Vereinigte Staaten Recalls
Verbraucherrückrufe von FDA, CPSC, NHTSA und USDA FSIS.
IQon Spectral CT; Product Code (REF): 728332;
IQon Spectral CT; Product Code (REF): 728332;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109
Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
WB Impression Post Closed Tray incl. guide screw, L 13mm, TAN/POM
WB Impression Post Closed Tray incl. guide screw, L 13mm, TAN/POM
The mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown.
1) UNIVERSAL KIT, Kit SKU DT22010B; 2) VAD ACCESS KIT, Kit SKU DT22790C; 3) VAD ACCESS KIT, Kit SKU DT22790C; 4) PROCEDURE TRAY, Kit SKU DYNDA2966A; 5) PROCEDURE TRAY, Kit SKU DYNDA2966A; 6) PEDIATRIC CENTRAL LINE SMALL, Kit SKU DYNDC2208C; 7) PORT DRESSING KIT, Kit SKU DYNDC3107; 8) NON STERILE DIALYSIS KIT, Kit SKU DYNDH1029C; 9) UMBILICAL TRAY W/3.5&5FR CATH, Kit SKU UVT1250; 10) UMBILICAL VESSEL TRAY, Kit SKU UVT835; 11) HEMODIALYSIS TRAY, Kit SKU VAR101HVAA.
1) UNIVERSAL KIT, Kit SKU DT22010B; 2) VAD ACCESS KIT, Kit SKU DT22790C; 3) VAD ACCESS KIT, Kit SKU DT22790C; 4) PROCEDURE TRAY, Kit SKU DYNDA2966A; 5) PROCEDURE TRAY, Kit SKU DYNDA2966A; 6) PEDIATRIC CENTRAL LINE SMALL, Kit SKU DYNDC2208C; 7) PORT DRESSING KIT, Kit SKU DYNDC3107; 8) NON STERILE DIALYSIS KIT, Kit SKU DYNDH1029C; 9) UMBILICAL TRAY W/3.5&5FR CATH, Kit SKU UVT1250; 10) UMBILICAL VESSEL TRAY, Kit SKU UVT835; 11) HEMODIALYSIS TRAY, Kit SKU VAR101HVAA.
Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.
UDENYCA, pegfilgrastim-cbqv injection, 6 mg/0.6mL Single Dose Prefilled Syringe, Rx only, Manufactured by Accord BioPharma, Inc., 8041 Arco Corporate Drive, Suite 200, Raleigh, NC 27617, NDC 69448-025-63
UDENYCA, pegfilgrastim-cbqv injection, 6 mg/0.6mL Single Dose Prefilled Syringe, Rx only, Manufactured by Accord BioPharma, Inc., 8041 Arco Corporate Drive, Suite 200, Raleigh, NC 27617, NDC 69448-025-63
Temperature Abuse. 116 cartons with specific serial numbers of Lot 2199821, were stored incorrectly in a controlled room temperature environment instead of the required refrigerated environment.
Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Testosterone 100mg, Anastrozole 4mg (4.5), 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4104-01.
Testosterone 100mg, Anastrozole 4mg (4.5), 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4104-01.
Presence of Foreign substance - potential presence of metal particulate matter
Spectral CT 7500 on Rails; Product Code (REF): 728334;
Spectral CT 7500 on Rails; Product Code (REF): 728334;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Testosterone, 25.0 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-1025-32.
Testosterone, 25.0 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-1025-32.
Presence of Foreign substance - potential presence of metal particulate matter
Testosterone, 62.5 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-1062-32.
Testosterone, 62.5 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-1062-32.
Presence of Foreign substance - potential presence of metal particulate matter
Testosterone 87.5mg, Triamcinolone Acetonide17.5 mcg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-6087-32.
Testosterone 87.5mg, Triamcinolone Acetonide17.5 mcg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-6087-32.
