Medizinprodukte Recalls

Defective surgical instruments, diagnostic equipment, and health device recalls.

Showing 24 of 2,119 results
Medical DeviceUSFDA Devices

PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353

GE Medical Systems Information Technologies Inc
FDA DevicesClass II

PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Sep 2, 2026GE Medical Systems Information Technologies Inc
Medical DeviceUSFDA Devices

PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

GE Medical Systems Information Technologies Inc
FDA DevicesClass II

PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Sep 2, 2026GE Medical Systems Information Technologies Inc
Medical DeviceUSFDA Devices

Medline convenience kits labeled as: 1) CMH LITHO MINOR PACK-LF, Medline Kit SKU DYNJ54362F; 2) EAST MAJOR LITHOTOMY, Medline Kit SKU DYNJ900957C; 3) GYN LAPAROTOMY, Medline Kit SKU DYNJ906804B; 4) UNIVERSAL GYN, Medline Kit SKU DYNJ910195A.

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits labeled as: 1) CMH LITHO MINOR PACK-LF, Medline Kit SKU DYNJ54362F; 2) EAST MAJOR LITHOTOMY, Medline Kit SKU DYNJ900957C; 3) GYN LAPAROTOMY, Medline Kit SKU DYNJ906804B; 4) UNIVERSAL GYN, Medline Kit SKU DYNJ910195A.

Kits contain a potentially contaminated component which were recalled by the manufacturer

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline convenience kits labeled as: 1) MAJOR GYN PACK, Medline Kit SKU DYNJG001030C; 2) GYN LITHOTOMY PACK, Medline Kit SKU DYNJT8166; 3) OPEN HEART PACK, Medline Kit SKU DYNJ47638B; 4) EAST DAVINCI, Medline Kit SKU DYNJ900949C; 5) THORACIC, Medline Kit SKU DYNJ903470G; 6) CV PART 1 AND 2, Medline Kit SKU DYNJ908879G; 7) GENERAL ROBOTIC, Medline Kit SKU DYNJ910178A; 8) AMB GYN LAPAROSCOPY KT, Medline Kit SKU DYNJ910281B; 9) DIEP FLAP, Medline Kit SKU DYNJ910832A; 10) C-SECTION, Medline Kit SKU DYNJ911236; 11) LAPAROTOMY, Medline Kit SKU DYNJ911243; 12) OPEN HEART A&B, Medline Kit SKU DYNJ911245; 13) TRANSPLANT LIVER BOWEL KIT-LF, Medline Kit SKU DYNJ911332; 14) TRANSPLANT LIVER BOWEL KIT-LF, Medline Kit SKU DYNJ911332A.

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits labeled as: 1) MAJOR GYN PACK, Medline Kit SKU DYNJG001030C; 2) GYN LITHOTOMY PACK, Medline Kit SKU DYNJT8166; 3) OPEN HEART PACK, Medline Kit SKU DYNJ47638B; 4) EAST DAVINCI, Medline Kit SKU DYNJ900949C; 5) THORACIC, Medline Kit SKU DYNJ903470G; 6) CV PART 1 AND 2, Medline Kit SKU DYNJ908879G; 7) GENERAL ROBOTIC, Medline Kit SKU DYNJ910178A; 8) AMB GYN LAPAROSCOPY KT, Medline Kit SKU DYNJ910281B; 9) DIEP FLAP, Medline Kit SKU DYNJ910832A; 10) C-SECTION, Medline Kit SKU DYNJ911236; 11) LAPAROTOMY, Medline Kit SKU DYNJ911243; 12) OPEN HEART A&B, Medline Kit SKU DYNJ911245; 13) TRANSPLANT LIVER BOWEL KIT-LF, Medline Kit SKU DYNJ911332; 14) TRANSPLANT LIVER BOWEL KIT-LF, Medline Kit SKU DYNJ911332A.

Kits contain a potentially contaminated component which were recalled by the manufacturer

Sep 2, 2026Medline Industries, LP
Medical DeviceUKUK

Filters should be used during the administration of Parenteral Nutrition for patients in all care settings DSI/2026/011

MHRAUK
uk-mhraMedical Device ...

Filters should be used during the administration of Parenteral Nutrition for patients in all care settings DSI/2026/011

Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome....

Sep 2, 2026MHRA
Medical DeviceUSFDA Devices

Medline medical convenience kit labeled as: TRACH TUBE KIT, Medline Kit SKU DYKM2330

Medline Industries, LP
FDA DevicesClass II

Medline medical convenience kit labeled as: TRACH TUBE KIT, Medline Kit SKU DYKM2330

Component labeling discrepancy

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Structur 2 SC Temporary Resin Cartridge Refill 75Gm, Product Codes: 1) 114-3933, Mfr Product Code 1479; Lots 2536179, 2606429 2) 114-3934, Mfr Product Code 1480; Lot 2539064 3) 114-3935, Mfr Product Code 1481; Lot 2533123 4) 114-3937, Mfr Product Code 1482; Lots 2533203, 2607170 is a fluorescent cold polymerizing paste-paste system for the production of temporary crowns, bridges, inlays and onlays.

