PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405637, 405707.
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405637, 405707.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet
BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use with the da Vinci¿ 5 system to aid in surgery. The components of the NIR handheld camera system include a camera control unit and a light source, NIR handheld camera, and additional components that together, connect to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from the NIR handheld camera system is displayed on the Tower monitor. Component: Yes, the SOCK is the component and integrated into the dV5 system
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use with the da Vinci¿ 5 system to aid in surgery. The components of the NIR handheld camera system include a camera control unit and a light source, NIR handheld camera, and additional components that together, connect to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from the NIR handheld camera system is displayed on the Tower monitor. Component: Yes, the SOCK is the component and integrated into the dV5 system
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
Brand Name: B. BRAUN MEDICAL INC. Product Name: SESK UPMC DOUBLE CATHETER TRAY Model/Catalog Number: 530134 Software Version: N/A Product Description: SESK UPMC DOUBLE CATHETER TRAY Component: No
Brand Name: B. BRAUN MEDICAL INC. Product Name: SESK UPMC DOUBLE CATHETER TRAY Model/Catalog Number: 530134 Software Version: N/A Product Description: SESK UPMC DOUBLE CATHETER TRAY Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333201); Spinocan 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333211); Spinocan 22 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333231)
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333201); Spinocan 25 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333211); Spinocan 22 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333231)
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
DeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)
DeRoyal Industries Inc. Ref 46-406, Covaderm Plus, Vascular Access Dressing 6"x6" (15cmx15cm)
Sterile wound dressing, lacks sterility assurance
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK HARRIS METHODIST Model/Catalog Number: 555498 Software Version: N/A Product Description: CESK HARRIS METHODIST Component: No
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK HARRIS METHODIST Model/Catalog Number: 555498 Software Version: N/A Product Description: CESK HARRIS METHODIST Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis
TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis
Due to incorrect functional length on device labeling.
Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System
Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.
Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 405723, 406704.
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 405723, 406704.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSK H M C 20061890 (560074), SSK S C H (560214), SSK BJC H S (ORIGINAL CUST) (560272), SSK CA P M C (ORIGINAL CUST) (560327), SSK T M H (560394), SSK S O 20070208 (560428), SSK L G H (ORIGINAL CUST) (560462), SSK H M H (560465), SSK BUPIVACAINE TRAY 2068824 (560476), SSK K F H (560497), SSK T J U H (560513), SSK O M C (560521), SSK CUSTOM SPINAL TRAY (560523), SSK O R M C (560533), SSK B S H S 20115122 (560538), SSK M M (560572), SSK M S 27GA PENCAN SPINAL TRAY (560580), SSK S T H & C (560584), SSK U C (560587), SSK P25BKG (560603), SSK P24BKG (560606), SSK S25BKG (560607), SSK M C - BUPIVACAINE TRAY (560609), SSK T M C (560613), SSK PENCIL POINT SPINAL TRAY (560619), SSK E O & S H-SPINAL (560621), SSK BSW SPINAL TRAY (560628), SSK M C OB DEPT - SPINAL (560645), SSK B H M (560646), SSK M G M C (560647), SSK S M H (560648), SSK H H S (560660), O S U M C SPINAL TRAY SSK (560667), A M C, SSK (560670), S B H - SPINAL TRAY SSK (560672);
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSK H M C 20061890 (560074), SSK S C H (560214), SSK BJC H S (ORIGINAL CUST) (560272), SSK CA P M C (ORIGINAL CUST) (560327), SSK T M H (560394), SSK S O 20070208 (560428), SSK L G H (ORIGINAL CUST) (560462), SSK H M H (560465), SSK BUPIVACAINE TRAY 2068824 (560476), SSK K F H (560497), SSK T J U H (560513), SSK O M C (560521), SSK CUSTOM SPINAL TRAY (560523), SSK O R M C (560533), SSK B S H S 20115122 (560538), SSK M M (560572), SSK M S 27GA PENCAN SPINAL TRAY (560580), SSK S T H & C (560584), SSK U C (560587), SSK P25BKG (560603), SSK P24BKG (560606), SSK S25BKG (560607), SSK M C - BUPIVACAINE TRAY (560609), SSK T M C (560613), SSK PENCIL POINT SPINAL TRAY (560619), SSK E O & S H-SPINAL (560621), SSK BSW SPINAL TRAY (560628), SSK M C OB DEPT - SPINAL (560645), SSK B H M (560646), SSK M G M C (560647), SSK S M H (560648), SSK H H S (560660), O S U M C SPINAL TRAY SSK (560667), A M C, SSK (560670), S B H - SPINAL TRAY SSK (560672);
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044
Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044
Temperature probe devices lack FDA clearance.
BD Pyxis MedStation ES REF: 323 Medication cabinet
BD Pyxis MedStation ES REF: 323 Medication cabinet
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
BD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
BD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
Atlan A350. Model Number: 8211500. anesthesia workstation
Atlan A350. Model Number: 8211500. anesthesia workstation
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK SHANDS JACKSONVILLE 20055717 Model/Catalog Number: 555884 Software Version: N/A Product Description: CESK SHANDS JACKSONVILLE 20055717 Component: No
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK SHANDS JACKSONVILLE 20055717 Model/Catalog Number: 555884 Software Version: N/A Product Description: CESK SHANDS JACKSONVILLE 20055717 Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.