Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 038 NB 6SH; Catalog No.: 10714001; Product/UPN No.: H787107140015 (Box), H787107140010 (Pouch); Box Quantity: 10 pouches;
Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 038 NB 6SH; Catalog No.: 10714001; Product/UPN No.: H787107140015 (Box), H787107140010 (Pouch); Box Quantity: 10 pouches;
The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
triumph - bonneville t100, bonneville t120, street scrambler, street twin motocyclette
triumph - bonneville t100, bonneville t120, street scrambler, street twin motocyclette
une combinaison de températures ambiantes élevées, d'une charge de courant élevée et d'une tension latérale excessive peut amener le câblage entre l'alternateur et le connecteur du harnais à former un pont. si les pales du connecteur forment un pont, un seuil thermique peut être atteint qui dépasse la capacité du connecteur et provoque la déformation du connecteur. cela peut entraîner un court-circuit, augmentant le risque d'incendie.
Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 035 NB 6SH; Catalog No.: 10732301; Product/UPN No.: H787107323015 (Box), H787107323010 (Pouch); Box Quantity: 10 units;
Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 035 NB 6SH; Catalog No.: 10732301; Product/UPN No.: H787107323015 (Box), H787107323010 (Pouch); Box Quantity: 10 units;
The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Medline Kits: 1) NEWBORN ADMISSION, Model Number: DYKB1038; 2) DRAINAGE TRAY, Model Number: DYNDA3188; 3) VA ORLANDO SHOULDER PACK, Model Number: DYNJG001027B; 4) TUMESCENT SYRINGE KIT, Model Number: DYNJTUMSYR; 5) OMC PTCA PACK, Model Number: DYNJT3719; 6) PORT PACK, Model Number: DYNJT5667; 7) WASTE MANAGEMENT KIT - 2, Model Number: DYNJWASTEKIT2; 8) CARDIAC CATH, Model Number: DYNJ20175J; 9) D&C PACK-POB, Model Number: DYNJ33521D; 10) BASIC PACK, Model Number: DYNJ38533F; 11) CATH PACK IMPLANT, Model Number: DYNJ39230D; 12) CERVICAL SPINE PACK, Model Number: DYNJ40886B; 13) MV-IR ANGIO DRAPE PACK-LF, Model Number: DYNJ41553F; 14) TV ENDOVASCULAR-LF, Model Number: DYNJ41558D; 15) MAJOR PACK-LF, Model Number: DYNJ44816J; 16) ANGIOGRAPHY PACK, Model Number: DYNJ48683A; 17) EXTREMITY PACK, Model Number: DYNJ55922C; 18) NEURO PACK, Model Number: DYNJ57785C; 19) PACEMAKER PACK, Model Number: DYNJ61377B; 20) VASCULAR PACK, Model Number: DYNJ62683D; 21) LINQ INSERTION PK W/OUT LIDO, Model Number: DYNJ63668B; 22) BASIC PACK, Model Number: DYNJ83324; 23) OMF MINOR PACK, Model Number: DYNJ83991B; 24) ARTERIOGRAM SUMMIT PACK, Model Number: DYNJ84040B; 25) HMWB NEURO NAMIC PACK, Model Number: DYNJ86600; 26) VCH CARDIAC CATH PACK, Model Number: DYNJ89064; 27) RADIOLOGY PACK, Model Number: DYNJ89080; 28) LITHOTOMY PACK, Model Number: DYNJ89959; 29) AHT SPINE PACK, Model Number: DYNJ903566D; 30) LAP, Model Number: DYNJ904576B; 31) EP MERCY JANESVILLE, Model Number: DYNJ907013; 32) LHK, US, RUBY MEMORIAL HOSP, Model Number: NAM0032; 33) CHERIAN CATH KIT, Model Number: VASC1086; 34) RIGHT HEART KIT - LAKE NONA, Model Number: VASC1120; 35) KIT ANGIO CSTM-SHAWNEE, Model Number: VASC1151; 36) LEFT HEART KIT - OAKBEND, Model Number: VASC1155; 37) IR 2 PORT-ARROWHEAD REGIONAL, Model Number: VASC1404; 38) LHK - ARROWHEAD REGIONAL MED, Model Number: VASC1405; 39) LHK - UMC DOUBLE TRANSDUCER, Model Number: VASC1415A; 40) SDMC, EP TWO PORT CUSTOM KIT, Model Number: VASC1533; 41) ARTERIOGRAM TRAY, Model Number: 00-402001V
