Vereinigte Staaten Recalls
Verbraucherrückrufe von FDA, CPSC, NHTSA und USDA FSIS.
Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A
Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38B; 6) L & D PACK-LF, Model Number: DYNJ0456816O; 7) L & D PACK-LF, Model Number: DYNJ0456816P; 8) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 10) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 11) DELIVERY PACK-LF, Model Number: DYNJ20546F; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 13) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 14) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 15) OB PACK-LF, Model Number: DYNJ26121I; 16) OB DELIVERY PACK, Model Number: DYNJ31024J; 17) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 18) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 19) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 20) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 21) VAG DELIVERY PACK, Model Number: DYNJ38481G; 22) VAG DELIVERY PACK, Model Number: DYNJ38481I; 23) VAG DELIVERY PACK, Model Number: DYNJ38481K; 24) OB PACK 2, Model Number: DYNJ39000D; 25) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 26) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 27) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 28) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 29) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 31) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 32) DELIVERY PACK, Model Number: DYNJ55373A; 33) VAGINAL DELIVERY PACK, Model Number: DYNJ59403F; 34) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 35) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 36) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 37) OB-VAG PACK 9128190, Model Number: DYNJ62341; 38) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 39) LABOR AND DELIVERY TRAY, Model Number: DYNJ62722B; 40) L & D PACK, Model Number: DYNJ63216A; 41) L&D PACK, Model Number: DYNJ63757A; 42) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 43) MINOR VAG PACK, Model Number: DYNJ68663; 44) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 45) GYN VAGINAL PACK, Model Number: DYNJ68681; 46) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 47) VAGINAL PACK, Model Number: DYNJ69774A; 48) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 49) DELIVERY PACK, Model Number: DYNJ87469A; 50) DELIVERY B, Model Number: DYNJ900351T; 51) VAGINAL DELIVERY, Model Number: DYNJ901387A; 52) FLOYD DELIVERY, Model Number: DYNJ906937D; 53) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 54) OB PACK, Model Number: DYNJT5186; 55) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 56) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 57) LABOR AND DELIVERY KIT, Model Number: MNS8950A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
busPIRone Hydrochloride Tablets, USP, 5 mg, 500-count bottle, Rx Only, Manufactured by Unichem Laboratories LTD., Ind, Area, Meerut Road, Ghazibad - 201 003, India Manufactured for: Unichem Pharmaceuticals (USA), Inc. East Brunswick, NJ 08816, NDC 29300-244-05.
busPIRone Hydrochloride Tablets, USP, 5 mg, 500-count bottle, Rx Only, Manufactured by Unichem Laboratories LTD., Ind, Area, Meerut Road, Ghazibad - 201 003, India Manufactured for: Unichem Pharmaceuticals (USA), Inc. East Brunswick, NJ 08816, NDC 29300-244-05.
Subpotent drug
Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).
Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF-CSC-20 G31221 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
DEX Ophthalmic Tissue Forceps, 27ga DEX Stiff Maculorhexis Forceps, Model/Catalog Number: DVF4019-27
DEX Ophthalmic Tissue Forceps, 27ga DEX Stiff Maculorhexis Forceps, Model/Catalog Number: DVF4019-27
Field Safety Corrective Action for IFU in DEX Forceps and Scissors.
See RES for complete list. Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326B; 4) DRAWER #5, Model Number: ACC010475; 5) DRAWER 1 NEURO CART, Model Number: ACC010542; 6) DRAWER 2 NEURO CART, Model Number: ACC010543; 7) DRAWER 4, Model Number: ACC010722; 8) SUCTION TRAY, Model Number: ACC010862; 9) MALIGNANT HYPERTHERMIA CART, Model Number: ACC010896;
See RES for complete list. Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326B; 4) DRAWER #5, Model Number: ACC010475; 5) DRAWER 1 NEURO CART, Model Number: ACC010542; 6) DRAWER 2 NEURO CART, Model Number: ACC010543; 7) DRAWER 4, Model Number: ACC010722; 8) SUCTION TRAY, Model Number: ACC010862; 9) MALIGNANT HYPERTHERMIA CART, Model Number: ACC010896;
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Low Heat Non-Fat Dried Milk Powder - paper bags 25 kg. (packed under 5 labels: Off Grade NFDM Variable; Extra Grade NFDM Standard rBST Free; Grade A NFDM LH Basic; Grade A NFDM LH Standard rBST Free; Grade A NFDM LH Tier 1 rBST Free)
Low Heat Non-Fat Dried Milk Powder - paper bags 25 kg. (packed under 5 labels: Off Grade NFDM Variable; Extra Grade NFDM Standard rBST Free; Grade A NFDM LH Basic; Grade A NFDM LH Standard rBST Free; Grade A NFDM LH Tier 1 rBST Free)
Potential contamination with Salmonella.
