Medical DeviceUSFDA Devices

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc)

Mentor Texas, LP.
FDA DevicesClass II

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc)

Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.

Apr 8, 2026Mentor Texas, LP.
Medical DeviceUSFDA Devices

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Abiomed, Inc.
FDA DevicesClass I

Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 7. Optical AIC w/Impella Connect, Pack'd, CA; Product Code: 0042-0040-CA. 8. Optical AIC w/Impella Connect, Pack'd, EU; Product Code: 0042-0040-EU. 9. Optical, AIC, Impella Connect, Pkgd, AU; Product Code: 0042-0040-AU.

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Apr 8, 2026Abiomed, Inc.
Medical DeviceUSFDA Devices

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc

Mentor Texas, LP.
FDA DevicesClass II

CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc

Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.

Apr 8, 2026Mentor Texas, LP.
Medical DeviceUSFDA Devices

Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc

Mentor Texas, LP.
FDA DevicesClass II

Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc

Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.

Apr 8, 2026Mentor Texas, LP.
Medical DeviceUSFDA Devices

Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001

Stryker Communications
FDA DevicesClass II

Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001

Surgical light assembly may not adequate support the weight of the ceiling cover.

Apr 8, 2026Stryker Communications
Medical DeviceUSFDA Devices

MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

Medtronic MiniMed, Inc.
FDA DevicesClass II

MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)

All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

Apr 8, 2026Medtronic MiniMed, Inc.
Medical DeviceUSFDA Devices

Burlington Medical, Thyroid Shield.

Burlington Medical, LLC
FDA DevicesClass II

Burlington Medical, Thyroid Shield.

Potential for attenuation degradation over time, decreasing the lifespan.

Apr 8, 2026Burlington Medical, LLC
Medical DeviceUSFDA Devices

eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149

Civco Medical Instruments Co. Inc.
FDA DevicesClass II

eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Apr 8, 2026Civco Medical Instruments Co. Inc.
Medical DeviceUSFDA Devices

Allura Xper FD20/20 OR Table; System Code: 722039;

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA DevicesClass II

Allura Xper FD20/20 OR Table; System Code: 722039;

Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

Apr 8, 2026PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Medical DeviceUSFDA Devices

MEDLINE ANTERIOR HIP PACK DYNJ64672B

Medline Industries, LP
FDA DevicesClass II

MEDLINE ANTERIOR HIP PACK DYNJ64672B

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Apr 8, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Sterile Radiology Procedure Kits, Model Number DYNDH1491B

Medline Industries, LP
FDA DevicesClass II

Sterile Radiology Procedure Kits, Model Number DYNDH1491B

There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility.

Apr 8, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.

I.T.S. GmbH
FDA DevicesClass II

HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Apr 1, 2026I.T.S. GmbH
Medical DeviceUSFDA Devices

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Apr 1, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61

Medtronic MiniMed, Inc.
FDA DevicesClass II

MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61

Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery).

Apr 1, 2026Medtronic MiniMed, Inc.
Medical DeviceUSFDA Devices

Philips X7-2 Ultrasound Transducer

Philips Ultrasound, Inc
FDA DevicesClass III

Philips X7-2 Ultrasound Transducer

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

Apr 1, 2026Philips Ultrasound, Inc
Medical DeviceUSFDA Devices

Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Apr 1, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B

Merit Medical Systems, Inc.
FDA DevicesClass I

16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Apr 1, 2026Merit Medical Systems, Inc.
Medical DeviceUSFDA Devices

I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.

I.T.S. GmbH
FDA DevicesClass II

I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Apr 1, 2026I.T.S. GmbH
Medical DeviceUSFDA Devices

ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C

Merit Medical Systems, Inc.
FDA DevicesClass I

ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C

16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Apr 1, 2026Merit Medical Systems, Inc.
Medical DeviceUSFDA Devices

Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

Medline Industries, LP
FDA DevicesClass II

Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.

Apr 1, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Philips OMNI III TEE Ultrasound Transducer

Philips Ultrasound, Inc
FDA DevicesClass III

Philips OMNI III TEE Ultrasound Transducer

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

Apr 1, 2026Philips Ultrasound, Inc
Medical DeviceUSFDA Devices

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

I.T.S. GmbH
FDA DevicesClass II

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Apr 1, 2026I.T.S. GmbH
Medical DeviceUSFDA Devices

Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.

I.T.S. GmbH
FDA DevicesClass II

Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Apr 1, 2026I.T.S. GmbH
Medical DeviceUSFDA Devices

Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Olympus Corporation of the Americas
FDA DevicesClass II

Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Apr 1, 2026Olympus Corporation of the Americas