOS MV pro 40; PALACOS MV pro 80 Model/Catalog Number: 5150633; 5150635 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: AN01 CAROTID ANESTH KIT-LF, Medline Kit SKU PHS972096014B
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: AN01 CAROTID ANESTH KIT-LF, Medline Kit SKU PHS972096014B
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Fran s PureBar Almondmilk Chocolate 46% Madagascar Plant-Based, net wt. 1.1oz, UPC 7 49668 24280 2. Ingredients: Sugar, cocoa butter, cocoa beans, almonds, sunflower lecithin, vanilla extract. May contain traces of milk, peanuts and sesame. Fran's Chocolates, Ltd. Seattle, WA 98108.
Fran s PureBar Almondmilk Chocolate 46% Madagascar Plant-Based, net wt. 1.1oz, UPC 7 49668 24280 2. Ingredients: Sugar, cocoa butter, cocoa beans, almonds, sunflower lecithin, vanilla extract. May contain traces of milk, peanuts and sesame. Fran's Chocolates, Ltd. Seattle, WA 98108.
undeclared hazelnut.