Medline Kits: 1) NEWBORN ADMISSION, Model Number: DYKB1038; 2) DRAINAGE TRAY, Model Number: DYNDA3188; 3) VA ORLANDO SHOULDER PACK, Model Number: DYNJG001027B; 4) TUMESCENT SYRINGE KIT, Model Number: DYNJTUMSYR; 5) OMC PTCA PACK, Model Number: DYNJT3719; 6) PORT PACK, Model Number: DYNJT5667; 7) WASTE MANAGEMENT KIT - 2, Model Number: DYNJWASTEKIT2; 8) CARDIAC CATH, Model Number: DYNJ20175J; 9) D&C PACK-POB, Model Number: DYNJ33521D; 10) BASIC PACK, Model Number: DYNJ38533F; 11) CATH PACK IMPLANT, Model Number: DYNJ39230D; 12) CERVICAL SPINE PACK, Model Number: DYNJ40886B; 13) MV-IR ANGIO DRAPE PACK-LF, Model Number: DYNJ41553F; 14) TV ENDOVASCULAR-LF, Model Number: DYNJ41558D; 15) MAJOR PACK-LF, Model Number: DYNJ44816J; 16) ANGIOGRAPHY PACK, Model Number: DYNJ48683A; 17) EXTREMITY PACK, Model Number: DYNJ55922C; 18) NEURO PACK, Model Number: DYNJ57785C; 19) PACEMAKER PACK, Model Number: DYNJ61377B; 20) VASCULAR PACK, Model Number: DYNJ62683D; 21) LINQ INSERTION PK W/OUT LIDO, Model Number: DYNJ63668B; 22) BASIC PACK, Model Number: DYNJ83324; 23) OMF MINOR PACK, Model Number: DYNJ83991B; 24) ARTERIOGRAM SUMMIT PACK, Model Number: DYNJ84040B; 25) HMWB NEURO NAMIC PACK, Model Number: DYNJ86600; 26) VCH CARDIAC CATH PACK, Model Number: DYNJ89064; 27) RADIOLOGY PACK, Model Number: DYNJ89080; 28) LITHOTOMY PACK, Model Number: DYNJ89959; 29) AHT SPINE PACK, Model Number: DYNJ903566D; 30) LAP, Model Number: DYNJ904576B; 31) EP MERCY JANESVILLE, Model Number: DYNJ907013; 32) LHK, US, RUBY MEMORIAL HOSP, Model Number: NAM0032; 33) CHERIAN CATH KIT, Model Number: VASC1086; 34) RIGHT HEART KIT - LAKE NONA, Model Number: VASC1120; 35) KIT ANGIO CSTM-SHAWNEE, Model Number: VASC1151; 36) LEFT HEART KIT - OAKBEND, Model Number: VASC1155; 37) IR 2 PORT-ARROWHEAD REGIONAL, Model Number: VASC1404; 38) LHK - ARROWHEAD REGIONAL MED, Model Number: VASC1405; 39) LHK - UMC DOUBLE TRANSDUCER, Model Number: VASC1415A; 40) SDMC, EP TWO PORT CUSTOM KIT, Model Number: VASC1533; 41) ARTERIOGRAM TRAY, Model Number: 00-402001V
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
Avitene Ultrafoam Microfibrillar Collagen Hemostat - 8cm x 6.25cm x 1cm. Catalog Number: 1050030.
Avitene Ultrafoam Microfibrillar Collagen Hemostat - 8cm x 6.25cm x 1cm. Catalog Number: 1050030.
Potential for product to contain foreign matter, confirmed to be inspect fragments.
Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.
Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.
Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.
Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08
Diversatek Healthcare Viper¿ 3-Stage Fixed Wire Balloon Dilator, 6mm-7mm-8mm, REF 1204-08
The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2138.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2138.
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
Medline Namic Syringes: 1) SYR CNTRL 10ML RING W/ROT NB, Model Number: DYNJSYR10CWRB;
Medline Namic Syringes: 1) SYR CNTRL 10ML RING W/ROT NB, Model Number: DYNJSYR10CWRB;
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BIOPSY PACK DYNJ38587A, DYNJ38587B BIOPSY/DRAINAGE TRAY-LF DYNJ30078D W LINQ PACK DYNJ59108B
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BIOPSY PACK DYNJ38587A, DYNJ38587B BIOPSY/DRAINAGE TRAY-LF DYNJ30078D W LINQ PACK DYNJ59108B
Unapproved design changes to the products outside of the 510(k) clearance.
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC.
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC.
Potential for detachment of a distal tip component of the device during use.
Soft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Quantity: 5 pouches;
Soft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Quantity: 5 pouches;
The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Arrow Echogenic Introducer Needle REF ASK-04001-BWK2 UDI code: (01)10801902220315(17)261231(11)251210(10)33F25J0347 This device is intended for injection or aspiration of fluids. The needle protection device covers the needle after use to help prevent needle sticks.
