Dispositivos Médicos Recalls

Defective surgical instruments, diagnostic equipment, and health device recalls.

Showing 24 of 2,119 results
Medical DeviceUSFDA Devices

See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182;

Medline Industries, LP
FDA DevicesClass II

See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182;

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

One Step 10A in vitro diagnostic test

DFI Co., Ltd.
FDA DevicesClass II

One Step 10A in vitro diagnostic test

The devices were distributed without required FDA premarket clearance or approval.

May 13, 2026DFI Co., Ltd.
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222

Erbe Medical, LLC
FDA DevicesClass II

ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

May 13, 2026Erbe Medical, LLC
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W;

Medline Industries, LP
FDA DevicesClass II

See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W;

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

DEX Ophthalmic Tissue Forceps, 25ga DEX Nano Tapered AWH Forceps, Model/Catalog Number: DVF4005-25

Katalyst Surgical, LLC
FDA DevicesClass II

DEX Ophthalmic Tissue Forceps, 25ga DEX Nano Tapered AWH Forceps, Model/Catalog Number: DVF4005-25

Field Safety Corrective Action for IFU in DEX Forceps and Scissors.

May 13, 2026Katalyst Surgical, LLC
Medical DeviceUSFDA Devices

One Step UTI in vitro diagnostic test REF: 3374

DFI Co., Ltd.
FDA DevicesClass II

One Step UTI in vitro diagnostic test REF: 3374

The devices were distributed without required FDA premarket clearance or approval.

May 13, 2026DFI Co., Ltd.
Medical DeviceUSFDA Devices

Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). The choice of catheter material for different angiographic procedures should be based on the physician s experience.

Cook Incorporated
FDA DevicesClass I

Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., polyethylene or nylon, non-braided or braided with 1:1 torque). The choice of catheter material for different angiographic procedures should be based on the physician s experience.

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

May 13, 2026Cook Incorporated
Medical DeviceUSFDA Devices

One Step pH in vitro diagnostic test REF: 31I4P

DFI Co., Ltd.
FDA DevicesClass II

One Step pH in vitro diagnostic test REF: 31I4P

The devices were distributed without required FDA premarket clearance or approval.

May 13, 2026DFI Co., Ltd.
Medical DeviceUSFDA Devices

ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

Erbe Medical, LLC
FDA DevicesClass II

ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

May 13, 2026Erbe Medical, LLC
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.

BioFire Diagnostics, LLC
FDA DevicesClass II

BioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.

Contamination to in-vitro diagnostic test may result in false positives.

May 13, 2026BioFire Diagnostics, LLC
Medical DeviceUSFDA Devices

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Cook Incorporated
FDA DevicesClass I

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

May 13, 2026Cook Incorporated
Medical DeviceUSFDA Devices

Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

Mint Medical GmbH
FDA DevicesClass II

Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.

May 13, 2026Mint Medical GmbH
Medical DeviceUSFDA Devices

One Step K in vitro diagnostic test REF: 81A4

DFI Co., Ltd.
FDA DevicesClass II

One Step K in vitro diagnostic test REF: 81A4

The devices were distributed without required FDA premarket clearance or approval.

May 13, 2026DFI Co., Ltd.
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.

Centinel Spine, Inc.
FDA DevicesClass II

Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.

Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.

May 13, 2026Centinel Spine, Inc.
Medical DeviceUSFDA Devices

Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807K; 5) D&C/HYSTEROSCOPY PACK, Model Number: DYNJ0547084O; 6) HYSTER/RESECTOSCOPE #78-RF, Model Number: DYNJ21824R; 7) MINOR VAGINAL #76-RF, Model Number: DYNJ27434R; 8) VAG HYST PACK, Model Number: DYNJ31344J; 9) VAG HYST PACK, Model Number: DYNJ31344K; 10) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393L; 11) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393M; 12) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393N; 13) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393O; 14) LAVH PACK, Model Number: DYNJ36712F; 15) D&C MINOR LITHOTOMY PACK, Model Number: DYNJ45090I; 16) D C HYSTEROSCOPY PACK-LF, Model Number: DYNJ45801C; 17) VAG HYST PACK (VHGSG)642-LF, Model Number: DYNJ47694I; 18) HYSTEROSCOPY D AND C PACK, Model Number: DYNJ51741B; 19) LAVH - N #668629 -N, Model Number: DYNJ58121C; 20) PK, GYN-MINOR-LITHO, Model Number: DYNJ59037B; 21) HYSTEROSCOPY PACK, Model Number: DYNJ59444G; 22) D AND C PACK, Model Number: DYNJ61282B; 23) VAGINAL HYSTERECTOMY, Model Number: DYNJ67161; 24) PERI GYN, Model Number: DYNJ67708A; 25) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713B; 26) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713C; 27) ANORECTAL PACK, Model Number: DYNJ83560; 28) D&C CDS, Model Number: DYNJ902560I; 29) HYSTEROSCOPY, Model Number: DYNJ903327K; 30) HYSTEROSCOPY, Model Number: DYNJ903475I; 31) ROBOTIC HYST, Model Number: DYNJ903798G; 32) TLH KIT, Model Number: DYNJ908149; 33) LAVH TOTE, Model Number: DYNJ908340; 34) LAVH TOTE, Model Number: DYNJ908340A; 35) LAVH TOTE, Model Number: DYNJ908340B; 36) LAVH TOTE, Model Number: DYNJ908340D; 37) LAVH, Model Number: DYNJ908986; 38) LAVH, Model Number: DYNJ908986A; 39) BRECKENRIDGE LAVH, Model Number: DYNJ910537; 40) LAVH GYN/ONC, Model Number: DYNJ910927; 41) VAG HYST PACK, Model Number: DYNJT3348; 42) LAVH PART B, Model Number: DYNJV0312G

