Safety Recall Directory
Browse and filter the latest official safety alerts globally.
Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS840402AI; 2. OPEN HEART CDS, Medline Kit Number/SKU CDS984890V; 3. RR-NEURO VASCULAR PACK-LF, Medline Kit Number/SKU DYNJ0394874L; 4. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993M; 5. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933G; 6. FESS-MB PACK-LF, Medline Kit Number/SKU DYNJ0888159I; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861C; 8. POSTERIOR LUMBAR PACK, Medline Kit Number/SKU DYNJ42626B; 9. CV CUTDOWN PACK, Medline Kit Number/SKU DYNJ45374F; 10. X RAY CLOSED PACK, Medline Kit Number/SKU DYNJ48949C; 11. RFD- AAA PACK, Medline Kit Number/SKU DYNJ54242B; 12. NEURO LAM, Medline Kit Number/SKU DYNJ64188A; 13. LIVE OAK SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ69223; 14. FEM POP PACK, Medline Kit Number/SKU DYNJ81423C; 15. ENS FREEFLAP PROCDURE - LF, Medline Kit Number/SKU DYNJ88249; 16. OPEN HEART A&B, Medline Kit Number/SKU DYNJ901203D; 17. NEURO ACD, Medline Kit Number/SKU DYNJ904549F; 18. KIT VP SHUNT RFD, Medline Kit Number/SKU DYNJ906303F; 19. ACDF-LF, Medline Kit Number/SKU DYNJ906681C; 20. ACDF-LF, Medline Kit Number/SKU DYNJ906681D; 21. ACDF-LF, Medline Kit Number/SKU DYNJ906681F; 22. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196B; 23. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196C; 24. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ908362F; 25. VP SHUNT, Medline Kit Number/SKU DYNJ909379; 26. VP SHUNT, Medline Kit Number/SKU DYNJ909379A.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS840402AI; 2. OPEN HEART CDS, Medline Kit Number/SKU CDS984890V; 3. RR-NEURO VASCULAR PACK-LF, Medline Kit Number/SKU DYNJ0394874L; 4. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993M; 5. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933G; 6. FESS-MB PACK-LF, Medline Kit Number/SKU DYNJ0888159I; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861C; 8. POSTERIOR LUMBAR PACK, Medline Kit Number/SKU DYNJ42626B; 9. CV CUTDOWN PACK, Medline Kit Number/SKU DYNJ45374F; 10. X RAY CLOSED PACK, Medline Kit Number/SKU DYNJ48949C; 11. RFD- AAA PACK, Medline Kit Number/SKU DYNJ54242B; 12. NEURO LAM, Medline Kit Number/SKU DYNJ64188A; 13. LIVE OAK SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ69223; 14. FEM POP PACK, Medline Kit Number/SKU DYNJ81423C; 15. ENS FREEFLAP PROCDURE - LF, Medline Kit Number/SKU DYNJ88249; 16. OPEN HEART A&B, Medline Kit Number/SKU DYNJ901203D; 17. NEURO ACD, Medline Kit Number/SKU DYNJ904549F; 18. KIT VP SHUNT RFD, Medline Kit Number/SKU DYNJ906303F; 19. ACDF-LF, Medline Kit Number/SKU DYNJ906681C; 20. ACDF-LF, Medline Kit Number/SKU DYNJ906681D; 21. ACDF-LF, Medline Kit Number/SKU DYNJ906681F; 22. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196B; 23. UNIVERSITY FREE FLAP, Medline Kit Number/SKU DYNJ907196C; 24. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ908362F; 25. VP SHUNT, Medline Kit Number/SKU DYNJ909379; 26. VP SHUNT, Medline Kit Number/SKU DYNJ909379A.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments
Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments
Includes an incorrect screw seat interface.