DEX Ophthalmic Tissue Forceps, 25ga DEX Maculorhexis Forceps, Model/Catalog Number: DVF4019-25
DEX Ophthalmic Tissue Forceps, 25ga DEX Maculorhexis Forceps, Model/Catalog Number: DVF4019-25
Field Safety Corrective Action for IFU in DEX Forceps and Scissors.
Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.
Halyard CARDIAC CATH PACK SAN23CARDQ kit. Model Number: DRCC36.
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Loard's Brownie Nut Fudge Ice Cream - 32 oz
Loard's Brownie Nut Fudge Ice Cream - 32 oz
Undeclared Milk, Walnuts, Eggs, Wheat, Soy.
Duloxetine Delayed-Release Capsules, 30 mg, Rx Only, a) 90 Capsules, NDC 27241-098-09, b) 30 capsules, NDC 27241-098-03, c) 1000 Capsules, NDC 27241-098-10, Marketed by: Ajanta Pharma USA Inc., Bridgewater, NJ 08807, Made in India
Duloxetine Delayed-Release Capsules, 30 mg, Rx Only, a) 90 Capsules, NDC 27241-098-09, b) 30 capsules, NDC 27241-098-03, c) 1000 Capsules, NDC 27241-098-10, Marketed by: Ajanta Pharma USA Inc., Bridgewater, NJ 08807, Made in India
CGMP Deviations: Presence of N-nitroso-Duloxetine impurity above FDA recommended limit of 0.83 ppm, identified at the 12-month and 18-month long-term stability intervals.
Loard's Lemon Chiffon Ice Cream - 32 oz
Loard's Lemon Chiffon Ice Cream - 32 oz
Undeclared Milk, Eggs, Yellow #5.
TPN bag (patient specific), Rx# 11-4909703-0-1, Compound Volume 1660 mL per bag, Rx only, Single Dose Injection, Refrigerated Injection, Central Admixture Pharmacy Services, Los Angeles, 13128 Imperial Hwy, Santa Fe Spr., CA 90670
TPN bag (patient specific), Rx# 11-4909703-0-1, Compound Volume 1660 mL per bag, Rx only, Single Dose Injection, Refrigerated Injection, Central Admixture Pharmacy Services, Los Angeles, 13128 Imperial Hwy, Santa Fe Spr., CA 90670
Incorrect Product Formulation: product did not contain insulin as listed in the label.
Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A
Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Halyard VA DETROIT CATH LAB PACK kit. Model Numbers: VMCD029.
Halyard VA DETROIT CATH LAB PACK kit. Model Numbers: VMCD029.
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069
Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
DELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 1.5% DEX. LM/LC, 2L 5PK, Part Number 054-20221, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.
DELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 1.5% DEX. LM/LC, 2L 5PK, Part Number 054-20221, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.
Lack of Assurance of Sterility: Potential leaks from perforations in bags.
DEX Ophthalmic Tissue Forceps, 25ga DEX Nano Tapered AWH Forceps, Model/Catalog Number: DVF4005-25
DEX Ophthalmic Tissue Forceps, 25ga DEX Nano Tapered AWH Forceps, Model/Catalog Number: DVF4005-25
Field Safety Corrective Action for IFU in DEX Forceps and Scissors.