Medline Industries, LP
FDA DevicesClass II

Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807K; 5) D&C/HYSTEROSCOPY PACK, Model Number: DYNJ0547084O; 6) HYSTER/RESECTOSCOPE #78-RF, Model Number: DYNJ21824R; 7) MINOR VAGINAL #76-RF, Model Number: DYNJ27434R; 8) VAG HYST PACK, Model Number: DYNJ31344J; 9) VAG HYST PACK, Model Number: DYNJ31344K; 10) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393L; 11) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393M; 12) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393N; 13) RRMC VAGINAL HYSTERECTOMY PACK, Model Number: DYNJ33393O; 14) LAVH PACK, Model Number: DYNJ36712F; 15) D&C MINOR LITHOTOMY PACK, Model Number: DYNJ45090I; 16) D C HYSTEROSCOPY PACK-LF, Model Number: DYNJ45801C; 17) VAG HYST PACK (VHGSG)642-LF, Model Number: DYNJ47694I; 18) HYSTEROSCOPY D AND C PACK, Model Number: DYNJ51741B; 19) LAVH - N #668629 -N, Model Number: DYNJ58121C; 20) PK, GYN-MINOR-LITHO, Model Number: DYNJ59037B; 21) HYSTEROSCOPY PACK, Model Number: DYNJ59444G; 22) D AND C PACK, Model Number: DYNJ61282B; 23) VAGINAL HYSTERECTOMY, Model Number: DYNJ67161; 24) PERI GYN, Model Number: DYNJ67708A; 25) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713B; 26) ROBOTIC HYSTERECTOMY PACK, Model Number: DYNJ69713C; 27) ANORECTAL PACK, Model Number: DYNJ83560; 28) D&C CDS, Model Number: DYNJ902560I; 29) HYSTEROSCOPY, Model Number: DYNJ903327K; 30) HYSTEROSCOPY, Model Number: DYNJ903475I; 31) ROBOTIC HYST, Model Number: DYNJ903798G; 32) TLH KIT, Model Number: DYNJ908149; 33) LAVH TOTE, Model Number: DYNJ908340; 34) LAVH TOTE, Model Number: DYNJ908340A; 35) LAVH TOTE, Model Number: DYNJ908340B; 36) LAVH TOTE, Model Number: DYNJ908340D; 37) LAVH, Model Number: DYNJ908986; 38) LAVH, Model Number: DYNJ908986A; 39) BRECKENRIDGE LAVH, Model Number: DYNJ910537; 40) LAVH GYN/ONC, Model Number: DYNJ910927; 41) VAG HYST PACK, Model Number: DYNJT3348; 42) LAVH PART B, Model Number: DYNJV0312G

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

May 13, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

DEX Ophthalmic Tissue Forceps, 25ga DEX Stiff Maculorhexis Forceps, Model/Catalog Number: DVF4019-25-S

Katalyst Surgical, LLC
FDA DevicesClass II

DEX Ophthalmic Tissue Forceps, 25ga DEX Stiff Maculorhexis Forceps, Model/Catalog Number: DVF4019-25-S

Field Safety Corrective Action for IFU in DEX Forceps and Scissors.

May 13, 2026Katalyst Surgical, LLC
Medical DeviceUSFDA Devices

Uric Acid in vitro diagnostic test REF: 31H0P

DFI Co., Ltd.
FDA DevicesClass II

Uric Acid in vitro diagnostic test REF: 31H0P

The devices were distributed without required FDA premarket clearance or approval.

May 13, 2026DFI Co., Ltd.

Install The Recall File App