Medical DeviceUSFDA Devices

Cytal Wound Matrix 2-Layer 5x5 cm. Product ID: WSM0505.

Integra LifeSciences Corp. (NeuroSciences)
FDA DevicesClass II

Cytal Wound Matrix 2-Layer 5x5 cm. Product ID: WSM0505.

Increased rate of out-of-specification endotoxin results.

Apr 22, 2026Integra LifeSciences Corp. (NeuroSciences)
Medical DeviceUSFDA Devices

B. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L.

B Braun Medical Inc
FDA DevicesClass I

B. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L.

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Apr 22, 2026B Braun Medical Inc
Vehicle RecallUSNHTSA

Toyota Motor Engineering & Manufacturing - Improperly Welded Seat Brackets

Toyota Motor Engineering & Manufacturing
NHTSASEATS

Toyota Motor Engineering & Manufacturing - Improperly Welded Seat Brackets

Improperly welded seat brackets can prevent the seat from adequately restraining an occupant during a crash, increasing the risk of injury.

Apr 22, 2026Toyota Motor Engineering & Manufacturing
Medical DeviceUSFDA Devices

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA, Inc
FDA DevicesClass II

ARTIS Pheno VE30A and VE40A, Model 10849000

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Apr 22, 2026Siemens Medical Solutions USA, Inc
Medical DeviceUSFDA Devices

Philips Spectral CT on Rails. Model Number: 728334.

PHILIPS MEDICAL SYSTEMS
FDA DevicesClass II

Philips Spectral CT on Rails. Model Number: 728334.

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Apr 22, 2026PHILIPS MEDICAL SYSTEMS
Vehicle RecallUSNHTSA

Volkswagen Group of America, Inc. - Fuel Tank May Leak After Crash

Volkswagen Group of America, Inc.
NHTSAFUEL SYSTEM, GA...

Volkswagen Group of America, Inc. - Fuel Tank May Leak After Crash

A fuel tank with a dislodged or separated fuel pressure sensor can leak fuel after a rollover crash, increasing the risk of a fire.

Apr 22, 2026Volkswagen Group of America, Inc.
Vehicle RecallUSNHTSA

Trails West Manufacturing of Idaho, Inc. - Cooktop Flame May Become Inverted

Trails West Manufacturing of Idaho, Inc.
NHTSAEQUIPMENT

Trails West Manufacturing of Idaho, Inc. - Cooktop Flame May Become Inverted

An inverted cooktop flame increases the risk of a fire.

Apr 22, 2026Trails West Manufacturing of Idaho, Inc.
PharmaceuticalUSFDA Drugs

HydroPeptide CLEAR ALLIANCE SERUM, 2.5% BENZOYL PEROXIDE, a)1 FL OZ/30ml; b) 2 FL OZ/60ml spray bottle; DIST BY HYDROPEPTIDE LLC, ISSAQUAH, WA 98027

Owen Biosciences Inc.
FDA DrugsClass II

HydroPeptide CLEAR ALLIANCE SERUM, 2.5% BENZOYL PEROXIDE, a)1 FL OZ/30ml; b) 2 FL OZ/60ml spray bottle; DIST BY HYDROPEPTIDE LLC, ISSAQUAH, WA 98027

Chemical contamination: Presence of benzene

Apr 22, 2026Owen Biosciences Inc.
Medical DeviceUSFDA Devices

Cytal Wound Matrix 2-Layer 7x10 cm. Product ID: WSM0710.

Integra LifeSciences Corp. (NeuroSciences)
FDA DevicesClass II

Cytal Wound Matrix 2-Layer 7x10 cm. Product ID: WSM0710.

Increased rate of out-of-specification endotoxin results.

Apr 22, 2026Integra LifeSciences Corp. (NeuroSciences)
Food & BeverageUKUK FSA

Good4U recalls Super Sprouts Super Greens due to possible presence of Salmonella

Good4U
UK FSAPublished

Good4U recalls Super Sprouts Super Greens due to possible presence of Salmonella

The possible presence of Salmonella in the product listed above. Symptoms caused by Salmonella usually include fever, diarrhoea and abdominal cramps. If you are experiencing these symptoms, you should stay away from work, school or nursery until you have stopped having symptoms for at least 48 hours to avoid passing it on to others.

