Vereinigte Staaten Recalls

Verbraucherrückrufe von FDA, CPSC, NHTSA und USDA FSIS.

Showing 24 of 17,563 results
Medical DeviceUSFDA Devices

Allura Xper FD20/10; Model Number: 722029;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD20/10; Model Number: 722029;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

SMV DSI, System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
FDA DevicesClass II

SMV DSI, System, Tomography, Computed, Emission

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Jan 21, 2026GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Medical DeviceUSFDA Devices

Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A

Heraeus Medical GmbH (Dental Division)
FDA DevicesClass II

Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A

increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.

Jan 21, 2026Heraeus Medical GmbH (Dental Division)
Food & BeverageUSFDA Food

Tops Strawberry Jam INGREDIENTS: Sugar, Strawberry Pulp 30%, Gelling Agent (Pectin - INS 440), Acidity Regulator (Citric Acid - INS 330), Preservative (Sodium Benzoate - INS 211, Potassium Metabisulphite - INS 224), Food Colour (Carmoisine - INS 122), Nature Identical Flavouring Substances - Strawberry. Allergen Advice: Contains Sulphite, May Contain Soy & Wheat. PRODUCT OF INDIA Brand Owned & Mktd. by: G.D. Foods Mfg. (I) Pvt. Ltd. C-1/119, Janak Puri, New Delhi - 110058 (INDIA) Mfd. by: G.D. Foods Mfg. (I) Pvt. Ltd. Plot No. SP-15&16. RIICO Industrial Area, Neemrana-301705, Dist. - Kotputli-Behror, Rajasthan (INDIA) Imported and Distributed by: Mangalm LLC California, USA Email: info@mangalm.com Phone: 925-302-6025 UPC: 8 904288 626322 Batch No.: 4F1101 Best Before: 10.06.26 NET QTY.: 16.75 oz (475g)

Mangalm LLC
FDA FoodClass II

Tops Strawberry Jam INGREDIENTS: Sugar, Strawberry Pulp 30%, Gelling Agent (Pectin - INS 440), Acidity Regulator (Citric Acid - INS 330), Preservative (Sodium Benzoate - INS 211, Potassium Metabisulphite - INS 224), Food Colour (Carmoisine - INS 122), Nature Identical Flavouring Substances - Strawberry. Allergen Advice: Contains Sulphite, May Contain Soy & Wheat. PRODUCT OF INDIA Brand Owned & Mktd. by: G.D. Foods Mfg. (I) Pvt. Ltd. C-1/119, Janak Puri, New Delhi - 110058 (INDIA) Mfd. by: G.D. Foods Mfg. (I) Pvt. Ltd. Plot No. SP-15&16. RIICO Industrial Area, Neemrana-301705, Dist. - Kotputli-Behror, Rajasthan (INDIA) Imported and Distributed by: Mangalm LLC California, USA Email: info@mangalm.com Phone: 925-302-6025 UPC: 8 904288 626322 Batch No.: 4F1101 Best Before: 10.06.26 NET QTY.: 16.75 oz (475g)

Unapproved color (carmoisine).

Jan 21, 2026Mangalm LLC
PharmaceuticalUSFDA Drugs

Claritin, 20 Count, UPC: 655708016305

GOLD STAR DISTRIBUTION INC
FDA DrugsClass II

Claritin, 20 Count, UPC: 655708016305

CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.

Jan 21, 2026GOLD STAR DISTRIBUTION INC
Medical DeviceUSFDA Devices

Allura Xper FD10/10; Model Number: 722005;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD10/10; Model Number: 722005;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

Allura Xper FD10/10; Model Number: 722027;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD10/10; Model Number: 722027;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
PharmaceuticalUSFDA Drugs

TESTO-100 CIII (testosterone 100 mg), 10 Sterile Pellets per carton, Rx only, Manufactured by: Advanced Pharmaceutical Technology, 132 South Central Avenue, Elmsford, NY 10523, NDC# 57377-100-01

Advanced Pharmaceutical Technology, Inc.
FDA DrugsClass II

TESTO-100 CIII (testosterone 100 mg), 10 Sterile Pellets per carton, Rx only, Manufactured by: Advanced Pharmaceutical Technology, 132 South Central Avenue, Elmsford, NY 10523, NDC# 57377-100-01

Lack of Assurance of Sterility: Potential microbial contamination of subcutaneous pellets.

Jan 21, 2026Advanced Pharmaceutical Technology, Inc.
PharmaceuticalUSFDA Drugs

Swan, White Clear Alcohol, 50%, 16oz, 12 count

GOLD STAR DISTRIBUTION INC
FDA DrugsClass II

Swan, White Clear Alcohol, 50%, 16oz, 12 count

CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.