WHILL Model F, with software versions before HMI: 2.23 / MC: 1.05; the device is an indoor/outdoor, foldable battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
WHILL Model F, with software versions before HMI: 2.23 / MC: 1.05; the device is an indoor/outdoor, foldable battery-operated 2-wheel drive (rear-wheel drive) powered wheelchair.
Cyber vulnerability of BLE communication was found by CISA VDP program.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO PACK, Medline kit number/SKU DYNJ64479; 2. CRANIO/NEURO/SPINE, Medline kit number/SKU DYNJ901072D.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO PACK, Medline kit number/SKU DYNJ64479; 2. CRANIO/NEURO/SPINE, Medline kit number/SKU DYNJ901072D.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808
Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Medline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.
Medline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317.
Medline medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. BLOOD CULTURE KIT, Medline Kit SKU DYNDH1520; 2. BLOOD DRAW KIT, Medline Kit SKU EBSI1317.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU CDS982389J; 3. VP SHUNT ADULT PACK-LF, Medline Kit Number/SKU DYNJ0117497G; 4. PICC LINE PACK, Medline Kit Number/SKU DYNJ04592K; 5. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993L; 6. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933F; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861B; 8. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861D; 9. EVD VP SHUNT DBS PACK 319705, Medline Kit Number/SKU DYNJ44603C; 10. BURR HOLE SHUNT PACK, Medline Kit Number/SKU DYNJ56819G; 11. CHRISTUS CHILDRENS VP SHUNT PK, Medline Kit Number/SKU DYNJ61174A; 12. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213C; 13. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213F; 14. NASAL ENDO PACK, Medline Kit Number/SKU DYNJ84501; 15. RO VP SHUNT PACK, Medline Kit Number/SKU PHS392834010D.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU CDS982389J; 3. VP SHUNT ADULT PACK-LF, Medline Kit Number/SKU DYNJ0117497G; 4. PICC LINE PACK, Medline Kit Number/SKU DYNJ04592K; 5. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993L; 6. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933F; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861B; 8. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861D; 9. EVD VP SHUNT DBS PACK 319705, Medline Kit Number/SKU DYNJ44603C; 10. BURR HOLE SHUNT PACK, Medline Kit Number/SKU DYNJ56819G; 11. CHRISTUS CHILDRENS VP SHUNT PK, Medline Kit Number/SKU DYNJ61174A; 12. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213C; 13. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213F; 14. NASAL ENDO PACK, Medline Kit Number/SKU DYNJ84501; 15. RO VP SHUNT PACK, Medline Kit Number/SKU PHS392834010D.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ZALE STEREOTACTIC CDS, Medline Kit Number/SKU CDS983173G; 2. EPISTAXIS TRAY, Medline Kit Number/SKU DYNDA3240; 3. RHINO PACK, Medline Kit Number/SKU DYNJ46504B; 4. MSC SEPTO RHINO PACK-LF, Medline Kit Number/SKU DYNJ51677B; 5. RHINO/SEPTO/FESS PACK, Medline Kit Number/SKU DYNJ55955C; 6. NASAL PACK, Medline Kit Number/SKU DYNJ67789A.
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ZALE STEREOTACTIC CDS, Medline Kit Number/SKU CDS983173G; 2. EPISTAXIS TRAY, Medline Kit Number/SKU DYNDA3240; 3. RHINO PACK, Medline Kit Number/SKU DYNJ46504B; 4. MSC SEPTO RHINO PACK-LF, Medline Kit Number/SKU DYNJ51677B; 5. RHINO/SEPTO/FESS PACK, Medline Kit Number/SKU DYNJ55955C; 6. NASAL PACK, Medline Kit Number/SKU DYNJ67789A.
