Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.
MEDLINE medical procedure convenience kit labeled as: 1) TOTAL KNEE-LF, Medline kit SKU CDSOPT590T; 2) TOTAL KNEE-LF, Medline kit SKU CDSOPT590U; 3) TOTAL KNEE CDS, Medline kit SKU CDS983930J; 4) HIP CDS-LF, Medline kit SKU CDS984556L; 5) HIP CDS-LF, Medline kit SKU CDS984556M; 6) TOTAL KNEE PACK, Medline kit SKU DYNJ905162Q.
The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Labeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.
Penumbra System RED 68 Reperfusion Catheter REF: RED68KIT
Penumbra System RED 68 Reperfusion Catheter REF: RED68KIT
Due to manufacturing issues, reperfusion catheters may fractures
Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.
Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.
Defects in the silicone seal, which may result in "doming" of the silicone seal due to improper sealing, potentially leading to fluid occlusion or leakage.
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
FARASTAR Ablation Generator EU, Pulsed Field Ablation Generator, Product ID 61M401; Percutaneous Cardiac Ablation Catheter
Boston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.
FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062
FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
MEDLINE convenience kits labeled as: 1) TOTAL JOINT KIT, Medline Kit SKU DYKS1574A; 2) TOTAL KNEE BASIC, Medline Kit SKU DYNJ902378J; 3) TOTAL HIP, Medline Kit SKU DYNJ903492K; 4) ANTERIOR HIP, Medline Kit SKU DYNJ903493I; 5) ANTERIOR HIP, Medline Kit SKU DYNJ903493J; 6) ANTERIOR HIP, Medline Kit SKU DYNJ903493K; 7) TOTAL HIP, Medline Kit SKU DYNJ906665G; 8) TOTAL KNEE, Medline Kit SKU DYNJ906666J; 9) TOTAL HIP-LF, Medline Kit SKU DYNJ909000A; 10) TOTAL KNEE-LF, Medline Kit SKU DYNJ909002A; 11) TOTAL HIP W NAVIGATION NO SYR, Medline Kit SKU DYNJ909068B; 12) TOTAL JOINT, Medline Kit SKU DYNJ909069B; 13) TOTAL KNEE, Medline Kit SKU DYNJ909070C; 14) TOTAL KNEE W NAVIGATION NO SYR, Medline Kit SKU DYNJ909071B; 15) TOTAL KNEE W NAVIGATION, Medline Kit SKU DYNJ909071C; 16) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534; 17) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534A; 18) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534B; 19) ANTERIOR HIP, Medline Kit SKU DYNJ909640F; 20) POSTERIOR HIP, Medline Kit SKU DYNJ909642D; 21) TOTAL KNEE, Medline Kit SKU DYNJ909643D; 22) TOTAL KNEE, Medline Kit SKU DYNJ909643F; 23) TOTAL JOINT, Medline Kit SKU DYNJ910002A; 24) TOTAL KNEE, Medline Kit SKU DYNJ910354A; 25) TOTAL HIP, Medline Kit SKU DYNJ910813; 26) ANTERIOR HIP, Medline Kit SKU DYNJ911083.
MEDLINE convenience kits labeled as: 1) TOTAL JOINT KIT, Medline Kit SKU DYKS1574A; 2) TOTAL KNEE BASIC, Medline Kit SKU DYNJ902378J; 3) TOTAL HIP, Medline Kit SKU DYNJ903492K; 4) ANTERIOR HIP, Medline Kit SKU DYNJ903493I; 5) ANTERIOR HIP, Medline Kit SKU DYNJ903493J; 6) ANTERIOR HIP, Medline Kit SKU DYNJ903493K; 7) TOTAL HIP, Medline Kit SKU DYNJ906665G; 8) TOTAL KNEE, Medline Kit SKU DYNJ906666J; 9) TOTAL HIP-LF, Medline Kit SKU DYNJ909000A; 10) TOTAL KNEE-LF, Medline Kit SKU DYNJ909002A; 11) TOTAL HIP W NAVIGATION NO SYR, Medline Kit SKU DYNJ909068B; 12) TOTAL JOINT, Medline Kit SKU DYNJ909069B; 13) TOTAL KNEE, Medline Kit SKU DYNJ909070C; 14) TOTAL KNEE W NAVIGATION NO SYR, Medline Kit SKU DYNJ909071B; 15) TOTAL KNEE W NAVIGATION, Medline Kit SKU DYNJ909071C; 16) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534; 17) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534A; 18) MHCZ TOTAL KNEE, Medline Kit SKU DYNJ909534B; 19) ANTERIOR HIP, Medline Kit SKU DYNJ909640F; 20) POSTERIOR HIP, Medline Kit SKU DYNJ909642D; 21) TOTAL KNEE, Medline Kit SKU DYNJ909643D; 22) TOTAL KNEE, Medline Kit SKU DYNJ909643F; 23) TOTAL JOINT, Medline Kit SKU DYNJ910002A; 24) TOTAL KNEE, Medline Kit SKU DYNJ910354A; 25) TOTAL HIP, Medline Kit SKU DYNJ910813; 26) ANTERIOR HIP, Medline Kit SKU DYNJ911083.
Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. Huons Co, Ltd. has issued a voluntary recall of this product due to quality issues.
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P
Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and mild to moderate obstructive sleep apnea (OSA).
Due to device not having FDA Clearance
3mensio Workstation (Vascular Fenestrated) software
3mensio Workstation (Vascular Fenestrated) software
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
SeaSpine Polyaxial Head REF Number: 41-3010
SeaSpine Polyaxial Head REF Number: 41-3010
Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. SM-LAPAROSCOPIC PROSTATE PAC DYNJ61206J
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. SM-LAPAROSCOPIC PROSTATE PAC DYNJ61206J
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4
Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
MEDLINE medical procedure convenience kit labeled as: HIP CDS-LF, Medline kit SKU CDS984556N
MEDLINE medical procedure convenience kit labeled as: HIP CDS-LF, Medline kit SKU CDS984556N
The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.