Jan 21, 2026GOLD STAR DISTRIBUTION INC
Medical DeviceUSFDA Devices

Allura Xper FD20 Biplane; Model Number: 722008;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD20 Biplane; Model Number: 722008;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

SMV DSX New Line, System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
FDA DevicesClass II

SMV DSX New Line, System, Tomography, Computed, Emission

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Jan 21, 2026GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Medical DeviceUSFDA Devices

Allura Xper FD20 OR Table; Model Number: 722023;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD20 OR Table; Model Number: 722023;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

Medicrea, IB3D Universal Implant Inserter, Rx Only REF: A24000000

Medicrea International
FDA DevicesClass II

Medicrea, IB3D Universal Implant Inserter, Rx Only REF: A24000000

Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.

Jan 21, 2026Medicrea International
Food & BeverageUSFDA Food

HerbsForever- Hingwastika Churna ; Ayruvedic Herbal Supplement; 90 VEGE Capsules; 100 GMS UPC: 8 07814 00622 4

Herbs Forever
FDA FoodClass I

HerbsForever- Hingwastika Churna ; Ayruvedic Herbal Supplement; 90 VEGE Capsules; 100 GMS UPC: 8 07814 00622 4

Potential undeclared allergen ingredient (Wheat)

Jan 21, 2026Herbs Forever
Medical DeviceUSFDA Devices

Allura Xper FD10; Model Number: 722003;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD10; Model Number: 722003;

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Jan 21, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Vehicle RecallUSNHTSA

Volkswagen Group of America, Inc. - Park Outside - High Voltage Battery May Catch Fire

Volkswagen Group of America, Inc.
NHTSAELECTRICAL SYSTEM

Volkswagen Group of America, Inc. - Park Outside - High Voltage Battery May Catch Fire

A vehicle fire while parked or driving can increase the risk of injury.

Jan 21, 2026Volkswagen Group of America, Inc.
Medical DeviceUSFDA Devices

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) ADMIN KIT, Medline kit SKU DYKS1339B; 2) ADMIN KIT, Medline kit SKU DYKS1339C; 3) QCASC WELCOME KIT, Medline kit SKU DYKS1359B; 4) ADMISSION KIT, Medline kit SKU DYKS1534D.

Medline Industries, LP
FDA DevicesClass II

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) ADMIN KIT, Medline kit SKU DYKS1339B; 2) ADMIN KIT, Medline kit SKU DYKS1339C; 3) QCASC WELCOME KIT, Medline kit SKU DYKS1359B; 4) ADMISSION KIT, Medline kit SKU DYKS1534D.

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

Jan 21, 2026Medline Industries, LP
PharmaceuticalUSFDA Drugs

Benadryl, 25'S 2PK Box, UPC 815556020316

GOLD STAR DISTRIBUTION INC
FDA DrugsClass II

Benadryl, 25'S 2PK Box, UPC 815556020316

CGMP Deviations: Insanitary conditions including rodent exposure/activity in their distribution center.

Jan 21, 2026GOLD STAR DISTRIBUTION INC
Vehicle RecallUSNHTSA

Alexis Fire Equipment Company - Loss of Vehicle Control from Loose Axle Clamp Fasteners

Alexis Fire Equipment Company
NHTSASUSPENSION

Alexis Fire Equipment Company - Loss of Vehicle Control from Loose Axle Clamp Fasteners

A loss of vehicle control increases the risk of a crash.

Jan 21, 2026Alexis Fire Equipment Company
Medical DeviceUSFDA Devices

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103.

Medline Industries, LP
FDA DevicesClass II

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103.

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

Jan 21, 2026Medline Industries, LP
Food & BeverageUSFDA Food

Ukrop's 6" Pound Cake with Buttercream Icing/Pink Roses NET WT 44 OZ (2.75lb/1.25kg) CONTAINS: Egg, Milk, Wheat Made in a production area that also uses soy and tree nuts. UPC 7 22515 91452 7 Ukrop's Homestyle Foods, LLC Richmond, VA 23230

Ukrop's Homestyle Foods, LLC
FDA FoodClass II

Ukrop's 6" Pound Cake with Buttercream Icing/Pink Roses NET WT 44 OZ (2.75lb/1.25kg) CONTAINS: Egg, Milk, Wheat Made in a production area that also uses soy and tree nuts. UPC 7 22515 91452 7 Ukrop's Homestyle Foods, LLC Richmond, VA 23230

Undeclared Soy. Label did not include soy, which was used in product.

Jan 21, 2026Ukrop's Homestyle Foods, LLC
Medical DeviceUSFDA Devices

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Elekta, Inc.
FDA DevicesClass II

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Jan 21, 2026Elekta, Inc.
Medical DeviceUSFDA Devices

VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

Canon Medical System, USA, INC.
FDA DevicesClass II

VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Jan 21, 2026Canon Medical System, USA, INC.
Medical DeviceUSFDA Devices

CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System

Thoratec LLC
FDA DevicesClass II

CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System

Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.

Jan 21, 2026Thoratec LLC