PharmaceuticalUSFDA Drugs

Max Packaging Antibacterial Towelette (Benzalkonium Chloride 0.13%), 1 towelette, 4x7, Max Packaging, Made in the USA, NDC 59647-750-01

ACME UNITED CORPORATION
FDA DrugsClass II

Max Packaging Antibacterial Towelette (Benzalkonium Chloride 0.13%), 1 towelette, 4x7, Max Packaging, Made in the USA, NDC 59647-750-01

CGMP Deviations

Mar 11, 2026ACME UNITED CORPORATION
Medical DeviceUSFDA Devices

Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

ESAOTE S.P.A.
FDA DevicesClass II

Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.

Mar 11, 2026ESAOTE S.P.A.
Vehicle RecallUSNHTSA

ARBOC Specialty Vehicles, LLC - Loss of Rear ABS System

ARBOC Specialty Vehicles, LLC
NHTSASERVICE BRAKES,...

ARBOC Specialty Vehicles, LLC - Loss of Rear ABS System

Loss of the rear ABS system can increase the risk of a crash.

Mar 11, 2026ARBOC Specialty Vehicles, LLC
Medical DeviceUSFDA Devices

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological

GE Healthcare
FDA DevicesClass II

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological

Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,

Mar 11, 2026GE Healthcare
PharmaceuticalUSFDA Drugs

Metoprolol Succinate Extended-Release Tablets, 50 mg, 100 tablets bottles (NDC: 45963-676-11) and 1,000 tablets bottles (NDC: 45963-676-96), Rx Only, Teva Pharmaceuticals, Parsippany, NJ 07054

Teva Pharmaceuticals USA, Inc
FDA DrugsClass II

Metoprolol Succinate Extended-Release Tablets, 50 mg, 100 tablets bottles (NDC: 45963-676-11) and 1,000 tablets bottles (NDC: 45963-676-96), Rx Only, Teva Pharmaceuticals, Parsippany, NJ 07054

Failed Dissolution Specifications

Mar 11, 2026Teva Pharmaceuticals USA, Inc
Medical DeviceUSFDA Devices

CS100 IABP. Software Version CS100 IABP Q.01.

Datascope Corp.
FDA DevicesClass II

CS100 IABP. Software Version CS100 IABP Q.01.

The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.

Mar 11, 2026Datascope Corp.
Medical DeviceUSFDA Devices

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Staar Surgical AG
FDA DevicesClass II

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Mar 11, 2026Staar Surgical AG
Food & BeverageUSFDA Food

Junebar Peanut Chocolate Chip All Natural Snack Bar; INGREDIENTS: ORGANIC PEANUT BUTTER, DATE PASTE, ORGANIC BLACK BEANS, ORGANIC SWEET POTATOES, GLUTEN FREE ROLLED OATS, SEMI-SWEET SOY FREE VEGAN SHOCOLATE CHIPS (ORGANIC CANE SUGAR, UNSWEETENED CHOCOLATE, COCOA BUTTER), PEANUTS, OAT FLOUR, VANILLA EXTRACT (VANILLA BEAN EXTRATIVES IN WATER, ALCOHOL), FLAX, CINNAMON, SEA SALT, BAKING POWDER; ALLERGEN WARNING: CONTAINS PEANUTS, TREE NUTS (ALMOND, COCONUT), PRODUCED ON EQUIPMENT ALSO HANDLING PEANUT, SESAME, PRODUCED IN A FACILITY THAT ALSO PROCESSES GLUTEN PRODUCTS. KEEP REFRIGERATED.; UPC 8 60012 38461 7

Juniper Granola, LLC
FDA FoodClass I

Junebar Peanut Chocolate Chip All Natural Snack Bar; INGREDIENTS: ORGANIC PEANUT BUTTER, DATE PASTE, ORGANIC BLACK BEANS, ORGANIC SWEET POTATOES, GLUTEN FREE ROLLED OATS, SEMI-SWEET SOY FREE VEGAN SHOCOLATE CHIPS (ORGANIC CANE SUGAR, UNSWEETENED CHOCOLATE, COCOA BUTTER), PEANUTS, OAT FLOUR, VANILLA EXTRACT (VANILLA BEAN EXTRATIVES IN WATER, ALCOHOL), FLAX, CINNAMON, SEA SALT, BAKING POWDER; ALLERGEN WARNING: CONTAINS PEANUTS, TREE NUTS (ALMOND, COCONUT), PRODUCED ON EQUIPMENT ALSO HANDLING PEANUT, SESAME, PRODUCED IN A FACILITY THAT ALSO PROCESSES GLUTEN PRODUCTS. KEEP REFRIGERATED.; UPC 8 60012 38461 7

Products contain undeclared milk and soy.

