iLet Bionic Pancreas, REF: BB1001
iLet Bionic Pancreas, REF: BB1001
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
Medline convenience kits: PORT A CATH DYNJ906818A
Medline convenience kits: PORT A CATH DYNJ906818A
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models
Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Medline convenience kits: BASIC DYNJ905869G BASIC PACK DYNJ68484B BIB GENERAL LAPAROSCOPY PA DYNJ58305 CHRISTUS CH PEDIATRIC BASIC DYNJ61167G CVC INSERTION BUNDLE CVI3940A DAVINCI PACK SURGICOUNT DYNJ45544F EPCH BASIC PACK DYNJ83094C GENERAL PACK DYNJ39682B DYNJ84882 GENERAL RFD DYNJ906336F GENERAL SURGERY PACK DYNJT6003 GV MACD PACK DYNJ46952M IMPLANT PACK - BBI AR DYNJ63508A KIT THORACOSCOPY DYNJ908872F LVAD TRAY DM1120A MAJOR UROLOGY PACK-LF DYNJ0368525M MICRODISECT DECOMPRESS-LF DYNJ911467 MINOR PACK DYNJ86972 MULTI PURPOSE PACK DYNJ58406B OPEN HEART DYNJ903245C P&K PACK DYNJT5838 PACK,PEDIATRIC MAJOR DYNJ906888F PCNL DYNJ904468C PEDIATRIC MINOR RFD DYNJ906296I PEDIATRIC PACK DYNJ54239K PEDIATRIC SURGERY DYNJ81474B PERLMUTTER PACK DYNJT7387 RFT ISC NOBLES MINOR PACK DYNJ43954B ROBOTICS FMOL LADY OF THE LAKE DYNJ89584 ROBOTICS PACK DYNJ83622B SCOLI PACK NTX DYNJ68415C SMALL ABDOMINAL PACK-LF PHS41749G STIM IMPLANT TRIALS DYNJ87790A SUPPLEMENTAL PACK DYNJ65244 TENEX PACK DYNDH2190 THORACIC PACK DYNJ56616G THORACIC PACK SHD DYNJ68416C THORACOSCOPY DYNJ9425843R THORACOTOMY PM DYNJ906815A THORACOTOMY THORACOSCOPY DYNJ902882D TRAY,NEONATAL PICC DRSG CHANGE DYNDC1533B UNIVERSAL-ASC DYNJ900071C
Medline convenience kits: BASIC DYNJ905869G BASIC PACK DYNJ68484B BIB GENERAL LAPAROSCOPY PA DYNJ58305 CHRISTUS CH PEDIATRIC BASIC DYNJ61167G CVC INSERTION BUNDLE CVI3940A DAVINCI PACK SURGICOUNT DYNJ45544F EPCH BASIC PACK DYNJ83094C GENERAL PACK DYNJ39682B DYNJ84882 GENERAL RFD DYNJ906336F GENERAL SURGERY PACK DYNJT6003 GV MACD PACK DYNJ46952M IMPLANT PACK - BBI AR DYNJ63508A KIT THORACOSCOPY DYNJ908872F LVAD TRAY DM1120A MAJOR UROLOGY PACK-LF DYNJ0368525M MICRODISECT DECOMPRESS-LF DYNJ911467 MINOR PACK DYNJ86972 MULTI PURPOSE PACK DYNJ58406B OPEN HEART DYNJ903245C P&K PACK DYNJT5838 PACK,PEDIATRIC MAJOR DYNJ906888F PCNL DYNJ904468C PEDIATRIC MINOR RFD DYNJ906296I PEDIATRIC PACK DYNJ54239K PEDIATRIC SURGERY DYNJ81474B PERLMUTTER PACK DYNJT7387 RFT ISC NOBLES MINOR PACK DYNJ43954B ROBOTICS FMOL LADY OF THE LAKE DYNJ89584 ROBOTICS PACK DYNJ83622B SCOLI PACK NTX DYNJ68415C SMALL ABDOMINAL PACK-LF PHS41749G STIM IMPLANT TRIALS DYNJ87790A SUPPLEMENTAL PACK DYNJ65244 TENEX PACK DYNDH2190 THORACIC PACK DYNJ56616G THORACIC PACK SHD DYNJ68416C THORACOSCOPY DYNJ9425843R THORACOTOMY PM DYNJ906815A THORACOTOMY THORACOSCOPY DYNJ902882D TRAY,NEONATAL PICC DRSG CHANGE DYNDC1533B UNIVERSAL-ASC DYNJ900071C
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Description/REF: ACCESS TRAY/ASK-04001-MC3
Description/REF: ACCESS TRAY/ASK-04001-MC3
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Description/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
Description/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C
MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Medline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C
Medline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F
Medline convenience kit: (82)LABOR & DELIVERY C-SECTION DYNJ907895F
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
DHP Topical Anesthetic Gel, Benzocaine 20%, Strawberry Flavor, 1 oz. (30 g), Manufactured for and Distributed by Dental Health Products, Inc, 2614 North Sugar Bush Road, New Franken, WI 54229. NDC 69634-020-30
DHP Topical Anesthetic Gel, Benzocaine 20%, Strawberry Flavor, 1 oz. (30 g), Manufactured for and Distributed by Dental Health Products, Inc, 2614 North Sugar Bush Road, New Franken, WI 54229. NDC 69634-020-30
Defective container:may contain bottles with incomplete seals
MEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
MEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSEL CATH KIT: 20GA X 5IN/AK-04550-S CVC KIT: 1-LUMEN 5 FR X 20 CM/AK-04301-SP CVC KIT: 16 GA X 6IN (16 CM)/AK-04306-SP CVC KIT: 16GA X 6 IN/AK-04800-S CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-12402 CVC KIT: 3-LUMEN 7 FR X 16 CM/AK-12703-CDC CVC KIT: 2L 4 FR X 13 CM/AK-14402 CVC KIT: 2-LUMEN 5 FR X 13 CM/AK-14502 CVC KIT: 3-LUMEN 5.5 FR X 30 CM/AK-14553 CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-14703-SP CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-15402 CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-15553 CVC KIT: 2-LUMEN 4 FR X 30 CM/AK-16402 CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-16553 CVC KIT: 5 FR X 20 CM/ASK-04301-HH2 CVC KIT 3L: 7 FR X 6 IN (16 CM)/ASK-12703-MGH3
Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSEL CATH KIT: 20GA X 5IN/AK-04550-S CVC KIT: 1-LUMEN 5 FR X 20 CM/AK-04301-SP CVC KIT: 16 GA X 6IN (16 CM)/AK-04306-SP CVC KIT: 16GA X 6 IN/AK-04800-S CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-12402 CVC KIT: 3-LUMEN 7 FR X 16 CM/AK-12703-CDC CVC KIT: 2L 4 FR X 13 CM/AK-14402 CVC KIT: 2-LUMEN 5 FR X 13 CM/AK-14502 CVC KIT: 3-LUMEN 5.5 FR X 30 CM/AK-14553 CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-14703-SP CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-15402 CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-15553 CVC KIT: 2-LUMEN 4 FR X 30 CM/AK-16402 CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-16553 CVC KIT: 5 FR X 20 CM/ASK-04301-HH2 CVC KIT 3L: 7 FR X 6 IN (16 CM)/ASK-12703-MGH3
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall documents for a full list of products)
MEDLINE Medical Procedure Kits labeled as: KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall documents for a full list of products)
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Med Pride, Bacitracin Zinc Ointment, 1oz (28.3g) per tube, 72 tubes per case, Manufactured for: Shield Line LLC, 59 Hook Road, Bayonne, NJ 07002 USA, Made in India, UPC X004WB4LKT.
Med Pride, Bacitracin Zinc Ointment, 1oz (28.3g) per tube, 72 tubes per case, Manufactured for: Shield Line LLC, 59 Hook Road, Bayonne, NJ 07002 USA, Made in India, UPC X004WB4LKT.
CGMP Deviations; deficiencies observed during FDA inspection
Airstream, Inc. - Printed Circuit Board May Overheat
Airstream, Inc. - Printed Circuit Board May Overheat
An overheated printed circuit board increases the risk of a fire.
safco SensiCaine-Ultra (20% Benzocaine), Topical Anesthetic Gel, Strawberry Flavor, 1 oz. (29.6mL), Distributed by: Safco Dental Supply Co., Buffalo Grove, IL 60099, Made in USA, NDC 67239-0223-1.
safco SensiCaine-Ultra (20% Benzocaine), Topical Anesthetic Gel, Strawberry Flavor, 1 oz. (29.6mL), Distributed by: Safco Dental Supply Co., Buffalo Grove, IL 60099, Made in USA, NDC 67239-0223-1.
Defective container:may contain bottles with incomplete seals
Medline convenience kits: DAVINCI MAIN DYNJ908031D
Medline convenience kits: DAVINCI MAIN DYNJ908031D
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.
MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
MEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
MEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Med Pride, Bacitracin Zinc Ointment, 500 units, 1oz (28.3g) tubes, Manufactured for: Shield Line LLC, 59 Hook Road, Bayonne, NJ 07002 USA, Made in India, UPC 3 52410 30354 8.
Med Pride, Bacitracin Zinc Ointment, 500 units, 1oz (28.3g) tubes, Manufactured for: Shield Line LLC, 59 Hook Road, Bayonne, NJ 07002 USA, Made in India, UPC 3 52410 30354 8.
CGMP Deviations; deficiencies observed during FDA inspection
Description/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHS1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU13; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UNHS1
Description/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHS1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU13; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UNHS1
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Description/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
Description/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.