Presence of Foreign substance - potential presence of metal particulate matter
Pro Numb Tattoo Numbing Spray, For Sensitive Skin, 4% Lidocaine, [1FL OZ/28.4 ML or 4 oz/120 mL] per bottle, Distributed by: Pro Numb Tattoo numbing spray LLC, Melbourne, FL 32907, NDC 83389-112-11
Pro Numb Tattoo Numbing Spray, For Sensitive Skin, 4% Lidocaine, [1FL OZ/28.4 ML or 4 oz/120 mL] per bottle, Distributed by: Pro Numb Tattoo numbing spray LLC, Melbourne, FL 32907, NDC 83389-112-11
cGMP deviations
Testosterone 100mg Anastrozole 4mg , 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4104-32
Testosterone 100mg Anastrozole 4mg , 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4104-32
Presence of Foreign substance - potential presence of metal particulate matter
Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH
Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
acyclovir sodium, 200 mg, 200mg added to 100 mL, 0.9% Sodium Chloride Injection, USP, For Intravenous Use, Single-Use Bag, Fagron Sterile Services, 20 Dan Road, Canton, MA 02021, NDC 71506-035-56
acyclovir sodium, 200 mg, 200mg added to 100 mL, 0.9% Sodium Chloride Injection, USP, For Intravenous Use, Single-Use Bag, Fagron Sterile Services, 20 Dan Road, Canton, MA 02021, NDC 71506-035-56
Lack of Assurance of Sterility
ketamine HCl, 1,000 mg, 1,000 mg per 100 mL (10 mg per mL) in Sodium Chloride Injection, Fagron Sterile Services, 20 Dan Road, Canton, MA 02021, NDC 71506-050-56
ketamine HCl, 1,000 mg, 1,000 mg per 100 mL (10 mg per mL) in Sodium Chloride Injection, Fagron Sterile Services, 20 Dan Road, Canton, MA 02021, NDC 71506-050-56
Lack of Assurance of Sterility
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
Reprocessed Abbott Inquiry Steerable Diagnostic Catheters: Abbott Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81105RH 81172RH; Abbott Inquiry Steerable Diagnostic Catheter 20 Elec. 2-10-2mm 81202RH; Abbott Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81483RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 2-5-2mm 81102RH 81104RH 81174RH 81531RH 81532RH; Inquiry Steerable Diagnostic Catheter 10 Elec. 5mm 81107RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 2-5-2mm 81402RH 81404RH 81405RH 81474RH; Inquiry Steerable Diagnostic Catheter 4 Elec. 5mm 81473RH
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
ESTRADIOL, 18 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 3521, NDC: 79559-3018-32.
ESTRADIOL, 18 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 3521, NDC: 79559-3018-32.
Presence of Foreign substance - potential presence of metal particulate matter
CREAMY PEANUT BUTTER NET WT 0.75 OZ(21g) DISTRIBUTED BY US FOODS, ROSEMONT, IL 60018; PEANUT BUTTER NET WT. 0.5OZ & NET WT 0.75OZ DIST BY: DYMA BRANDS, INC. ATLANTA, GA 30328; Product also sold under the following brand names: FLAVOR Fresh PEANUT BUTTER NET WT 0.75 OZ(21g) DISTRIBUTED BY: DYMA BRANDS, INC. ATLANTA, GA 30328; HOUSE RECIPE CREAMY PEANUT BUTTER NET WT 0.75 OZ (21.2g) DISTRIBUTED BY SYSCO COROPRATION, HOUSTON, TEXAS 77077; Katy's Kitchen SMOOTH PEANUT BUTTER NET WT 0.75 OZ (21g) MARKETED BY INDEPENDENT MARKETING ALLIANCE HOUSTIN, TX 77043 PEANUT BUTTER creamy NET WT 0.75 OZ (21g) FOR DISTRIBUTION EXCLUSIVELY BY GORDON FOOD SERVICE WYOMING, MI 49509; CREAMY PEANUT BUTTER NET WT 1.12 OZ(32g) POCO PAC DISTRIBUTED BY DYMA BRANDS, INC. ATLANTA, GA 30328
CREAMY PEANUT BUTTER NET WT 0.75 OZ(21g) DISTRIBUTED BY US FOODS, ROSEMONT, IL 60018; PEANUT BUTTER NET WT. 0.5OZ & NET WT 0.75OZ DIST BY: DYMA BRANDS, INC. ATLANTA, GA 30328; Product also sold under the following brand names: FLAVOR Fresh PEANUT BUTTER NET WT 0.75 OZ(21g) DISTRIBUTED BY: DYMA BRANDS, INC. ATLANTA, GA 30328; HOUSE RECIPE CREAMY PEANUT BUTTER NET WT 0.75 OZ (21.2g) DISTRIBUTED BY SYSCO COROPRATION, HOUSTON, TEXAS 77077; Katy's Kitchen SMOOTH PEANUT BUTTER NET WT 0.75 OZ (21g) MARKETED BY INDEPENDENT MARKETING ALLIANCE HOUSTIN, TX 77043 PEANUT BUTTER creamy NET WT 0.75 OZ (21g) FOR DISTRIBUTION EXCLUSIVELY BY GORDON FOOD SERVICE WYOMING, MI 49509; CREAMY PEANUT BUTTER NET WT 1.12 OZ(32g) POCO PAC DISTRIBUTED BY DYMA BRANDS, INC. ATLANTA, GA 30328
Foreign material. During production, the firm found pieces of blue plastic in a filter.
Testosterone 75mg, Anastrozole 4mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4079-32.
Testosterone 75mg, Anastrozole 4mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-4079-32.
Presence of Foreign substance - potential presence of metal particulate matter
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Surepulse VS Cap Medium REF: SP-161-A1 component of VS Newborn Heart Rate Monitor
Surepulse VS Cap Medium REF: SP-161-A1 component of VS Newborn Heart Rate Monitor
Retroactively reported; Labeling contains incorrect sizing guide for Cap, a component of the heart rate monitor.
ESTRADIOL, 20 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-3020-32.
ESTRADIOL, 20 mg, 1 Sterile Pellet, Asteria Health 432 Industrial Ln, Birmingham, AL 35211, NDC: 79559-3020-32.
Presence of Foreign substance - potential presence of metal particulate matter