Henry Schein Inc.
FDA DevicesClass II

Structur 2 SC Temporary Resin Cartridge Refill 75Gm, Product Codes: 1) 114-3933, Mfr Product Code 1479; Lots 2536179, 2606429 2) 114-3934, Mfr Product Code 1480; Lot 2539064 3) 114-3935, Mfr Product Code 1481; Lot 2533123 4) 114-3937, Mfr Product Code 1482; Lots 2533203, 2607170 is a fluorescent cold polymerizing paste-paste system for the production of temporary crowns, bridges, inlays and onlays.

Specific lots were distributed in the US with incorrect packaging and Instructions for Use (IFU) from the U.S. FDA-cleared device.

Sep 2, 2026Henry Schein Inc.
Medical DeviceUSFDA Devices

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Midmark Corporation
FDA DevicesClass II

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

Sep 2, 2026Midmark Corporation
Medical DeviceUSFDA Devices

Brand Name: Halyard Product Name: DAVINCI TRUCUSTOM BUNDLE containing TRAY FOLEY 16FR W/STATLOCK Model/Catalog Number: SAMM064-27

AVID Medical, Inc.
FDA DevicesClass II

Brand Name: Halyard Product Name: DAVINCI TRUCUSTOM BUNDLE containing TRAY FOLEY 16FR W/STATLOCK Model/Catalog Number: SAMM064-27

Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.

Sep 2, 2026AVID Medical, Inc.
Medical DeviceUSFDA Devices

Medline convenience kits labeled as: 1) LAMI FUSION, Medline Kit SKU CDS982035Y; 2) NON MB MJR NEURO CRANI SM PK, Medline Kit SKU DYNJT6104; 3) CRANIOTOMY, Medline Kit SKU DYNJ903471I; 4) KIT CRANIOTOMY RFD, Medline Kit SKU DYNJ906496K.

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits labeled as: 1) LAMI FUSION, Medline Kit SKU CDS982035Y; 2) NON MB MJR NEURO CRANI SM PK, Medline Kit SKU DYNJT6104; 3) CRANIOTOMY, Medline Kit SKU DYNJ903471I; 4) KIT CRANIOTOMY RFD, Medline Kit SKU DYNJ906496K.

Kits contain a potentially contaminated component which were recalled by the manufacturer

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline convenience kits labeled as: 1) MAJOR PROCEDURE, Medline Kit SKU CDS982040U; 2) OPERATIVE LAP P&S SURGICAL, Medline Kit SKU DYNJ89608; 3) LAPAROSCOPY, Medline Kit SKU DYNJ903479L; 4) LAPAROSCOPY, Medline Kit SKU DYNJ903479M.

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits labeled as: 1) MAJOR PROCEDURE, Medline Kit SKU CDS982040U; 2) OPERATIVE LAP P&S SURGICAL, Medline Kit SKU DYNJ89608; 3) LAPAROSCOPY, Medline Kit SKU DYNJ903479L; 4) LAPAROSCOPY, Medline Kit SKU DYNJ903479M.

Kits contain a potentially contaminated component which were recalled by the manufacturer

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).

Medline Industries, LP
FDA DevicesClass II

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD7801780-18) (Teleflex Product Number/SKU 780-18).

Medline Industries, LP
FDA DevicesClass II

Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD7801780-18) (Teleflex Product Number/SKU 780-18).

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

IMRIS Imaging Inc
FDA DevicesClass II

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Sep 2, 2026IMRIS Imaging Inc
Medical DeviceUSFDA Devices

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

IMRIS Imaging Inc
FDA DevicesClass II

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Sep 2, 2026IMRIS Imaging Inc
Medical DeviceUSFDA Devices

Medline convenience kits labeled as: 1) C-SECTION, Medline Kit SKU DYNJ904612L; 2) C-SECTION, Medline Kit SKU DYNJ911755; 3) AHR C-SECTION KIT, Medline Kit SKU DYNJ912014.

Medline Industries, LP
FDA DevicesClass II

Medline convenience kits labeled as: 1) C-SECTION, Medline Kit SKU DYNJ904612L; 2) C-SECTION, Medline Kit SKU DYNJ911755; 3) AHR C-SECTION KIT, Medline Kit SKU DYNJ912014.

Kits contain a potentially contaminated component which were recalled by the manufacturer

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories

Olympus Corporation of the Americas
FDA DevicesClass II

POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories

Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.