Mar 11, 2026Juniper Granola, LLC
PharmaceuticalUSFDA Drugs

Med Nap BZK Antiseptic Towelettes (Benzalkonium Chloride 0.13%), 100 Towelettes, Manufactured by: Acme United Corporation, 1 Waterview Dr, Shelton, CT 06481, NDC 0924-0246-01.

ACME UNITED CORPORATION
FDA DrugsClass II

Med Nap BZK Antiseptic Towelettes (Benzalkonium Chloride 0.13%), 100 Towelettes, Manufactured by: Acme United Corporation, 1 Waterview Dr, Shelton, CT 06481, NDC 0924-0246-01.

CGMP Deviations

Mar 11, 2026ACME UNITED CORPORATION
Medical DeviceUSFDA Devices

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A

Aesculap Inc
FDA DevicesClass II

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A

There is the potential for the length of the trocar shaft to be too long.

Mar 11, 2026Aesculap Inc
Medical DeviceUSFDA Devices

Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

Mar 11, 2026Olympus Corporation of the Americas
Vehicle RecallUSNHTSA

Volkswagen Group of America, Inc. - Loose Transmission Ground Wire May Cause Fire

Volkswagen Group of America, Inc.
NHTSAELECTRICAL SYSTEM

Volkswagen Group of America, Inc. - Loose Transmission Ground Wire May Cause Fire

An open circuit may draw a higher current and increase the risk of a fire.

Mar 11, 2026Volkswagen Group of America, Inc.
Medical DeviceUSFDA Devices

Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

Mar 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL

Roche Diagnostics Operations, Inc.
FDA DevicesClass II

cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL

Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

Mar 11, 2026Roche Diagnostics Operations, Inc.
Medical DeviceUSFDA Devices

Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.

Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.

Mar 11, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable

Diagnostica Stago, Inc.
FDA DevicesClass II

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable

The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

Mar 11, 2026Diagnostica Stago, Inc.
PharmaceuticalUSFDA Drugs

WP WipesPlus Hand Sanitizing Wipes (Benzalkonium Chloride 0.13%), 5x7, 100 CT, Progressive Products, 211 S. Ridge Street, Rye Brook, NY 10573. NDC 67151-727-01.

ACME UNITED CORPORATION
FDA DrugsClass II

WP WipesPlus Hand Sanitizing Wipes (Benzalkonium Chloride 0.13%), 5x7, 100 CT, Progressive Products, 211 S. Ridge Street, Rye Brook, NY 10573. NDC 67151-727-01.

CGMP Deviations

Mar 11, 2026ACME UNITED CORPORATION
Medical DeviceUSFDA Devices

Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A

B Braun Medical Inc
FDA DevicesClass II

Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A

The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.

Mar 11, 2026B Braun Medical Inc
PharmaceuticalUSFDA Drugs

SILINTAN 25/pills, packaged in a 25-count bottle, Shanghai Chinese Medical Works, Shanghai, China

ANTHONY TRINH, 123Herbals LLC
FDA DrugsClass I

SILINTAN 25/pills, packaged in a 25-count bottle, Shanghai Chinese Medical Works, Shanghai, China

Marketed Without an Approved NDA/ANDA: FDA analysis revealed the presence of undeclared meloxicam.

Mar 11, 2026ANTHONY TRINH, 123Herbals LLC
Medical DeviceUSFDA Devices

Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.

Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

Mar 11, 2026Olympus Corporation of the Americas
Food & BeverageUSFDA Food

Premium Food Jordan Almonds, packed in 2.0 oz and 3.5 oz clear flexible plastic packaging.

Bader Enterprises, Inc.
FDA FoodClass II

Premium Food Jordan Almonds, packed in 2.0 oz and 3.5 oz clear flexible plastic packaging.

Contains undeclared allergen (wheat) and colors (FD&C Blue 1 Lake and FD&C Blue 2 Lake).

Mar 11, 2026Bader Enterprises, Inc.
Medical DeviceUSFDA Devices

Brand Name: HiResolution Bionic Ear System Product Name: Sky CI M90 Sound Processor Model/Catalog Number: CI-5295-150

Advanced Bionics, LLC
FDA DevicesClass II

Brand Name: HiResolution Bionic Ear System Product Name: Sky CI M90 Sound Processor Model/Catalog Number: CI-5295-150

Behind-the-ear sound processer packaging label is different then included product.

Mar 11, 2026Advanced Bionics, LLC
Vehicle RecallUSNHTSA

Volvo Car USA, LLC - Seat Belt Warning System Malfunction/FMVSS 208

Volvo Car USA, LLC
NHTSASEAT BELTS

Volvo Car USA, LLC - Seat Belt Warning System Malfunction/FMVSS 208

An audible warning chime that does not properly alert of an unbelted seat belt can increase the risk of injury during a crash.

Mar 11, 2026Volvo Car USA, LLC
Medical DeviceUSFDA Devices

Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A

B Braun Medical Inc
FDA DevicesClass II

Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A

The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.

Mar 11, 2026B Braun Medical Inc