Arrow Echogenic Introducer Needle REF ASK-04001-BWK2 UDI code: (01)10801902220315(17)261231(11)251210(10)33F25J0347 This device is intended for injection or aspiration of fluids. The needle protection device covers the needle after use to help prevent needle sticks.
Due to receiving a notice from a supplier that the liquid adhesive were incorrectly manufactured
Medline Kits: 1) SYR CNTRL 10ML RING W/ROT NB, Model Number: DYNJSYR10CWRB; 2) MANIFOLD KIT LOW PRESSURE OFF, Model Number: VASCSLPOFF1; 3) EP MANIFOLD KIT, Model Number: VASC1018A
Medline Kits: 1) SYR CNTRL 10ML RING W/ROT NB, Model Number: DYNJSYR10CWRB; 2) MANIFOLD KIT LOW PRESSURE OFF, Model Number: VASCSLPOFF1; 3) EP MANIFOLD KIT, Model Number: VASC1018A
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Model Number: TB2-0535FC.
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Model Number: TB2-0535FC.
Potential for detachment of a distal tip component of the device during use.
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes E P DRAPE PACK-LF DYNJ0373061J
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes E P DRAPE PACK-LF DYNJ0373061J
Unapproved design changes to the products outside of the 510(k) clearance.
Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 15mm-16.5mm-18mm, REF 1205-18
Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 15mm-16.5mm-18mm, REF 1205-18
The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.
Medline Syringes: 1) SYRINGE 12ML CONTROL W/ROTATOR, Model Number: DYNJSYR12CWR; 2) SYRINGE 10ML CONTROL W/ROTATOR, Model Number: DYNJSYR10CWR; 3) SYR CONTROL PAD 12ML L/L, Model Number: DYNJSYR12CP; 4) TBG PRESS 72IN M/F ROT BRAID, Model Number: DYNJBRT72R; 5) SYRINGE 10ML CONTROL MLL, Model Number: DYNJSYR10C; 6) SYRINGE 8ML CONTROL MLL, Model Number: DYNJSYR8C; 7) SYRINGE 12ML CONTROL MLL, Model Number: DYNJSYR12C; 8) SYR CONTROL PAD 10ML L/L ROT, Model Number: DYNJSYR10CRP; 9) SYR CONTROL PAD 12ML L/L ROT, Model Number: DYNJSYR12CRP; 10) CTRL SYRINGE 6ML RING ROT L/L, Model Number: DYNJSYRC6CWR
Medline Syringes: 1) SYRINGE 12ML CONTROL W/ROTATOR, Model Number: DYNJSYR12CWR; 2) SYRINGE 10ML CONTROL W/ROTATOR, Model Number: DYNJSYR10CWR; 3) SYR CONTROL PAD 12ML L/L, Model Number: DYNJSYR12CP; 4) TBG PRESS 72IN M/F ROT BRAID, Model Number: DYNJBRT72R; 5) SYRINGE 10ML CONTROL MLL, Model Number: DYNJSYR10C; 6) SYRINGE 8ML CONTROL MLL, Model Number: DYNJSYR8C; 7) SYRINGE 12ML CONTROL MLL, Model Number: DYNJSYR12C; 8) SYR CONTROL PAD 10ML L/L ROT, Model Number: DYNJSYR10CRP; 9) SYR CONTROL PAD 12ML L/L ROT, Model Number: DYNJSYR12CRP; 10) CTRL SYRINGE 6ML RING ROT L/L, Model Number: DYNJSYRC6CWR
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2114.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2114.
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
Soft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014; Product/UPN No.: H787107140145 (Box), H787107140140 (Pouch); Box Quantity: 5 pouches;
Soft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014; Product/UPN No.: H787107140145 (Box), H787107140140 (Pouch); Box Quantity: 5 pouches;
The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EXTREMITY PACK DYNJ34846C HAND PACK DYNJ83218A, DYNJ85396 LAMINECTOMY DYNJ905156R MAJOR EXTREMITY DYNJ905159L, DYNJ905159M MAJOR EXTREMITY PACK DYNJ69926A PODIATRY PACK DYNJ41250B
Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EXTREMITY PACK DYNJ34846C HAND PACK DYNJ83218A, DYNJ85396 LAMINECTOMY DYNJ905156R MAJOR EXTREMITY DYNJ905159L, DYNJ905159M MAJOR EXTREMITY PACK DYNJ69926A PODIATRY PACK DYNJ41250B
Unapproved design changes to the products outside of the 510(k) clearance.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.