Apr 22, 2026Good4U
Medical DeviceUSFDA Devices

Brand Name: Elegance¿ Anterior Cervical Plate Product Name: ELEGANCE 1 LEVEL CERVICAL PLATE, ELEGANCE 2 LEVEL CERVICAL PLATE, ELEGANCE 3 LEVEL CERVICAL PLATE, ELEGANCE 4 LEVEL CERVICAL PLATE Model/Catalog Number: See Excel sheet attached named "FDA 5072 Product Information" Software Version: Not applicable Product Description: The Elegance¿Anterior Cervical plate is designed for use as a cervical plate system. It has a shape which restores the intervertebral height and lordosis. The Elegance¿ Anterior Cervical plate consists in a variety of shapes and sizes of cervical plates with rounded corners, featuring bone screw s holes, locking systems and bone screws. The plates go from one (1) to five (5) levels, and the screws are selfdrilling or self- tapping, and fixed or variable. The fixation is provided by the bone screws inserted into the vertebral body by means of an anterior approach. The Elegance¿plates, as well as the bone screws, are made of compliant ASTM F136 Titanium alloy. It is essential to insert the implants with the instrumentation specifically designed for this purpose. For more description of the instrumentation, read the technical documentation associated to the Elegance¿ Anterior Cervical plate. Detailed information concerning the surgical technique of the Elegance¿Anterior Cervical plate is available upon request, please contact CLARIANCE or its local representative.

Clariance-SAS
FDA DevicesClass II

Brand Name: Elegance¿ Anterior Cervical Plate Product Name: ELEGANCE 1 LEVEL CERVICAL PLATE, ELEGANCE 2 LEVEL CERVICAL PLATE, ELEGANCE 3 LEVEL CERVICAL PLATE, ELEGANCE 4 LEVEL CERVICAL PLATE Model/Catalog Number: See Excel sheet attached named "FDA 5072 Product Information" Software Version: Not applicable Product Description: The Elegance¿Anterior Cervical plate is designed for use as a cervical plate system. It has a shape which restores the intervertebral height and lordosis. The Elegance¿ Anterior Cervical plate consists in a variety of shapes and sizes of cervical plates with rounded corners, featuring bone screw s holes, locking systems and bone screws. The plates go from one (1) to five (5) levels, and the screws are selfdrilling or self- tapping, and fixed or variable. The fixation is provided by the bone screws inserted into the vertebral body by means of an anterior approach. The Elegance¿plates, as well as the bone screws, are made of compliant ASTM F136 Titanium alloy. It is essential to insert the implants with the instrumentation specifically designed for this purpose. For more description of the instrumentation, read the technical documentation associated to the Elegance¿ Anterior Cervical plate. Detailed information concerning the surgical technique of the Elegance¿Anterior Cervical plate is available upon request, please contact CLARIANCE or its local representative.

Due to a quality issue regarding the presence of burr on the finished cervical plate product.

Apr 22, 2026Clariance-SAS
Medical DeviceUSFDA Devices

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Beta Bionics, Inc.
FDA DevicesClass II

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.

Apr 22, 2026Beta Bionics, Inc.
Medical DeviceUSFDA Devices

B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.

B Braun Medical Inc
FDA DevicesClass I

B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Apr 22, 2026B Braun Medical Inc
Vehicle RecallUSNHTSA

Winnebago Industries, Inc. - Park Outside: Generator May Turn On Without Operator Action

Winnebago Industries, Inc.
NHTSAEQUIPMENT

Winnebago Industries, Inc. - Park Outside: Generator May Turn On Without Operator Action

People in an enclosed space, such as a garage, with a running generator may be at an increased risk of injury or death from carbon monoxide poisoning.

Apr 22, 2026Winnebago Industries, Inc.
PharmaceuticalUSFDA Drugs

QC Quality Choice, Menthol- Cough Suppressant Oral Anesthetic, Cough Drops, Sugar Free, Black Cherry Flavor, 25-count bag, Distributed By: CDMA, Inc., Novi MI 48375, Made in China, NDC: 83698-616-25, UPC: 635515993372.

Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
FDA DrugsClass II

QC Quality Choice, Menthol- Cough Suppressant Oral Anesthetic, Cough Drops, Sugar Free, Black Cherry Flavor, 25-count bag, Distributed By: CDMA, Inc., Novi MI 48375, Made in China, NDC: 83698-616-25, UPC: 635515993372.

This recall is being initiated following FDA's recommendation based on certain observations noted during an August 15, 2025, inspection of the manufacturing facility that may bear on product quality.