DEX Ophthalmic Tissue Forceps, 27ga DEX Stiff Super Grip Forceps, Model/Catalog Number: DVF4034-27
DEX Ophthalmic Tissue Forceps, 27ga DEX Stiff Super Grip Forceps, Model/Catalog Number: DVF4034-27
Ophthalmic Tissue Forceps (DEX") are sterile, hand-held ophthalmic surgical instruments designed for grasping, manipulating, compressing, pulling, or joining eye and/or surrounding tissues during surgical procedures. These devices fall under the GMDN code 62478 entitled Ophthalmic soft-tissue surgical forceps, probe-like, single use . Double pouched in tyvek pouches, sold in boxes of 5 with an inner and outer box.
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC
BioCera Fibers, Bioresorbable Bone Void Filler, REF: ORB-0320C-KC, ORB-0310C-KC, ORB-0304C-KC, ORB-0302C-KC, ORB-0305D-KC
Resorbable bone void filler falls outside standard specifications.
Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4A; 6) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4B; 7) VEIN PROCEDURE KIT, Model Number:DYNDA2076C; 8) CARDIAC CATH PACK-LF, Model Number:DYNJ0150605AC; 9) CARDIAC CATH PACK-LF, Model Number:DYNJ0368386D; 10) CATH LAB PACK-LF, Model Number:DYNJ0648473Y; 11) NBMC CATH PACK-LF, Model Number:DYNJ0803367I; 12) RADIOLOGY PACK-LF, Model Number:DYNJ0928124C; 13) ANGIO PACK-LF, Model Number:DYNJ0954970I; 14) SPECIALS TRAY RADIOLOGY, Model Number:DYNJ16803D; 15) CARDIAC CATH PACK, Model Number:DYNJ19946M; 16) CATH LAB PACK, Model Number:DYNJ30955I; 17) CATH LAB PACK, Model Number:DYNJ31773K; 18) CATH PACK-LF, Model Number:DYNJ33061D; 19) DB CCL OAKWOOD PACK-LF, Model Number:DYNJ33620D; 20) CARDIAC CATH PACK MRH-LF, Model Number:DYNJ35300M; 21) CATH LAB PK, Model Number:DYNJ35643C; 22) CARDIAC CATH PACK-LF, Model Number:DYNJ39066B; 23) ANGIOGRAPHY PACK, Model Number:DYNJ42134G; 24) CATH PACK-LF, Model Number:DYNJ43101C; 25) OR ANGIO PACK-LF, Model Number:DYNJ43415D; 26) OR ANGIO PACK-LF, Model Number:DYNJ43415F; 27) ANGIOPLASTY PACK, Model Number:DYNJ44700J; 28) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ44850J; 29) RADIOLOGICAL PACK, Model Number:DYNJ44889O; 30) CARDIAC CATH PACK, Model Number:DYNJ47370B; 31) ANGIOGRAPHY PACK, Model Number:DYNJ53129D; 32) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ56772G; 33) DR. MOHINDRA PACK, Model Number:DYNJ57602G; 34) NE ENDO VASCULAR PACK, Model Number:DYNJ57940B; 35) CARDIAC CATH PACK - Q1, Model Number:DYNJ58048C; 36) OR ANGIO, Model Number:DYNJ59931B; 37) LPCH PERRY CATH PACK, Model Number:DYNJ67378C; 38) HOSPITAL REGIONAL DE CONCEPCIO, Model Number:DYNJ67918A; 39) CATH LAB PACK, Model Number:DYNJ68633; 40) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197B; 41) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197C; 42) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197D; 43) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197F; 44) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197G; 45) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197I; 46) BASIC CARDIAC CATH PACK, Model Number:DYNJ81593B; 47) CATH LAB PACK, Model Number:DYNJ83073; 48) UMC EL PASO CATH PACK, Model Number:DYNJ83089; 49) DR. MOHINDRA PACK, Model Number:DYNJ83819A; 50) VZ VAS/ANG PACK, Model Number:DYNJ83981; 51) WCPN CARDIOLOGY PACK, Model Number:DYNJ84539; 52) CATH LAB, Model Number:DYNJ85511; 53) CATH LAB, Model Number:DYNJ85511A; 54) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJT1520K; 55) CVC BUNDLE KIT, Model Number:ECVC6225A; 56) CVC 3L 7F 20CM VANTEX, Model Number:ECVC7265A; 57) ANGIOGRAPHY TRAY, Model Number:MNS9130
Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4A; 6) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4B; 7) VEIN PROCEDURE KIT, Model Number:DYNDA2076C; 8) CARDIAC CATH PACK-LF, Model Number:DYNJ0150605AC; 9) CARDIAC CATH PACK-LF, Model Number:DYNJ0368386D; 10) CATH LAB PACK-LF, Model Number:DYNJ0648473Y; 11) NBMC CATH PACK-LF, Model Number:DYNJ0803367I; 12) RADIOLOGY PACK-LF, Model Number:DYNJ0928124C; 13) ANGIO PACK-LF, Model Number:DYNJ0954970I; 14) SPECIALS TRAY RADIOLOGY, Model Number:DYNJ16803D; 15) CARDIAC CATH PACK, Model Number:DYNJ19946M; 16) CATH LAB PACK, Model Number:DYNJ30955I; 17) CATH LAB PACK, Model Number:DYNJ31773K; 18) CATH PACK-LF, Model Number:DYNJ33061D; 19) DB CCL OAKWOOD PACK-LF, Model Number:DYNJ33620D; 20) CARDIAC CATH PACK MRH-LF, Model Number:DYNJ35300M; 21) CATH LAB PK, Model Number:DYNJ35643C; 22) CARDIAC CATH PACK-LF, Model Number:DYNJ39066B; 23) ANGIOGRAPHY PACK, Model Number:DYNJ42134G; 24) CATH PACK-LF, Model Number:DYNJ43101C; 25) OR ANGIO PACK-LF, Model Number:DYNJ43415D; 26) OR ANGIO PACK-LF, Model Number:DYNJ43415F; 27) ANGIOPLASTY PACK, Model Number:DYNJ44700J; 28) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ44850J; 29) RADIOLOGICAL PACK, Model Number:DYNJ44889O; 30) CARDIAC CATH PACK, Model Number:DYNJ47370B; 31) ANGIOGRAPHY PACK, Model Number:DYNJ53129D; 32) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ56772G; 33) DR. MOHINDRA PACK, Model Number:DYNJ57602G; 34) NE ENDO VASCULAR PACK, Model Number:DYNJ57940B; 35) CARDIAC CATH PACK - Q1, Model Number:DYNJ58048C; 36) OR ANGIO, Model Number:DYNJ59931B; 37) LPCH PERRY CATH PACK, Model Number:DYNJ67378C; 38) HOSPITAL REGIONAL DE CONCEPCIO, Model Number:DYNJ67918A; 39) CATH LAB PACK, Model Number:DYNJ68633; 40) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197B; 41) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197C; 42) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197D; 43) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197F; 44) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197G; 45) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197I; 46) BASIC CARDIAC CATH PACK, Model Number:DYNJ81593B; 47) CATH LAB PACK, Model Number:DYNJ83073; 48) UMC EL PASO CATH PACK, Model Number:DYNJ83089; 49) DR. MOHINDRA PACK, Model Number:DYNJ83819A; 50) VZ VAS/ANG PACK, Model Number:DYNJ83981; 51) WCPN CARDIOLOGY PACK, Model Number:DYNJ84539; 52) CATH LAB, Model Number:DYNJ85511; 53) CATH LAB, Model Number:DYNJ85511A; 54) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJT1520K; 55) CVC BUNDLE KIT, Model Number:ECVC6225A; 56) CVC 3L 7F 20CM VANTEX, Model Number:ECVC7265A; 57) ANGIOGRAPHY TRAY, Model Number:MNS9130
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
DELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 2.5% DEX. LM/LC, 2L 5PK, Part Number 054-20222, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.
DELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for Intraperitoneal Dialysis Only, 2.5% DEX. LM/LC, 2L 5PK, Part Number 054-20222, Fresenius Medical Care North America, 920 Winter Street, Waltham, MA 02451.
Lack of Assurance of Sterility: Potential leaks from perforations in bags.
Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case
Medical Action Industries Inc. Pack Cath BHH, REF: BHCA49K, Sterile EO, Rx Only, QTY: 5/Case
Convenience kit contains a syringe in which the rotating adapter may unwind during use, resulting in a loose connection and/or full disconnection between the syringe and manifold"