Medizinprodukte Recalls
Defective surgical instruments, diagnostic equipment, and health device recalls.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Focalyx Fusion
Focalyx Fusion
Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.
Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Product Name: Lumipulse ¿-Amyloid Plasma Controls Model/Catalog Number: 81559 Software Version: Not Applicable Product Description: Lumipulse ¿-Amyloid Plasma Controls: This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range. Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen), 2 ¿2 Concentrations L1 Level 1 (2¿ 1.5 mL) L2 Level 2 (2¿ 1.5 mL) Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27
Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Male Fertility Sperm Test for Home Use (Cassette)
Male Fertility Sperm Test for Home Use (Cassette)
Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.
Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25
Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
PRIMSAFLEX M100 SET, Product Code 106697; Dialyzer, High Permeability With Or Without Sealed Dialysate System
PRIMSAFLEX M100 SET, Product Code 106697; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;
AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
PRISMAFLEX HF1000 SET, Product Code 107140; Dialyzer, High Permeability With Or Without Sealed Dialysate System
PRISMAFLEX HF1000 SET, Product Code 107140; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
Nasal feeding tube packaged without the nasal transfer tube component.