Sep 2, 2026Olympus Corporation of the Americas
Medical DeviceUSFDA Devices

CONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable Driver REF# CFBC-5502B 5.5mm GENESYs CrossFT Suture Anchor with two, #2 (5 metric) Hi-Fi Sutures Sterile

ConMed Corporation
FDA DevicesClass II

CONMED Linvatec GENESYS CrossFT Suture Anchor with Disposable Driver REF# CFBC-5502B 5.5mm GENESYs CrossFT Suture Anchor with two, #2 (5 metric) Hi-Fi Sutures Sterile

Suture anchor may contain incorrect patient implant card which may result in incorrect device identification and traceability.

Sep 2, 2026ConMed Corporation
Medical DeviceUSFDA Devices

Fresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.

Fresenius Medical Care Renal Therapies Group, LLC
FDA DevicesClass II

Fresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.

Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) has received complaints regarding flow-, no water-, or temperature-related alarms with the 2008T¿ BlueStar Hemodialysis machines.

Sep 2, 2026Fresenius Medical Care Renal Therapies Group, LLC
Medical DeviceUSFDA Devices

Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation

Merge Healthcare, Inc.
FDA DevicesClass II

Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation

Software issue may cause measurements to be displayed inaccurately.

Sep 2, 2026Merge Healthcare, Inc.
Medical DeviceUSFDA Devices

Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Canon Medical System, USA, INC.
FDA DevicesClass II

Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.

Sep 2, 2026Canon Medical System, USA, INC.
Medical DeviceUSFDA Devices

Medline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit SKU DYNJ40154D; 3) ENT PACK, Medline Kit SKU DYNJ40154F; 4) ENT SCOPE PACK, Medline Kit SKU DYNJ44480D; 5) ENT NASAL ORAL PACK MOSES, Medline Kit SKU DYNJ58928A; 6) HEAD & NECK PACK, Medline Kit SKU DYNJ83296; 7) TRANSPHENOIDAL PACK, Medline Kit SKU DYNJ89006; 8) SINUS, Medline Kit SKU DYNJ900537D; 9) SINUS, Medline Kit SKU DYNJ900537F; 10) ENT HEAD & NECK, Medline Kit SKU DYNJ903679F; 11) ENT HEAD & NECK, Medline Kit SKU DYNJ903679G; 12) ENT HEAD & NECK, Medline Kit SKU DYNJ903679I; 13) ENT HEAD & NECK, Medline Kit SKU DYNJ903679J; 14) ENT HEAD & NECK, Medline Kit SKU DYNJ903679K.

Medline Industries, LP
FDA DevicesClass II

Medline medical convenience kits, labeled as: 1) SINUS PACK-LF, Medline Kit SKU DYNJ0966146F; 2) ENT PACK, Medline Kit SKU DYNJ40154D; 3) ENT PACK, Medline Kit SKU DYNJ40154F; 4) ENT SCOPE PACK, Medline Kit SKU DYNJ44480D; 5) ENT NASAL ORAL PACK MOSES, Medline Kit SKU DYNJ58928A; 6) HEAD & NECK PACK, Medline Kit SKU DYNJ83296; 7) TRANSPHENOIDAL PACK, Medline Kit SKU DYNJ89006; 8) SINUS, Medline Kit SKU DYNJ900537D; 9) SINUS, Medline Kit SKU DYNJ900537F; 10) ENT HEAD & NECK, Medline Kit SKU DYNJ903679F; 11) ENT HEAD & NECK, Medline Kit SKU DYNJ903679G; 12) ENT HEAD & NECK, Medline Kit SKU DYNJ903679I; 13) ENT HEAD & NECK, Medline Kit SKU DYNJ903679J; 14) ENT HEAD & NECK, Medline Kit SKU DYNJ903679K.

Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU DYNJ58012A; 7) NEURO PACK SJ, Medline Kit SKU DYNJ63477A; 8) BASIC NEURO PACK, Medline Kit SKU DYNJ83071A.

Medline Industries, LP
FDA DevicesClass II

Medline medical convenience kits, labeled as: 1) NEURO CDS, Medline Kit SKU CDS985246J; 2) NEURO PACK, Medline Kit SKU DYNJ32770V; 3) NEURO BUNDLE-LF, Medline Kit SKU DYNJ40095D; 4) NEURO PACK-LF, Medline Kit SKU DYNJ42844D; 5) NEURO PACK, Medline Kit SKU DYNJ45121I; 6) NEURO LAMINECTOMY PACK, Medline Kit SKU DYNJ58012A; 7) NEURO PACK SJ, Medline Kit SKU DYNJ63477A; 8) BASIC NEURO PACK, Medline Kit SKU DYNJ83071A.

Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.

Sep 2, 2026Medline Industries, LP
Medical DeviceUSFDA Devices

Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03

AVID Medical, Inc.
FDA DevicesClass II

Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03

Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.