Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Medtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter
Medtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter
Certain lots of product have the potential for a sterile barrier breach.
Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter
Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter
Certain lots of product have the potential for a sterile barrier breach.
Dexcom G7 Sensors
Dexcom G7 Sensors
AA-Med Pty Ltd is recalling two lots of Dexcom G7 sensors. These lots were identified for disposal and destruction in the USA due to incomplete sterilisation and deployment failure during testing....
Field Safety Notices: 11 to 15 May 2026
Field Safety Notices: 11 to 15 May 2026
List of Field Safety Notices from 11 to 15 May 2026....
Kimal Procedure Packs containing recalled components: Namic Angiographic Syringe with the risk of syringe disconnection; Namic Manifolds with the risk of foreign particulates. Important guidance for use in urgent procedures where there are no alternatives. (DSI/2026/002)
Kimal Procedure Packs containing recalled components: Namic Angiographic Syringe with the risk of syringe disconnection; Namic Manifolds with the risk of foreign particulates. Important guidance for use in urgent procedures where there are no alternatives. (DSI/2026/002)
Medline Industries LP and Kimal PLC have identified safety issues with the Namic Angiographic Control Syringes with RA and Namic White Star Off Handle Manifold devices, and have issued field safety no...
Medtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Medtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER
BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Medline has identified the presence of particulate within the fluid path of the Manifolds.
Namic Preceptor Manifold, custom, Medline Product Number/SKU 64037107; Extravascular blood pressure transducer
Namic Preceptor Manifold, custom, Medline Product Number/SKU 64037107; Extravascular blood pressure transducer
Medline has identified the presence of particulate within the fluid path of the Manifolds.
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3¿ cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan¿ helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: No Product Description: 11-1424 - ICEMAN W/,UNIV,LOOP,NS,EH The IceMan CLASSIC3¿ cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan¿ helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 LUCAS 3.1 Model Number: 99576 Catalog Number: 99576-000063 UDI-DI code: 00883873834329 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 LUCAS 3.1 Model Number: 99576 Catalog Number: 99576-000063 UDI-DI code: 00883873834329 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
Due to a subset of chest compression systems/devices where the service performed on the systems were not appropriately documented (The documents for the service performed to justify the passing results were not include in the service records).
On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33
On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33
Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications.
Medline medical convenience kits, containing Namic Manifold, Labeled as: HEART CATH PACK-LF, Medline Kit Number/SKU DYNJ36478B
Medline medical convenience kits, containing Namic Manifold, Labeled as: HEART CATH PACK-LF, Medline Kit Number/SKU DYNJ36478B
Medline has identified the presence of particulate within the fluid path of the Manifolds.
Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27
Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used are the same for each user. The watchOS app cannot be installed independently from the iOS phone app. Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used are the same for each user. The watchOS app cannot be installed independently from the iOS phone app. Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Daig Livewire Steerable, Product Number 401904; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Daig Livewire Steerable, Product Number 401904; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-15552-VFIE; Arrow-Clark VectorFlow Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: REF: CS-15192-VFIM, CS-15232-VFIM, CS-15272-VFIM; Arrow-Clark VectorFlow Retrograde Chronic Hemodialysis Catheter, REF: ACS-15192-VF, ACS-15232-VF, ACS-15272-VF, ACS-15552-VF, CS-15192-VF, CS-15192-VFE, CS-15232-VF, CS-15232-VFE, CS-15272-VF, CS-15272-VFE, CS-15312-VF, CS-15312-VFE, CS-15422-VF, CS-15422-VFE, CS-15552-VF, CS-15552-VFE; Arrow-Clark VectorFlow Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: CS-15192-VFM, CS-15232-VFM, CS-15272-VFM
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-15552-VFIE; Arrow-Clark VectorFlow Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: REF: CS-15192-VFIM, CS-15232-VFIM, CS-15272-VFIM; Arrow-Clark VectorFlow Retrograde Chronic Hemodialysis Catheter, REF: ACS-15192-VF, ACS-15232-VF, ACS-15272-VF, ACS-15552-VF, CS-15192-VF, CS-15192-VFE, CS-15232-VF, CS-15232-VFE, CS-15272-VF, CS-15272-VFE, CS-15312-VF, CS-15312-VFE, CS-15422-VF, CS-15422-VFE, CS-15552-VF, CS-15552-VFE; Arrow-Clark VectorFlow Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: CS-15192-VFM, CS-15232-VFM, CS-15272-VFM
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.
QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.
Product has the potential for false positive results