Apr 22, 2026Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
Medical DeviceUSFDA Devices

Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable

THERAKOS DEVELOPMENT LIMITED
FDA DevicesClass II

Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable

Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.

Apr 22, 2026THERAKOS DEVELOPMENT LIMITED
PharmaceuticalUSFDA Drugs

Memantine Hydrochloride Extended-Release, Capsules, 7 mg, 100 Capsules (10 x 10 blister packs), Rx only, Packaged and Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-6734-61

The Harvard Drug Group LLC
FDA DrugsClass II

Memantine Hydrochloride Extended-Release, Capsules, 7 mg, 100 Capsules (10 x 10 blister packs), Rx only, Packaged and Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-6734-61

Failed Dissolution Specifications

Apr 22, 2026The Harvard Drug Group LLC
Medical DeviceUSFDA Devices

Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;

Gentuity, LLC
FDA DevicesClass II

Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;

The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.

Apr 22, 2026Gentuity, LLC
PharmaceuticalUSFDA Drugs

0.9% Sodium Chloride Injection, USP, 10 mL Ampules, Rx only, Single Dose, Preservative Free, Manufactured for: Spectra Medical Devices, LLC, Wilmington, MA, 01887, By: Houns Co., Ltd, Jecheon, Korea, 27159, NDC 65282-1510-1.

Huons Co., Ltd.
FDA DrugsClass II

0.9% Sodium Chloride Injection, USP, 10 mL Ampules, Rx only, Single Dose, Preservative Free, Manufactured for: Spectra Medical Devices, LLC, Wilmington, MA, 01887, By: Houns Co., Ltd, Jecheon, Korea, 27159, NDC 65282-1510-1.

Lack of Assurance of Sterility

Apr 22, 2026Huons Co., Ltd.
PharmaceuticalUSFDA Drugs

QC Quality Choice, Menthol-Cough Suppressant Oral Anesthetic, Cough Drops, Menthol Flavor, 30-count bag, Distributed By: CDMA Inc., Novi, MI 48375, Made in China, NDC: 83698-675-30, UPC: 635515986718.

Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
FDA DrugsClass II

QC Quality Choice, Menthol-Cough Suppressant Oral Anesthetic, Cough Drops, Menthol Flavor, 30-count bag, Distributed By: CDMA Inc., Novi, MI 48375, Made in China, NDC: 83698-675-30, UPC: 635515986718.

This recall is being initiated following FDA's recommendation based on certain observations noted during an August 15, 2025, inspection of the manufacturing facility that may bear on product quality.

Apr 22, 2026Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
PharmaceuticalUSFDA Drugs

exchange select, Menthol- Cough Suppressant Oral Anesthetic, Cough Drops, Menthol Flavor, 30-count bag, Manufactured for Your Military Exchanges By: Medical Group Care, LLC, 1035 Collier Center Way, STE 5, Naples, FL, Made in China, NDC: 83698-581-30, UPC: 614299398887.

Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
FDA DrugsClass II

exchange select, Menthol- Cough Suppressant Oral Anesthetic, Cough Drops, Menthol Flavor, 30-count bag, Manufactured for Your Military Exchanges By: Medical Group Care, LLC, 1035 Collier Center Way, STE 5, Naples, FL, Made in China, NDC: 83698-581-30, UPC: 614299398887.

This recall is being initiated following FDA's recommendation based on certain observations noted during an August 15, 2025, inspection of the manufacturing facility that may bear on product quality.

Apr 22, 2026Xiamen Kang Zhongyuan Biotechnology Co., Ltd.
Medical DeviceUSFDA Devices

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.

B Braun Medical Inc
FDA DevicesClass I

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Apr 22, 2026B Braun Medical Inc
Medical DeviceUSFDA Devices

B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.

B Braun Medical Inc
FDA DevicesClass I

B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Apr 22, 2026B Braun Medical Inc
Vehicle RecallUSNHTSA

Hexagon Purus Systems USA LLC - Digital Display May Show Incorrect Air Brake Pressure/FMVSS 121

Hexagon Purus Systems USA LLC
NHTSAELECTRICAL SYSTEM

Hexagon Purus Systems USA LLC - Digital Display May Show Incorrect Air Brake Pressure/FMVSS 121

A display that does not show accurate information may fail to alert the driver to an actual brake issue, increasing the risk of a crash.

Apr 21, 2026Hexagon Purus Systems USA LLC