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
Medline and Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. LVAD DRIVELINE KIT (Centurion), Medline Kit SKU DT17360 (Centurion); 2. LONG TERM/INPATIENT IV SECUREMENT SET (Centurion), Medline Kit SKU IVS1975A; 3. CVC DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT5565B; 4. BSI PICC DRESSING CHANGE KIT, Medline Kit SKU EBSI1172B; 5. BSI CVC DRESSING CHANGE KIT, Medline Kit SKU EBSI1171A; 6. BSI SENSITIVE SKIN DRSNG CHNGE, Medline Kit SKU EBSI1173A; 7. LVAD MANAGEMENT SYSTEM - DAILY (Centurion), Medline Kit SKU DM810; 8. WEEKLY LVAD TRAY (Centurion), Medline Kit SKU DM100; 9. DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT7340; 10. SENSITIVE SKIN LVAD TRAY (Centurion), Medline Kit SKU DM660; 11. PORT DRESSING CHANGE KIT, Medline Kit SKU EBSI1365A.
Medline and Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: 1. LVAD DRIVELINE KIT (Centurion), Medline Kit SKU DT17360 (Centurion); 2. LONG TERM/INPATIENT IV SECUREMENT SET (Centurion), Medline Kit SKU IVS1975A; 3. CVC DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT5565B; 4. BSI PICC DRESSING CHANGE KIT, Medline Kit SKU EBSI1172B; 5. BSI CVC DRESSING CHANGE KIT, Medline Kit SKU EBSI1171A; 6. BSI SENSITIVE SKIN DRSNG CHNGE, Medline Kit SKU EBSI1173A; 7. LVAD MANAGEMENT SYSTEM - DAILY (Centurion), Medline Kit SKU DM810; 8. WEEKLY LVAD TRAY (Centurion), Medline Kit SKU DM100; 9. DRESSING CHANGE TRAY (Centurion), Medline Kit SKU DT7340; 10. SENSITIVE SKIN LVAD TRAY (Centurion), Medline Kit SKU DM660; 11. PORT DRESSING CHANGE KIT, Medline Kit SKU EBSI1365A.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Corp. Micro Bird, Inc. - Contact with Drive Belt Pully May Damage and Cause Oil Lines to Leak
Corp. Micro Bird, Inc. - Contact with Drive Belt Pully May Damage and Cause Oil Lines to Leak
A damaged oil like may leak, increasing the risk of a fire in the presence of an ignition source.
Specially SELECTED brand Vanilla Creme Brulee; NET WT 3.6oz (103g); INGREDIENTS: CREAM (MILK, CREAM), FROZEN YOLK MIX (EGG YOLK, SUGAR) SUGAR, NATURAL VANILLA FLAVOR, TAPIOCA STARCH, CARRAGEENAN; TOPPING INGREDIENTS: CARAMEL FLAKES (SUGAR, GLUCOSE SYRUP); CONTAINS: MILK, EGGS. MAY CONTAIN: WHEAT, SOY; DIST.SOLD EXCLUSIVELY BY: ALDI BATAVIA, IL; packaged in glass jars
Specially SELECTED brand Vanilla Creme Brulee; NET WT 3.6oz (103g); INGREDIENTS: CREAM (MILK, CREAM), FROZEN YOLK MIX (EGG YOLK, SUGAR) SUGAR, NATURAL VANILLA FLAVOR, TAPIOCA STARCH, CARRAGEENAN; TOPPING INGREDIENTS: CARAMEL FLAKES (SUGAR, GLUCOSE SYRUP); CONTAINS: MILK, EGGS. MAY CONTAIN: WHEAT, SOY; DIST.SOLD EXCLUSIVELY BY: ALDI BATAVIA, IL; packaged in glass jars
Product may contain foreign objects, specifically glass.
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
Philips SmartPath to dStream for XR and 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781270. 2. Model Number (REF): 782113.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Mesh Style Tip Protectors CSW-03-2.0 Open-end style CSW-04-4.0 Closed-end style
Mesh Style Tip Protectors CSW-03-2.0 Open-end style CSW-04-4.0 Closed-end style
Product does not have FDA clearance.
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.
Depuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.